Development of imaging energy analyzer using multipole Wien filter
Source abstract
Abstract
We discussed a new design of a Wien filter energy analyzer for an energy-filtered X-ray photoemission electron microscopy system. We have demonstrated that the second-order aberration and the third-order aperture aberration can be corrected by the multipole Wien filter by adjusting multipole components of electric and magnetic fields up to octupole components. The three-dimensional charge simulation method indicated that 12 electrodes and magnetic poles can effectively reproduce these ideal electric and magnetic fields.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Electron and X-Ray Spectroscopy Techniques · Advanced Electron Microscopy Techniques and Applications · Advanced X-ray Imaging Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (7)
Show source-provided references
1. Tonner · J. Electron. Spectrosc. Relat. Phenom. · 1997 · 84 · 21110.1016/S0368-2048(97)00005-4 ↗
2. Bauer · J. Phys. Condens.: Mat. · 2001 · 13 · 1139
3. Yasufuku · Chem. Lett. · 2002 · 8 · 84210.1246/cl.2002.842 ↗
4. Rotermund · Nature · 1990 · 343 · 35510.1038/343355a0 ↗
5. Rose · Optik · 1987 · 77 · 26
6. Hawkes · 1989
7. M. Kato, Dissertation, The University of Tokyo, 1997