Personalized hip implants manufacturing and testing
Source abstract
Abstract
Abstract Two models of Ti6Al4V personalized femoral stems for hip replacement have been designed and laser sintered with different sizes of fenestrated architecture that mimics the natural structure of bone, ensuring postoperative bone ingrowth and increasing the elasticity of the entire structure. They were tested statically and dynamically versus a commercial femoral stem. Mechanical tests were performed in order to determine the fatigue limit using the Locati method. The tests were conducted in a thermostatic bath (37°±1°) with the implants immersed in distilled water salted solution 0.91%. For probe embedment poly-methyl methacrylate (PMMA) was used. The characteristic curves of the two personalized fenestrated implants reveal an elastic behaviour by their nonlinear appearance. After dynamic tests an inverse relationship between displacements obtained in the static tests and the fatigue limit was observed. Large fenestrations conferred the desired elasticity to the implant, but contributed to a life service reduction. The fatigue limit for both implants was much above the minimum value specified by ISO 7602: 2010, so both models can be safely used in the medical practice, leading to increased life service of implants.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Dental materials and restorations · Bone Tissue Engineering Materials
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (22)
Show source-provided references
1. Darrell · 2017
2. Problem of stress shielding and improvement to the hip implant designs: a review10.3923/jms.2007.460.467 ↗
3. H. Orban, D. Bunea, L. Moldovan, I. Antoniac, D. Gheorghiu, Semenescu A., Componentă femurală pentru proteză totală de șold, Brevet de invenție RO 117890 B, Oficiul de stat pentru invenții și mărci, 30 Sept. 2002.
4. Next generation orthopaedic implants by additive manufacturing using electron beam melting
5. Fully porous 3D printed titanium femoral stem to reduce stress-shielding following total hip arthroplasty
6. Direct metal fabrication of titanium implants with tailored materials and mechanical properties using electron beam melting technology10.1016/j.msec.2007.04.022 ↗
7. Compression-compression fatigue of selective electron beam melted cellular titanium (Ti-6Al-4V)10.1002/jbm.b.31901 ↗
8. MIMICS: Software & Services for Biomedical Engineering, http://biomedical.materialise.com/mimics (Accessed 28.03.2015).
9. SolidWorks: 3D CAD solutions, http://www.solidworks.com/sw/products/3d-cad/packages.htm (Accessed 15.06.2016).
10. EOS e-Manufacturing Solutions: Industrial 3D Printing, http://www.eos.info/en (Accessed 23.05.2016)
11. ISO 7206-4:2002: Implants for surgery – Partial and total hip prostheses, Part 4-Determination of endurance properties of stemmed femoral components.
12. ISO 7206-8:2002: Implants for surgery – Partial and total hip prostheses, Part 8- Endurance performances of stemmed femoral components.
13. Pacioga · 2017
14. 2017
15. 2017
16. 2017
17. 2017
18. Testing titanium implants
19. Campioni · 2013 · 81
20. Computer simulation on fatigue behavior of cemented hip prostheses: a physiological model10.1016/j.cmpb.2004.05.004 ↗
21. Long-term clinical consequences of stress-shielding after total hip arthroplasty without cement10.2106/00004623-199707000-00006 ↗
22. Andreykiv · 2006 · 77