Abstract

Abstract We are developing a method to conduct SIMS analysis at high mass resolution (m/Δm>50,000), to facilitate the examination and study of complex organic and biomolecules on surfaces. The approach uses a primary-ion beam probe (rastered 25 keV Ga+ ion source), providing high (100 nm) spatial resolution, and an ion cyclotron resonance (ICR) cell, capable of mass analysis at a resolution in excess of 105 and mass accuracy of less than 1.0 ppm. The apparatus includes a time-of-flight (ToF) mass analyzer, offering rapid chemical mapping at low (m/Δm 300), although, readily detected by the ToF analyzer, are inefficiently transmitted to the ICR cell. This has stimulated the design of a new ion optics coupling arrangement, which provides a higher mass resolution over a wider mass range.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Ion-surface interactions and analysis · Mass Spectrometry Techniques and Applications · Analytical chemistry methods development

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (4)

Show source-provided references
1. Pacholski · Chem. Rev. · 1999 · 99 · 297710.1021/cr980137w ↗
2. Benninghoven · J. Vac. Sci. Technol. A · 1985 · 3 · 45110.1116/1.573015 ↗
3. Kotula · Microsc. Microanal. · 2003 · 9 · 110.1017/S1431927603030058 ↗
4. Kovtoun · J. Am. Soc. Mass Spectrom. · 2002 · 13 · 13510.1016/S1044-0305(01)00346-4 ↗