Wireless temperature measurement based on magnetic harmonic signal during magnetic heating operation for hyperthermia
Source abstract
Abstract
We developed a wireless temperature measurement method based on magnetic harmonic signals during magnetic heating and evaluated highly concentrated magnetic nanoparticles (MNPs) for hyperthermia applications. XRD confirmed that the iron oxide composition of Resovist® remained unchanged after concentration, while TEM and DLS analyses demonstrated reversible dispersibility between concentrated and diluted states. Magnetization measurements showed an increase in saturation magnetization, and the AC hysteresis loop area of fourfold-concentrated Resovist® was 3.8 times larger than that of the original sample, indicating enhanced heating efficiency. The detectable distance and volume for temperature measurement improved from 5 mm to 12 mm and from 10 µL to 5 µL, respectively. The developed system achieved noninvasive temperature measurement with an error of 0.07°C and constant-temperature control with a deviation of 0.56°C, demonstrating sufficient precision for biomedical applications.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Characterization and Applications of Magnetic Nanoparticles · Nanoparticle-Based Drug Delivery · Ultrasound and Hyperthermia Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Graduate School of Engineering, Tohoku University 1 , Sendai,
Graduate School of Biomedical Engineering, Tohoku University 2 , Sendai,
Funding
Nakatani Foundation for Advancement of Measuring Technologies in Biomedical Engineering · Iketani Science and Technology Foundation · Japan Agency for Medical Research and Development · Tohoku University Business Incubation Program · JSPS KAKENHI · JST FOREST Program
References (25)
Show source-provided references
1. Heating technology for malignant tumors: A review10.1080/02656736.2020.1779357 ↗
2. Magnetic nanoparticles in cancer theranostics10.7150/thno.11544 ↗
3. Theranostics based on magnetic nanoparticles and polymers: Intelligent design for efficient diagnostics and therapy10.3389/fchem.2020.00561 ↗
4. Animal experiment with PID-based automatic temperature control system for magnetic hyperthermia
5. Wireless temperature monitoring by using magnetic nanoparticles for biomedical applications on magnetic hyperthermia treatment10.1063/9.0000557 ↗
6. Optical fiber-based MR-compatible sensors for medical applications: An overview10.3390/s131014105 ↗
7. Closed-loop temperature-controlled magnetic hyperthermia therapy with magnetic guidance of superparamagnetic iron-oxide nanoparticles10.1002/adtp.202100237 ↗
8. Real-time infrared thermography detection of magnetic nanoparticle hyperthermia in a murine model under a non-uniform field configuration10.3109/02656736.2013.839056 ↗
9. Magnetic resonance thermometry and its biological applications – Physical principles and practical considerations10.1016/j.pnmrs.2019.01.003 ↗
10. Magnetic resonance thermometry: Methodology, pitfalls and practical solutions10.3109/02656736.2015.1108462 ↗
11. MR thermometry10.1002/jmri.21265 ↗
12. Magnetic nanoparticle temperature estimation10.1118/1.3106342 ↗
13. Heating efficiency in magnetic nanoparticle hyperthermia10.1016/j.jmmm.2013.11.006 ↗
14. Biocompatible nanoclusters with high heating efficiency for systemically delivered magnetic hyperthermia10.1021/acsnano.8b06542 ↗
15. Heating magnetic fluid with alternating magnetic field10.1016/s0304-8853(02)00706-0 ↗
16. Magnetic particle hyperthermia-biophysical limitations of a visionary tumour therapy10.1016/j.jmmm.2006.10.1156 ↗
17. Ultra-short pulse magnetic fields on effective magnetic hyperthermia for cancer therapy10.1063/9.0000558 ↗
18. Enhancing heating efficiency of magnetic hyperthermia using pulsed magnetic fields10.1063/9.0000702 ↗
19. Ferromagnetic resonance heating of magnetic nanoparticles resovist for biomedical applications10.1109/TMAG.2023.3285234 ↗
20. Ferucarbotran (Resovist): A new clinically approved RES-specific contrast agent for contrast-enhanced MRI of the liver: Properties, clinical development, and applications10.1007/s00330-002-1721-7 ↗
21. Mr thermometry accuracy and prospective imaging-based patient selection in MR-guided hyperthermia treatment for locally advanced cervical cancer10.3390/cancers13143503 ↗
22. Combining magnetic particle imaging and magnetic fluid hyperthermia in a theranostic platform10.1088/1361-6560/aa5601 ↗
23. 202010.1080/02656736.2020.1853252 ↗
24. Harmonic dependence of thermal magnetic particle imaging10.1038/s41598-023-42620-1 ↗
25. Simultaneous magnetic particle imaging (MPI) and temperature mapping using multi-color MPI