journal article·2021-06-06Open access

The Effect of Bottom Profile Dimples on the Femoral Head on Wear in Metal-on-Metal Total Hip Arthroplasty

J. Jamari, Muhammad Imam Ammarullah, Amir Putra Md Saad, Ardiyansyah Syahrom, Mohammad Uddin, Emile van der Heide, Hasan Basri

Abstract

Wear and wear-induced debris is a significant factor in causing failure in implants. Reducing contact pressure by using a textured surface between the femoral head and acetabular cup is crucial to improving the implant’s life. This study presented the effect of surface texturing as dimples on the wear evolution of total hip arthroplasty. It was implemented by developing finite element analysis from the prediction model without dimples and with bottom profile dimples of flat, drill, and ball types. Simulations were carried out by performing 3D physiological loading of the hip joint under normal walking conditions. A geometry update was initiated based on the patient’s daily routine activities. Our results showed that the addition of dimples reduced contact pressure and wear. The bottom profile dimples of the ball type had the best ability to reduce wear relative to the other types, reducing cumulative linear wear by 24.3% and cumulative volumetric wear by 31% compared to no dimples. The findings demonstrated that surface texturing with appropriate dimple bottom geometry on a bearing surface is able to extend the lifetime of hip implants.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Advanced materials and composites · Orthopedic Infections and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Mechanical Engineering, Faculty of Engineering, Diponegoro University, Tembalang, Semarang 50275, Central Java, Indonesia

Applied Mechanics and Design, School of Mechanical Engineering, Faculty of Mechanical Engineering, Universiti Teknologi Malaysia, Skudai 81310, Johor Bahru, Malaysia

Medical Device and Technology Center (MEDiTEC), Institute of Human-Centered and Engineering (IHumEn), Universiti Teknologi Malaysia, Skudai 81310, Johor Bahru, Malaysia

UniSA STEM, University of South Australia, Mawson Lakes Campus, Mawson Lakes, SA 5095, Australia

Laboratory for Surface Technology and Tribology, Faculty of Engineering Technology, University of Twente, P.O. Box 217, 7500 AE Enschede, The Netherlands

Department of Mechanical Engineering, Faculty of Engineering, Sriwijaya University, Indralaya 30662, South Sumatera, Indonesia

Funding

Sriwijaya University

References (30)

Show source-provided references
1. Recent Updates for Biomaterials Used in Total Hip Arthroplasty10.1186/s40824-018-0144-8 ↗
2. Long-Term Contact-Coupled Wear Prediction for Metal-on-Metal Total Hip Joint Replacement10.1243/13506501JET592 ↗
3. Australian Orthopaedic Association, and National Joint Replacement Registry (2020). Annu. Report 2020, Australian Orthopaedic Association. [2020th ed.].
4. Mahyudin, F., and Hermawan, H. (2016). Biomaterials and Medical Devices—A Perspective from an Emerging Country, Springer.10.1007/978-3-319-14845-8 ↗
5. A Novel Low Wearing Differential Hardness, Ceramic-on-Metal Hip Joint Prosthesis10.1016/S0021-9290(01)00096-3 ↗
6. Contact Mechanics and Elastohydrodynamic Lubrication in a Novel Metal-on-Metal Hip Implant with an Aspherical Bearing Surface10.1016/j.jbiomech.2009.11.018 ↗
7. Metal-on-Metal Hip Arthroplasty: A Review of Adverse Reactions and Patient Management10.3390/jfb6030486 ↗
8. Diamond-like Carbon Coatings with Zirconium-Containing Interlayers for Orthopedic Implants10.1016/j.jmbbm.2017.01.023 ↗
9. The Analysis of the Dimple Arrangement of the Artificial Hip Joint to the Performance of Lubrication10.1088/1757-899X/620/1/012116 ↗
10. The Analysis of Dimple Geometry on Artificial Hip Joint to the Performance of Lubrication10.1088/1742-6596/1198/4/042012 ↗
11. Contact Pressure Analysis of Acetabular Cup Surface with Dimple Addition on Total Hip Arthroplasty Using Finite Element Method10.1088/1757-899X/1034/1/012001 ↗
12. Investigation on Inner Flow Field Characteristics of Groove Textures in Fully Lubricated Thrust Bearings10.1108/ILT-08-2017-0224 ↗
13. Mechanical Micro-Texturing of Ti-6Al-4V Surfaces for Improved Wettability and Bio-Tribological Performances10.1016/j.surfcoat.2018.05.056 ↗
14. The impact of surface and geometry on coefficient of friction of artificial hip joints10.1016/j.jmbbm.2017.05.011 ↗
15. A Finite Element Method Comparison of Wear in Two Metal-on-Metal Total Hip Prostheses10.1243/09544119JEIM148 ↗
16. Development of Computational Wear Simulation of Metal-on-Metal Hip Resurfacing Replacements10.1016/j.jbiomech.2007.09.020 ↗
17. Jamari, J., Saputra, E., Anwar, I.B., and Van Der Heide, E. (2019). Study of an Additional Layer of Cement Mantle Hip Joints for Reducing Cracks. J. Funct. Biomater., 10.10.3390/jfb10030040 ↗
18. Predicting the Wear of Hard-on-Hard Hip Joint Prostheses10.1016/j.wear.2013.01.009 ↗
19. Contact Stress and Wear Analysis of Zirconia Against Alumina for Normal and Physically Demanding Loads in Hip Prosthesis10.1007/s42235-020-0078-4 ↗
20. Computational Wear Assessment of Hard on Hard Hip Implants Subject to Physically Demanding Tasks10.1007/s11517-017-1739-2 ↗
21. Temporal and Spatial Distributions of Directional Counterface Motion at the Acetabular Bearing Surface in Total Hip Arthroplasty
22. Effect of Microseparation on Contact Mechanics in Ceramic-on-Ceramic Hip Joint Replacements
23. Hip contact forces and gait patterns from routine activities10.1016/S0021-9290(01)00040-9 ↗
24. Contact and Rubbing of Flat Surfaces10.1063/1.1721448 ↗
25. Wear and Lubrication of Metal-on-Metal Hip Implants10.1097/00003086-199912000-00003 ↗
26. Comparison of Alloys and Designs in a Hip Simulator Study of Metal on Metal Implants10.1097/00003086-199608001-00015 ↗
27. Retrieval Study of Uncemented Metal-Metal Hip Prostheses Revised for Early Loosening10.1016/S0142-9612(02)00410-6 ↗
28. Hip Simulator Wear Comparison of Metal-on-Metal, Ceramic-on-Ceramic and Crosslinked UHMWPE Bearings10.1016/j.wear.2005.02.104 ↗
29. Accuracy of Measurement of Polyethylene Wear with Use of Radiographs of Total Hip Replacements10.2106/00004623-200312000-00016 ↗
30. Predictive Wear Modeling of the Articulating Metal-on-Metal Hip Replacements10.1002/jbm.b.33568 ↗