A Comparison Between Two Bearing Surfaces for Total Hip Arthroplasty—Ceramic-on-Ceramic and Metal–Polycarbonate–Urethane—A Pseudo-Randomized Study

Daniel Donaire Hoyas, Eladio Jiménez Mejías, Jesús Moreta, Manuel Sumillera García, Alberto Albert Ullibarri, Jorge Albareda Albareda

Journal of Functional Biomaterials· Volume 16· Issue 10

Abstract

Background: Polycarbonate–urethane (PCU) is a recently developed bearing surface used in prosthetic hip surgery. It offers several theoretical advantages, including an elasticity modulus similar to that of natural cartilage, good lubrication properties, low wear, and the possibility of using large heads. However, comparative clinical experience is limited. The purpose of this study was to analyze the results of the PCU bearing surface and compare them with those of ceramic-on-ceramic (CoC) bearings using the same femoral stem model. (2) Methods: Following a propensity score matching analysis of a prospectively collected database, patients with a primary total hip arthroplasty aged between 18 and 60 years were included. Subjects were divided into two groups (PCU and CoC). Demographic, patient satisfaction, and implant survival data were recorded. Clinical results were evaluated using the Harris Hip Score (HHS) and the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC). (3) Results: A total of 105 patients were included in each group. All patients exhibited a positive evolution on both the HHS and the WOMAC subscales between pre-op and one year post-op, no statistically significant differences being found between the groups with respect to improvement on the HHS (p = 0.172) or the pain (p = 0.523), stiffness (p = 0.448), and physical function (p = 0.255) subscales of the WOMAC. Head sizes in the PCU group were found to be larger, but this was not seen to have any effect on the patients’ clinical status or the prostheses’ dislocation rate. Although the complication rate was similar across the groups (p = 0.828), the incidence of squeaking was higher in the PCU group (p = 0.010). No differences were observed when comparing the implant survival rate (p = 0.427). nor in mean patient satisfaction (p = 0.138). (4) Conclusions: No differences were found in terms of clinical results, complications, implant survival, or patient satisfaction between the bearing surfaces under analysis, indicating that all of them are valid alternatives in total hip replacement, although the higher proportion of squeaking observed makes it advisable to exercise some caution.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Orthopedic Infections and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Servicio Cirugía Ortopédica y Traumatología, Hospital Universitario Poniente, 04700 El Ejido, Spain

Departamento de Medicina Preventiva y Salud Pública, Facultad de Medicina de la Universidad de Granada, 18016 Granada, Spain

Servicio Cirugía Ortopédica y Traumatología, Hospital San Juan de Dios Santurtzi, 48980 Santurce, Spain

Servicio de Cirugía Ortopédica y Traumatología, Hospital Universitario Marqués de Valdecilla, 39008 Santander, Spain

Servicio de Cirugía Ortopédica y Traumatología, Hospital Clínico Universitario Lozano Blesa, 50009 Zaragoza, Spain

References (37)

Show source-provided references
1. Bearing Surfaces for Total Hip Arthroplasty10.5435/JAAOS-D-15-00754 ↗
2. Outcomes of different bearings in total hip arthroplasty—Implant survival, revision causes, and patient-reported outcome
3. Biologic and Tribologic Considerations of Alternative Bearing Surfaces10.1097/00003086-200401000-00017 ↗
4. Biotribology of artificial hip joints10.5312/wjo.v6.i1.77 ↗
5. Total Hip Arthroplasty Bearing Surface Trends in the United States From 2007 to 2014: The Rise of Ceramic on Polyethylene10.1016/j.arth.2017.12.040 ↗
6. New alternate bearing surfaces in total hip arthroplasty: A review of the current literature10.1016/j.jcot.2017.10.013 ↗
7. Evaluation of the wear performance of a polycarbonate-urethane acetabular component in a hip joint simulator and comparison with UHMWPE and cross-linked UHMWPE10.1177/0885328210394471 ↗
8. The use of compliant layer prosthetic components in orthopedic joint repair and replacement: A review10.1002/jbm.b.33130 ↗
9. Wear rate evaluation of a novel polycarbonate-urethane cushion form bearing for artificial hip joints10.1016/j.actbio.2010.07.011 ↗
10. Cushion bearings versus large diameter head metal-on-metal bearings in total hip arthroplasty: A short-term metal ion study10.1007/s00402-011-1364-8 ↗
11. The Use of Polycarbonate-Urethane as an Acetabular Shell Bearing Surface: A 5-Year Prospective Study10.5301/hipint.5000483 ↗
12. Three-year results of a polycarbonate urethane acetabular bearing in total hip arthroplasty
13. Traumatic arthritis of the hip after dislocation and acetabular fractures: Treatment by mold arthroplasty. An end-result study using a new method of result evaluation10.2106/00004623-196951040-00012 ↗
14. Validation study of WOMAC: A health status instrument for measuring clinically-important patient-relevant outcomes following total hip or knee arthroplasty in osteoarthritis
15. Basgul, C., Spece, H., Kurtz, M.A., and Kurtz, S.M. (2025). Total Hip Arthroplasty Patient-Reported Outcome Measures Plateau by One Year: A Scoping Review and Meta-Analysis. J. Arthroplast., 1–10.10.1016/j.arth.2025.07.023 ↗
16. When does patient function “Plateau” after total joint arthroplasty? A cohort study10.1007/s00264-024-06248-8 ↗
17. Propensity score matching with R: Conventional methods and new features10.21037/atm-20-3998 ↗
18. Best Practice Guidelines for Propensity Score Methods in Medical Research: Consideration on Theory, Implementation, and Reporting. A Review10.1016/j.arthro.2021.06.037 ↗
19. R Core Team (2019, April 01). R: A Language and Environment for Statistical Computing. Available online: https://www.r-project.org/.
20. MatchIt: Nonparametric Preprocessing for Parametric Causal Inference10.18637/jss.v042.i08 ↗
21. The effect of bearing type on the outcome of total hip arthroplasty10.1080/17453674.2017.1405669 ↗
22. Updates in biomaterials of bearing surfaces in total hip arthroplasty10.1186/s42836-021-00092-6 ↗
23. Advances in zirconia toughened alumina biomaterials for total joint replacement10.1016/j.jmbbm.2013.03.022 ↗
24. Study of the Polycarbonate-Urethane/Metal Contact in Different Positions during Gait Cycle10.1155/2014/548968 ↗
25. A tribological study of UHMWPE acetabular cups and polyurethane compliant layer acetabular cups10.1002/1097-4636(2000)53:6<710::AID-JBM14>3.0.CO;2-R ↗
26. In vitro macrophage response to polyethylene and polycarbonate-urethane particles10.1002/jbm.a.32529 ↗
27. Is a Ceramic-on-Ceramic Bearing Really Superior to Ceramic-on-Polyethylene for Primary Total Hip Arthroplasty? A Systematic Review and Meta-Analysis of Randomised Controlled Trials
28. The hydroxyapatite-coated JRI-furlong hip. Outcome in patients under the age of 55 years10.1302/0301-620X.87B1.14656 ↗
29. Ben-Shlomo, Y., Blom, A., Boulton, C., Brittain, R., Clark, E., Dawson-Bowling, S., Deere, K., Esler, C., Espinoza, O., and Goldberg, A. (2021). The National Joint Registry 18th Annual Report 2021 [Internet], National Joint Registry.
30. Head size in primary total hip arthroplasty10.1302/2058-5241.3.170061 ↗
31. Mathematical evaluation of jumping distance in total hip arthroplasty10.3109/17453670902988378 ↗
32. Range of Motion and Stability in Total Hip Arthroplasty With 28-, 32-, 38-, and 44-mm Femoral Head Sizes10.1016/j.arth.2004.07.008 ↗
33. Bony impingement affects range of motion after total hip arthroplasty: A subject-specific approach10.1002/jor.20541 ↗
34. Effect of head diameter on passive and active dynamic hip dislocation10.1002/jor.22659 ↗
35. Biomechanics of Large Femoral Heads10.1097/01.blo.0000150117.42360.f9 ↗
36. Prevention of Dislocation After Total Hip Arthroplasty10.1016/j.arth.2018.01.047 ↗
37. The effect of acetabular cup orientations on limiting hip rotation10.1016/S0883-5403(99)90110-9 ↗