A Parametric Study of Flushing Conditions for Improvement of Angioscopy Visibility

Kohei Mitsuzuka, Yujie Li, Toshio Nakayama, Hitomi Anzai, Daisuke Goanno, Simon Tupin, Mingzi Zhang, Haoran Wang, Kazunori Horie, Makoto Ohta

Abstract

During an angioscopy operation, a transparent liquid called dextran is sprayed out from a catheter to flush the blood away from the space between the camera and target. Medical doctors usually inject dextran at a constant flow rate. However, they often cannot obtain clear angioscopy visibility because the flushing out of the blood is insufficient. Good flushing conditions producing clear angioscopy visibility will increase the rate of success of angioscopy operations. This study aimed to determine a way to improve the clarity for angioscopy under different values for the parameters of the injection waveform, endoscope position, and catheter angle. We also determined the effect of a stepwise waveform for injecting the dextran only during systole while synchronizing the waveform to the cardiac cycle. To evaluate the visibility of the blood-vessel walls, we performed a computational fluid dynamics (CFD) simulation and calculated the visible area ratio (VAR), representing the ratio of the visible wall area to the total area of the wall at each point in time. Additionally, the normalized integration of the VAR called the area ratio (ARVAR) represents the ratio of the visible wall area as a function of the dextran injection period. The results demonstrate that the ARVAR with a stepped waveform, bottom endoscope, and three-degree-angle catheter results in the highest visibility, around 25 times larger than that under the control conditions: a constant waveform, a center endoscope, and 0 degrees. This set of conditions can improve angioscopy visibility.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Coronary Interventions and Diagnostics · Intracranial Aneurysms: Treatment and Complications · Cardiac and Coronary Surgery Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Institute of Fluid Science, Tohoku University, 2-1-1 Katahira, Aoba-ku, Sendai 980-8577, Japan

Graduate School of Biomedical Engineering, Tohoku University, 6-6 Azaaoba, Aramaki, Aoba-ku, Sendai 980-8579, Japan

Centre for Healthy Futures, Torrens University Australia, 1-51 Foveaux Street, Sydney, NSW 2010, Australia

National Institute of Technology, Nara College, 22 Yatacho, Yamatokoriyama 639-1080, Japan

Macquarie Medical School, Faculty of Medicine, Health, and Human Sciences, Macquarie University, 75 Talavera Rd., Sydney, NSW 2109, Australia

Department of Cardiology, Sendai Kousei Hospital, 4-15 Hirose, Aoba-ku, Sendai 980-0873, Japan

Funding

Japan Society for the Promotion of Science · Collaborative Research Project 2020, Institute of Fluid Science, Tohoku University

References (36)

Show source-provided references
1. Mortality from Ischemic Heart Disease: Analysis of Data from the World Health Organization and Coronary Artery Disease Risk Factors from NCD Risk Factor Collaboration10.1161/CIRCOUTCOMES.118.005375 ↗
2. Institute of Medicine (US) Committee on Social Security (2010). Cardiovascular Disability Criteria Cardiovascular Disability: Updating the Social Security Listings, National Academies Press.
3. Clinical Application of Coronary Angioscopy for Diagnosing and Treating Coronary Artery Disease
4. Assessment of Plaque Vulnerability by Angioscopic Classification of Plaque Color10.1016/j.ahj.2004.03.047 ↗
5. Intravascular Ultrasound as a Clinical Adjunct for Carotid Plaque Characterization10.1016/j.jvs.2013.09.028 ↗
6. Carotid Artery Stenting: Investigation of Plaque Protrusion Incidence and Prognosis10.1016/j.jcin.2017.01.029 ↗
7. Plaque Protrusion Detected by Intravascular Ultrasound during Carotid Artery Stenting10.1016/j.jstrokecerebrovasdis.2014.06.007 ↗
8. Optical Coherence Tomography after Carotid Stenting: Rate of Stent Malapposition, Plaque Prolapse and Fibrous Cap Rupture According to Stent Design10.1016/j.ejvs.2013.03.005 ↗
9. Yoshimura, S., Kawasaki, M., Yamada, K., Hattori, A., Nishigaki, K., Minatoguchi, S., and Iwama, T. (2013). Optical Coherence Tomography (OCT): A New Imaging Tool during Carotid Artery Stenting. Optical Coherence Tomography, IntechOpen.10.5772/53962 ↗
10. Comparative Evaluation of Angioscopy and Intravascular Ultrasound for Assessing Plaque Protrusion during Carotid Artery Stenting Procedures10.1016/j.wneu.2019.01.102 ↗
11. Continuing Evolution of Therapy for Coronary Artery Disease. Initial Results from the Era of Coronary Angioplasty10.1161/01.CIR.89.5.2015 ↗
12. Implication of Plaque Color Classification for Assessing Plaque Vulnerability: A Coronary Angioscopy and Optical Coherence Tomography Investigation10.1016/j.jcin.2007.11.001 ↗
13. Incomplete Neointimal Coverage of Sirolimus-Eluting Stents: Angioscopic Findings10.1016/j.jacc.2005.11.092 ↗
14. Early Detection of Vulnerable Atherosclerotic Plaque for Risk Reduction of Acute Aortic Rupture and Thromboemboli and Atheroemboli Using Non-Obstructive Angioscopy10.1253/circj.CJ-15-0126 ↗
15. The Healing Process of Infarct-Related Plaques: Insights from 18 Months of Serial Angioscopic Follow-Up10.1016/S0735-1097(01)01673-4 ↗
16. Experiments in Intracardiac Surgery10.1016/S0096-5588(20)31936-X ↗
17. Improving the Visual Field in Coronary Artery by with Non-Obstructive Angioscopy: Dual Infusion Method10.1007/s10554-017-1079-1 ↗
18. The Role of Coronary Angioscopy in the BRS Era10.7793/jcoron.23.005 ↗
19. Flush Flow Behaviour Affected by the Morphology of Intravascular Endoscope: A Numerical Simulation and Experimental Study10.3389/fphys.2021.733767 ↗
20. Yamakoshi, K., and Tanaka, S. (2011). Endoscope and Blood Vessel Endoscope System. (P2011-87859A), JP Patent.
21. Okayama, K., Nanto, S., and Sakata, Y. (2019). Vascular Endoscope Catheter and Vascular Endoscope. (P2019-115755A), JP Patent.
22. Faisal, S. (2018). Hemodynamics of Endoscopic Imaging of Chronic Total Occlusions. [Master’s Thesis, University of Washington].
23. Evaluation of 4D Flow MRI-Based Non-Invasive Pressure Assessment in Aortic Coarctations10.1016/j.jbiomech.2019.07.004 ↗
24. Xie, X., Wang, Y., Zhu, H., Zhou, H., and Zhou, J. (2013). Impact of Coronary Tortuosity on Coronary Blood Supply: A Patient-Specific Study. PLoS ONE, 8.10.1371/journal.pone.0064564 ↗
25. Effects of Arterial Geometry on Aneurysm Growth: Three-Dimensional Computational Fluid Dynamics Study10.3171/jns.2004.101.4.0676 ↗
26. Modeling of Light Transmittance Measurement in a Finite Layer of Whole Blood-a Collimated Transmittance Problem in Monte Carlo Simulation and Diffusion Model
27. Turbulence Modeling in Three-Dimensional Stenosed Arterial Bifurcations10.1115/1.2401182 ↗
28. Numerical Simulation of Hemodynamic Parameters of Turbulent and Pulsatile Blood Flow in Flexible Artery with Single and Double Stenoses10.1007/s12206-015-0752-3 ↗
29. Physiologic Blood Flow Is Turbulent10.1038/s41598-020-72309-8 ↗
30. Effects of Wall Compliance on Multiharmonic Pulsatile Flow in Idealized Cerebral Aneurysm Models: Comparative PIV Experiments10.1007/s00348-020-02998-4 ↗
31. Optimization of Flow Diverters for Cerebral Aneurysms10.1016/j.jocs.2011.12.006 ↗
32. Multiobjective Design Optimization of Stent Geometry with Wall Deformation for Triangular and Rectangular Struts10.1007/s11517-018-1864-6 ↗
33. Development of Computer-Based Simulator for Catheter Navigation in Blood Vessels (1st Report, Evaluation of Fundamental Parameters of Guidewire and Blood Vessel)10.1299/kikaic.72.2137 ↗
34. Development of Computer-Based Simulator for Catheter Navigation in Blood Vessels (2nd Report, Evaluation of Torquability of Guidewire)10.1299/kikaic.73.2988 ↗
35. Catheter and Guidewire Simulator for Intravascular Surgery (Comparison between Simulation Results and Medical Images)10.1007/978-3-642-03906-5_35 ↗
36. Takashima, K., Horie, S., Takenaka, M., Mukai, T., Ishida, K., and Ueda, Y. (2012, January 5–7). Fundamental Study on Medical Tactile Sensor Composed of Organic Ferroelectrics. Proceedings of the 2012 Fifth International Conference on Emerging Trends in Engineering and Technology, Himeji, Japan.10.1109/ICETET.2012.13 ↗