Photoacoustic spectroscopy with a widely tunable narrowband fiber-feedback optical parametric oscillator
Source abstract
Abstract
Trace gas analysis is a key tool for the investigation of man-made environmental pollution as well as for early detection of respiratory diseases. To detect tiny concentrations, sensitive methods such as cavity ring down spectroscopy or plasmonic sensors have been used. Here, we demonstrate the combination of the photoacoustic effect in a classical cell with a novel, rapidly tunable, narrowband fiber-feedback optical parametric oscillator. The high sensitivity of photoacoustic cells and the extremely narrow linewidth as well as the wide and rapid tunability of the fiber-feedback optical parametric oscillator enable a high resolution of the rotational and vibrational bands of molecules in the near-infrared region. Photoacoustic spectra of methane, carbon dioxide, and water at ambient pressure are obtained in a broad spectral range and compared to high-resolution transmission molecular absorption database. In particular, scanning the entire carbon dioxide overtone around 4965 cm−1 at 2000 ppm takes 185 s with a signal-to-noise ratio of 31. This approach enables a wide tunability in the entire near- and mid-infrared spectral region suitable for many environmental and medical applications.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Spectroscopy and Laser Applications · Photoacoustic and Ultrasonic Imaging · Spectroscopy Techniques in Biomedical and Chemical Research
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
4th Physics Institute and Research Center SCoPE, University of Stuttgart 1 , Pfaffenwaldring 57, 70569 Stuttgart,
SI Stuttgart Instruments GmbH 2 , Curiestraße 2, 70563 Stuttgart,
Funding
Carl-Zeiss-Stiftung · Deutsche Forschungsgemeinschaft · Center for Integrated Quantum Science and Technology · Baden-Württemberg Stiftung · Bundesministerium für Bildung und Forschung
References (17)
Show source-provided references
1. Nanoantenna-enhanced gas sensing in a single tailored nanofocus10.1038/nmat3029 ↗
2. Nonlinear interferometer for Fourier-transform mid-infrared gas spectroscopy using near-infrared detection10.1364/oe.415365 ↗
3. Cavity ring-down spectroscopy: Experimental schemes and applications10.1080/014423500750040627 ↗
4. LXVIII. Upon the production of sound by radiant energy10.1080/14786448108627053 ↗
5. Optical frequency comb photoacoustic spectroscopy10.1039/c8cp05666h ↗
6. Time-resolved mid-infrared dual-comb spectroscopy10.1038/s41598-019-53825-8 ↗
7. Dual-comb photoacoustic spectroscopy10.1038/s41467-020-16917-y ↗
8. Hydrogen peroxide detection with quartz-enhanced photoacoustic spectroscopy using a distributed-feedback quantum cascade laser10.1063/1.4863955 ↗
9. Air quality monitoring with photoacoustic spectroscopy10.1002/opph.201700002 ↗
10. Quartz-enhanced photoacoustic spectroscopy10.1364/ol.27.001902 ↗
11. Off-axis quartz-enhanced photoacoustic spectroscopy using a pulsed nanosecond mid-infrared optical parametric oscillator10.1364/ol.41.004118 ↗
12. Mid-infrared Fourier-transform spectroscopy with a high-brilliance tunable laser source: Investigating sample areas down to 5 μm diameter10.1364/oe.23.011105 ↗
13. Ultra-stable high average power femtosecond laser system tunable from 133 to 20 μm10.1364/ol.41.004863 ↗
14. Photoacoustic spectroscopy in trace gas monitoring10.1002/9780470027318 ↗
15. HITRAN Application Programming Interface (HAPI): A comprehensive approach to working with spectroscopic data10.1016/j.jqsrt.2016.03.005 ↗
16. Hecht · 2018 · 1710.1515/9783110526653 ↗
17. Photoacoustic spectroscopy for the determination of lung cancer biomarkers—A preliminary investigation10.3390/s17010210 ↗