Evaluation of the spatial resolution of Gafchromic™ HD-V2 radiochromic film characterized by the modulation transfer function

Tatsuhiko Miyatake, Sadaoki Kojima, Hironao Sakaki, Thanh-Hung Dinh, Ibuki Takemoto, Masayasu Hata, Masaharu Nishikino, Yukinobu Watanabe, Masahiko Ishino, Michiaki Mori, James Kevin Koga, Yoichi Yamamoto, Fuyumi Ito, Masaki Kando, Toshiyuki Shirai, Kiminori Kondo

AIP Advances· Volume 13· Issue 8

Abstract

The Modulation Transfer Function (MTF) characteristics of radiochromic film (RCF) dosimetry with optical systems were evaluated with an RCF, HD-V2, which is transferred with the line patterns of the resolution test chart. In the evaluation using a GT-X980 flatbed scanner, a high contrast spatial resolution with an MTF of ≥0.8 is limited to about 83.3 μm due to artifacts such as scattering of the HD-V2 transmitted light. To achieve high contrast at higher spatial frequencies, a test bench microdensitometer (TBMD) was developed, and its optical performance, dose equivalent response, and spatial resolution characteristics with MTF were evaluated. The TBMD had a minimum readout focal diameter of 13 μm, and its high contrast spatial resolution with an MTF of ≥0.8 is about 11.9 μm. The result that the optical density error increases with decreasing focal diameter in the TBMD supports our hypothesis that these errors are based on the non-uniformity of the shading and surface roughness of the HD-V2 active layer monomers.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Advanced Radiotherapy Techniques · Radiation Effects and Dosimetry · Radiation Detection and Scintillator Technologies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Interdisciplinary Graduate School of Engineering Sciences, Kyushu University 1 , 6-1, Kasuga-Koen, Kasuga, Fukuoka 816-8580, Japan

Kansai Photon Science Institute, National Institutes for Quantum Science and Technology 2 , 8-1-7 Umemidai, Kizugawa, Kyoto 619-0215, Japan

National Institute of Radiological Sciences, National Institutes for Quantum Science and Technology 3 , 4-9-1 Anagawa, Inage-ku, Chiba 263-8555, Japan

Funding

JST-Mirai Program · KAKENHI

References (33)

Show source-provided references
1. Measurements of energetic proton transport through magnetized plasma from intense laser interactions with solids10.1103/physrevlett.84.670 ↗
2. Forward ion acceleration in thin films driven by a high-intensity laser10.1103/physrevlett.84.4108 ↗
3. Intense high-energy proton beams from Petawatt-Laser irradiation of solids10.1103/physrevlett.85.2945 ↗
4. Particle selection for laser-accelerated proton therapy feasibility study10.1118/1.1586268 ↗
5. Toward integrated laser-driven ion accelerator systems at the photo-medical research center in Japan10.1016/j.nima.2010.01.063 ↗
6. J. Phys. Conf. Ser. · 2010 · 244 · 02205110.1088/1742-6596/244/2/022051 ↗
7. High resolution laser-driven proton radiography10.1063/1.1494128 ↗
8. Multi-MeV proton source investigations in ultraintense laser-foil interactions10.1103/PhysRevLett.92.055003 ↗
9. Proton radiography of a laser-driven implosion10.1103/PhysRevLett.97.045001 ↗
10. Applications for nuclear phenomena generated by ultra-intense lasers10.1126/science.1080552 ↗
11. Fast ignition by intense laser-accelerated proton beams10.1103/physrevlett.86.436 ↗
12. Tumour irradiation in mice with a laser-accelerated proton beam10.1038/s41567-022-01520-3 ↗
13. Laser-PIXE using laser-accelerated proton beams10.1038/s41598-019-42997-y ↗
14. Ultralow emittance, multi-MeV proton beams from a laser virtual-cathode plasma accelerator10.1103/PhysRevLett.92.204801 ↗
15. Use of GafChromic film to diagnose laser generated proton beams10.1063/1.2901603 ↗
16. Radiochromic film imaging spectroscopy of laser-accelerated proton beams10.1063/1.3086424 ↗
17. CR-39 track detector for multi-MeV ion spectroscopy10.1038/s41598-017-02331-w ↗
18. Radiochromic film dosimetry: Recommendations of AAPM radiation therapy committee task group 5510.1118/1.598407 ↗
19. A method of dosimetry for synchrotron microbeam radiation therapy using radiochromic films of different sensitivity10.1088/0031-9155/53/23/014 ↗
20. Characterization of a Gafchromic film for the two-dimensional profile measurement of low-energy heavy-ion beams10.1016/j.nima.2016.04.055 ↗
21. Absolute calibration of GafChromic film for very high flux laser driven ion beams10.1063/1.5086822 ↗
22. Detection and analysis of laser driven proton beams by calibrated Gafchromic HD-V2 and MD-V3 radiochromic films10.1063/1.5049499 ↗
23. Optimization of radiochromic film stacks to diagnose high-flux laser-accelerated proton beams10.1063/5.0020568 ↗
24. Dosimetric calibration of GafChromic HD-V2, MD-V3, and EBT3 films for dose ranges up to 100 kGy10.1063/5.0043628 ↗
25. Characterization and use of EBT radiochromic film for IMRT dose verification10.1118/1.2360012 ↗
26. A comprehensive study of the Gafchromic EBT2 radiochromic film. A comparison with EBT10.1118/1.3512792 ↗
27. Clinical use of EBT model GafchromicTM film in radiotherapy10.1118/1.2362876 ↗
28. The artefacts of radiochromic film dosimetry with flatbed scanners and their causation by light scattering from radiation-induced polymers10.1088/0031-9155/59/13/3575 ↗
29. Influence factors of resolution in laser accelerated proton radiography and image deblurring10.1063/5.0039364 ↗
30. Characterization of radiochromic films as a micrometer-resolution dosimeter by confocal Raman spectroscopy10.1002/mp.13778 ↗
31. G. Ashland (R). Dosimetry media, type hd-v2, http://www.gafchromic.com/documents/gafchromic-hdv2.pdf, 2023.
32. MTF measurement method for medical displays by using a bar-pattern image10.1889/1.2372416 ↗
33. Radiochromic film for medical radiation dosimetry10.1016/s0927-796x(03)00034-2 ↗