journal article·2005-02

Development of imaging energy analyzer using multipole Wien filter

H. Niimi, M. Kato, T. Tsutsumi, T. Kawasaki, H. Matsudaira, S. Suzuki, W.-J. Chun, Y. Kitajima, M. Kudo, K. Asakura

Applied Surface Science· Volume 241· Issue 1-2
Publisher page

Abstract

We discussed a new design of a Wien filter energy analyzer for an energy-filtered X-ray photoemission electron microscopy system. We have demonstrated that the second-order aberration and the third-order aperture aberration can be corrected by the multipole Wien filter by adjusting multipole components of electric and magnetic fields up to octupole components. The three-dimensional charge simulation method indicated that 12 electrodes and magnetic poles can effectively reproduce these ideal electric and magnetic fields.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Electron and X-Ray Spectroscopy Techniques · Advanced Electron Microscopy Techniques and Applications · Advanced X-ray Imaging Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (7)

Show source-provided references
1. Tonner · J. Electron. Spectrosc. Relat. Phenom. · 1997 · 84 · 21110.1016/S0368-2048(97)00005-4 ↗
2. Bauer · J. Phys. Condens.: Mat. · 2001 · 13 · 1139
3. Yasufuku · Chem. Lett. · 2002 · 8 · 84210.1246/cl.2002.842 ↗
4. Rotermund · Nature · 1990 · 343 · 35510.1038/343355a0 ↗
5. Rose · Optik · 1987 · 77 · 26
6. Hawkes · 1989
7. M. Kato, Dissertation, The University of Tokyo, 1997