Digital measurement of positron lifetime
Source abstract
Abstract
A prototype of a digital positron lifetime apparatus is presented. We demonstrate that direct digitizing of detector pulses and subsequent simple analysis with software, can be used to replace the conventional analog electronics chain (constant-fraction discriminators, time-to-amplitude converter and multichannel analyzer). In this work, we use a fast digital oscilloscope. The quality of the lifetime data is shown to be as good as with a usual apparatus. For a pulsed positron lifetime beam the digital system is particularly suitable as no coincidence detection is needed and only one analog pulse has to be analyzed.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Muon and positron interactions and applications · Particle Detector Development and Performance · Advancements in Battery Materials
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (6)
Show source-provided references
1. A. Dupasquier, A.P. Mills, Jr. (Ed.), Positron Spectroscopy of Solids, IOS Press, Amsterdam, 1995
2. K. Saarinen, P. Hautojärvi, C. Corbel, in: M. Stavola (Ed.), Identification of Defects in Semiconductors, Academic Press, New York, 1998
3. R. Krause-Rehberg, H.S. Leipner, Positron Annihilation in Semiconductors, Springer, Heidelberg, 1999.10.1007/978-3-662-03893-2 ↗
4. Codino · Nucl. Instr. and Meth. A · 2000 · 440 · 19110.1016/S0168-9002(99)00813-X ↗
5. Bec̆vár̆ · Nucl. Instr. and Meth. A · 2000 · 443 · 55710.1016/S0168-9002(99)01156-0 ↗
6. Schödlbauer · Nucl. Instr. and Meth. B · 1988 · 34 · 25810.1016/0168-583X(88)90752-5 ↗