journal article·2012-07

Corrosion and ion release behavior of Cu/Ti film prepared via physical vapor deposition in vitro as potential biomaterials for cardiovascular devices

Hengquan Liu, Deyuan Zhang, Feng Shen, Gui Zhang, Shenhua Song

Applied Surface Science· Volume 258· Issue 19
Publisher page

Abstract

Abstract Cu/Ti films of various Cu/Ti ratios were prepared on a TiNi alloy via vacuum arc plasma deposition. The phase composition, structure, and concentration of elements were investigated via X-ray diffraction and X-photoelectron energy spectrum. The hemolysis ratio and platelet adhesion of the different films were characterized to evaluate blood compatibility. The corrosion and ion release behavior were investigated via a typical immersion test and electrochemical method. The growth of endothelial cells (ECs) was investigated, and methylthiazolyte-trazolium method was employed to evaluate the effect of Cu2+. The sophisticated films showed good compatibility. However, with increasing quality ratio of Cu/Ti, the hemolysis ratio increased, and some platelets started to break slightly. The Cu2+ release was gradually stabilized. The open circuit potential of the Cu/Ti film-modified samples was lower than that of the TiNi substrate. The polarization test result indicates that the passivation stability performance of Cu/Ti film samples is less than the TiNi substrate, and is favorable to Cu2+ release. The adhesion and proliferation of ECs would be inhibited with 10 wt.% Cu concentration of the film, and ECs would undergo apoptosis at >50 wt.% concentration. A Cu/Ti film with good compatibility and anti-endothelialization has potential applications for special cardiovascular devices.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Metal and Thin Film Mechanics · Corrosion Behavior and Inhibition · Polymer Surface Interaction Studies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (24)

Show source-provided references
1. Long-duration temporary vena cava filter: a prospective 104-case multicenter study10.1016/j.jvs.2006.02.004 ↗
2. Inferior vena cave filter tilt and asymmetry: comparison of the over-the-wire stainless-steel and titanium Green field Inferior vena cave filters10.1016/S1051-0443(97)70706-1 ↗
3. Coronary stents: a materials perspective10.1016/j.biomaterials.2006.11.042 ↗
4. Development in intracoronary stents
5. In vivo investigation of blood compatibility of titanium oxide films10.1002/(SICI)1097-4636(199810)42:1<128::AID-JBM16>3.0.CO;2-H ↗
6. Passive stent coatings in the drug-eluting era
7. Comparison of diamond-like carbon-coated stents versus uncoated stainless steel stents in coronary artery disease10.1016/j.amjcard.2003.10.048 ↗
8. Hemocompatibility of titanium oxide films10.1016/S0142-9612(03)00046-2 ↗
9. Corrosion susceptibility Investigation of Ti–O films modified L-605 vascular stents by cyclic potentiodynamic polarzition measurement10.1016/j.surfcoat.2011.04.048 ↗
10. Biomaterial-associated thrombosis: roles of coagulation factors, complement, platelets and leukocytes10.1016/j.biomaterials.2004.01.023 ↗
11. Blood compatibility of sputter-deposited alumina films10.1002/jbm.820280212 ↗
12. Efficacy of anticoagulation post-inferior vena caval filter placement
13. Heparin and low-molecular-weight heparin: the seventh ACCP conference on antithrombotic and thrombolytic therapy10.1378/chest.126.3_suppl.188S ↗
14. Filter complications and their management
15. Copper toxicity, oxidative stress, and antioxidant nutrients10.1016/S0300-483X(03)00159-8 ↗
16. The lattice parameter of the α copper–titanium solid solution10.1107/S0021889870006787 ↗
17. Water quality determination of copper-2,9-dimethy-1,10-phenanthroline was conducted using spectrophotometric methods, UDC614.777:543.42:546.56.
18. Cytotoxicity of fusarium mycotoxins to mammalian cell cultures as determined by the MTT bioassay10.1016/j.fct.2005.01.016 ↗
19. Yang · 2003
20. Fretting study of surface modified Ni–Ti shape memory alloy10.1023/A:1014735225436 ↗
21. Hemocompatibility of surface-modified, silicon-incorporated, diamond-like carbon films10.1016/j.actbio.2008.07.031 ↗
22. Effect of copper on blood coagulation of Oreochromis mossambicus (Cichlidae)
23. Biomaterial associated thrombosis: roles of coagulation factors, complement, platelets and leukocytes10.1016/j.biomaterials.2004.01.023 ↗
24. Detection of shortened activated partial thromboplastin times: an evaluation of different commercial reagents10.1016/j.thromres.2007.05.006 ↗