Abstract
The tissue equivalent proportional counter (TEPC) is often used in the measurement of ionizing radiation. However, there are many factors that can cause the micro-dose spectrum’s distortion and affect the accuracy of the measurement. In the TEPC design process, several factors can be analyzed through simulation methods. The simulation capabilities of the TEPC physical process are established and verified in this paper. The wall material, tissue equivalent gas, and electrode are selected as the main objects, and the corresponding design choices are given. The wall material has a great impact on the spectrum, and y¯F and y¯D show that A-150 plastic is closer to the human tissue in the TEPC physical process simulation. Both the two tissue equivalent gases have good simulation accuracy. After further analysis, propane-based equivalent gas is the better choice. Among the three electrode sizes, the size of 0.006 35 cm is the most suitable one for TEPC. The analysis results and analysis methods can provide references for the important elements design of TEPC.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Radiation Therapy and Dosimetry · Advanced Radiotherapy Techniques · Radiation Detection and Scintillator Technologies
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
College of Nuclear Science and Technology, Naval University of Engineering , Wuhan City 430033, China
References (14)
Show source-provided references
1. Development of micro-dose and its application
2. Evaluation of silicon based micro-dosimetry for boron neutron capture therapy quality assurance10.1016/j.ejmp.2019.09.072 ↗
3. Influence of TEPC wall on micro-dosimetric and relative biological effectiveness calculation measurement of heavy ion beam
4. Principles of experimental micro-dosimetry10.1093/oxfordjournals.rpd.a082801 ↗
5. Measurement of lineal-energy distributions for neutrons of 8 keV to 65 MeV by using a tissue-equivalent proportional counter10.1093/oxfordjournals.rpd.a006073 ↗
6. Study of neutron and γ discrimination with TEPC10.7538/yzk.2016.50.04.0577 ↗
7. Accurate simulations of TEPC neutron spectra using Geant410.1016/j.radphyschem.2015.06.015 ↗
8. Internal micro-dosimetry of alpha-emitting radionuclides10.1007/s00411-019-00826-w ↗
9. A wall-less tissue equivalent proportional counter as connecting bridge from micro-dosimetry to nano-dosimetry10.1016/j.radphyschem.2020.108729 ↗
10. Monte Carlo simulation of a new TEPC for micro-dosimetry at nanometric level: Response against a carbon ion beam10.1016/j.radmeas.2018.03.006 ↗
11. 1996
12. International Commission on Radiation Limits and Measurements, Micro-dosimetry.ICRU Report No. 36. Bethesda MD, 1983.
13. X. P. Zhao, “Tissue equivalent proportional counter applied to the micro-dose of beta nuclei and beta electrons,” Master thesis, Tsinghua University, China, 2010.
14. Numerical modelling of tissue-equivalent proportional counters10.1093/oxfordjournals.rpd.a082803 ↗