Reducing the foreign body response on human cochlear implants and their materials in vivo with photografted zwitterionic hydrogel coatings

Ryan Horne, Nir Ben-Shlomo, Megan Jensen, Morgan Ellerman, Caleb Escudero, Rong Hua, Douglas Bennion, C Allan Guymon, Marlan R. Hansen

Acta Biomaterialia· Volume 166

Abstract

The foreign body response to implanted materials often complicates the functionality of sensitive biomedical devices. For cochlear implants, this response can reduce device performance, battery life and preservation of residual acoustic hearing. As a permanent and passive solution to the foreign body response, this work investigates ultra-low-fouling poly(carboxybetaine methacrylate) (pCBMA) thin film hydrogels that are simultaneously photo-grafted and photo-polymerized onto polydimethylsiloxane (PDMS). The cellular anti-fouling properties of these coatings are robustly maintained even after six-months subcutaneous incubation and over a broad range of cross-linker compositions. On pCBMA-coated PDMS sheets implanted subcutaneously, capsule thickness and inflammation are reduced significantly in comparison to uncoated PDMS or coatings of polymerized poly(ethylene glycol dimethacrylate) (pPEGDMA). Further, capsule thickness is reduced over a wide range of pCBMA cross-linker compositions. On cochlear implant electrode arrays implanted subcutaneously for one year, the coating bridges over the exposed platinum electrodes and dramatically reduces the capsule thickness over the entire implant. Coated cochlear implant electrode arrays could therefore lead to persistent improved performance and reduced risk of residual hearing loss. More generally, the in vivo anti-fibrotic properties of pCBMA coatings also demonstrate potential to mitigate the fibrotic response on a variety of sensing/stimulating implants. STATEMENT OF SIGNIFICANCE: This article presents, for the first time, evidence of the in vivo anti-fibrotic effect of zwitterionic hydrogel thin films photografted to polydimethylsiloxane (PDMS) and human cochlear implant arrays. The hydrogel coating shows no evidence of degradation or loss of function after long-term implantation. The coating process enables full coverage of the electrode array. The coating reduces fibrotic capsule thickness 50-70% over a broad range of cross-link densities for implantations from six weeks to one year.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Advanced Sensor and Energy Harvesting Materials · Conducting polymers and applications · Polymer Surface Interaction Studies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (41)

Show source-provided references
1. Continuous glucose monitoring: 40 years, what we′ve learned and what's next10.1002/cphc.201300172 ↗
2. Wire tethering or ‘bowstringing’ as a long-term hardware-related complication of deep brain stimulation10.1159/000236369 ↗
3. Modulating the foreign body response of implants for diabetes treatment10.1016/j.addr.2021.01.011 ↗
4. Intracochlear inflammatory response to cochlear implant electrodes in humans10.1097/MAO.0000000000000540 ↗
5. Foreign body reaction to biomaterials10.1016/j.smim.2007.11.004 ↗
6. The formation and function of granulomas10.1146/annurev-immunol-032712-100022 ↗
7. Foreign body or hypersensitivity granuloma of the inner ear after cochlear implantation: one possible cause of a soft failure?10.1097/MAO.0b013e31818c33cf ↗
8. Effects of loss of residual hearing on speech performance with the CI422 and the Hybrid-L electrode10.1179/1754762815Y.0000000008 ↗
9. Post hybrid cochlear implant hearing loss and endolymphatic hydrops10.1097/MAO.0000000000001199 ↗
10. Intracochlear fibrosis and the foreign body response to cochlear implant biomaterials10.1002/lio2.329 ↗
11. The profile of adsorbed plasma and serum proteins on methacrylic acid copolymer beads: effect on complement activation10.1016/j.biomaterials.2016.11.036 ↗
12. Foreign body reaction to subcutaneous implants10.1007/978-3-319-18603-0_6 ↗
13. Covalently grafted 2-methacryloyloxyethyl phosphorylcholine networks inhibit fibrous capsule formation around silicone breast implants in a porcine model10.1021/acsami.0c07629 ↗
14. Conformal hydrogel coatings on catheters to reduce biofouling10.1021/acs.langmuir.8b03074 ↗
15. Ultralow-fouling, functionalizable, and hydrolyzable zwitterionic materials and their derivatives for biological applications10.1002/adma.200901407 ↗
16. An introduction to zwitterionic polymer behavior and applications in solution and at surfaces10.1039/C8CS00508G ↗
17. Functionalizable and nonfouling zwitterionic carboxybetaine hydrogels with a carboxybetaine dimethacrylate crosslinker10.1016/j.biomaterials.2010.09.067 ↗
18. Zwitterionic carboxybetaine polymer surfaces and their resistance to long-term biofilm formation10.1016/j.biomaterials.2009.05.058 ↗
19. Zwitterionic poly(carboxybetaine) hydrogels for glucose biosensors in complex media10.1016/j.bios.2010.10.031 ↗
20. Lubrication at physiological pressures by polyzwitterionic brushes10.1126/science.1169399 ↗
21. Biomimetic phosphorylcholine polymer grafting from polydimethylsiloxane surface using photo-induced polymerization10.1016/j.biomaterials.2006.05.046 ↗
22. Zwitterionic PEG-PC hydrogels modulate the foreign body response in a modulus-dependent manner10.1021/acs.biomac.8b00444 ↗
23. Vascular catheters with a nonleaching poly-sulfobetaine surface modification reduce thrombus formation and microbial attachment10.1126/scitranslmed.3004120 ↗
24. Carboxybetaine, sulfobetaine, and cationic block copolymer coatings: a comparison of the surface properties and antibiofouling behavior10.1002/app.35233 ↗
25. Superlow fouling sulfobetaine and carboxybetaine polymers on glass slides10.1021/la062175d ↗
26. Antifouling and mechanical properties of photografted zwitterionic hydrogel thin-film coatings depend on the cross-link density10.1021/acsbiomaterials.1c00852 ↗
27. A. Peel, R. Horne, D.M. Bennion, C.A. Guymon, M.R. Hansen, Durability of photografted zwitterionic hydrogel coatings for reduction of the foreign body response to cochlear implants, bioRxiv (2022) 2022.06.24.497518.10.1101/2022.06.24.497518 ↗
28. Photograftable zwitterionic coatings prevent staphylococcus aureus and staphylococcus epidermidis adhesion to PDMS surfaces10.1021/acsabm.0c01147 ↗
29. Cochlear implant material effects on inflammatory cell function and foreign body response10.1016/j.heares.2022.108597 ↗
30. Hydrogel coated and dexamethasone releasing cochlear implants: quantification of fibrosis in guinea pigs and evaluation of insertion forces in a human cochlea model10.1002/jbm.b.33187 ↗
31. Polymer coatings of cochlear implant electrode surface - an option for improving electrode-nerve-interface by blocking fibroblast overgrowth10.1371/journal.pone.0157710 ↗
32. Biodegradable polymeric coatings on cochlear implant surfaces and their influence on spiral ganglion cell survival10.1002/jbm.b.33110 ↗
33. Antifouling photograftable zwitterionic coatings on PDMS substrates10.1021/acs.langmuir.8b00838 ↗
34. Zwitterionic Photografted Coatings of Cochlear Implant Biomaterials Reduce Friction and Insertion Forces10.1097/MAO.0000000000003288 ↗
35. Mechanical properties and thermal behaviour of PEGDMA hydrogels for potential bone regeneration application10.1016/j.jmbbm.2011.04.004 ↗
36. Delayed loss of hearing after hearing preservation cochlear implantation: human temporal bone pathology and implications for etiology10.1016/j.heares.2015.08.018 ↗
37. Timing of acoustic hearing changes after cochlear implantation10.1002/lary.29984 ↗
38. Access and polarization electrode impedance changes in electric-acoustic stimulation cochlear implant users with delayed loss of acoustic hearing10.1007/s10162-021-00809-z ↗
39. Delayed changes in auditory status in cochlear implant users with preserved acoustic hearing10.1016/j.heares.2017.04.005 ↗
40. Cochlear implants: causes, effects and mitigation strategies for the foreign body response and inflammation10.1016/j.heares.2022.108536 ↗
41. Advances in hearing preservation in cochlear implant surgery10.1097/MOO.0000000000000742 ↗