journal article·2021-02-01Open access

Control of gas concentration distribution in a semiconductor process chamber using CT-TDLAS measurement

Daisuke Hayashi, Yuhei Sakaguchi, Masakazu Minami

AIP Advances· Volume 11· Issue 2

Abstract

Methane (CH4) concentration distribution in a semiconductor process chamber was controlled using the measurement of computed tomography-tunable diode laser absorption spectroscopy (CT-TDLAS) and the feedback control toward the feeding CH4 concentrations and flow rates. CH4 diluted with nitrogen was fed into the chamber through a shower head having three separate and concentric areas. Thirty-two laser paths were configured in the chamber to collect the infrared absorption spectra for the CT-TDLAS measurement. The computed tomography calculation using the 32 spectra reconstructed the two-dimensional CH4 concentration distribution in the chamber. The measured concentration distribution was updated once per second. Based on the measured concentration distribution, the feedback control algorithm determined the feeding CH4 concentration and flow rate of each shower head area. In this work, we set the target distribution as a ring shape. In the control algorithm, first the feeding CH4 flow rate of each shower head area was adjusted to match the concentration peak radius in the measured distribution to the radius in the target distribution. Then, the feeding CH4 concentration of each area was adjusted in sequence to match the measured average concentration of each area to the corresponding concentration in the target distribution. The algorithm worked successfully, and the concentration distribution reached the target distribution. The extension of the application and its limitations were also discussed.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Spectroscopy and Laser Applications · Atmospheric and Environmental Gas Dynamics · Laser Design and Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

HORIBA STEC, Co., Ltd. , 11-5 Hokotate-cho, Kamitoba, Minami-ku, Kyoto 601-8116, Japan

References (26)

Show source-provided references
1. J. Phys. D: Appl. Phys. · 2014 · 47 · 27300110.1088/0022-3727/47/27/273001 ↗
2. Jpn. J. Appl. Phys., Part 1 · 2018 · 57 · 06JA0110.7567/jjap.57.06ja01 ↗
3. J. Phys. D: Appl. Phys. · 2018 · 51 · 15520110.1088/1361-6463/aab322 ↗
4. Jpn. J. Appl. Phys., Part 1 · 2014 · 53 · 03DC0210.7567/jjap.53.03dc02 ↗
5. Proc. SPIE · 2015 · 9428 · 94280910.1117/12.2178326 ↗
6. Proc. SPIE · 2018 · 10585 · 105851Q10.1117/12.2297213 ↗
7. Prog. Ene. Comb. Sci. · 2017 · 59 · 110.1016/j.pecs.2016.11.002 ↗
8. Appl. Opt. · 2005 · 44 · 657810.1364/ao.44.006578 ↗
9. Chem. Eng. J. · 2010 · 158 · 210.1016/j.cej.2008.10.026 ↗
10. Appl. Opt. · 2011 · 50 · A2910.1364/ao.50.000a29 ↗
11. IEEE Trans. Instrum. Meas. · 2014 · 63 · 306710.1109/tim.2014.2315737 ↗
12. Appl. Phys. Lett. · 2014 · 104 · 03410110.1063/1.4862754 ↗
13. Appl. Phys. Lett. · 2014 · 104 · 15410610.1063/1.4871976 ↗
14. Appl. Spectrosc. · 2020 · 74 · 21010.1177/0003702819888214 ↗
15. Int. J. Mech. Syst. Eng. · 2015 · 1 · 10910.15344/2455-7412/2015/109 ↗
16. Opt. Express · 2013 · 21 · 115210.1364/oe.21.001152 ↗
17. ECS J. Solid State Sci. Tech. · 2018 · 7 · Q21110.1149/2.0231811jss ↗
18. J. Phys. D: Appl. Phys. · 2019 · 52 · 48510710.1088/1361-6463/ab3c30 ↗
19. J. Vib. Test. Sys. Dyn. · 2020 · 4 · 29710.5890/jvtsd.2020.12.001 ↗
20. Appl. Therm. Eng. · 2014 · 73 · 145310.1016/j.applthermaleng.2014.05.063 ↗
21. ECS J. Electrochem. Soc. · 2003 · 150 · G41810.1149/1.1578481 ↗
22. Jpn. J. Appl. Phys., Part 1 · 2008 · 48 · 08HC0110.1143/jjap.48.08hc01 ↗
23. J. Vac. Sci. Technol. B · 2010 · 28 · 78910.1116/1.3456619 ↗
24. J. Vac. Sci. Technol. A · 2015 · 33 · 04130110.1116/1.4920933 ↗
25. IEEE Trans. Plasma Sci. · 2020 · 48 · 360610.1109/tps.2020.3023732 ↗
26. Rev. Chem. Eng. · 2007 · 23 · 65