journal article·2018-10-01Open access

Development of an CNT field emission UHV ionization gauge with customized electronics

Jian Sun, Detian Li, Yongjun Cheng, Huzhong Zhang, Yongjun Wang, Gang Li, Xiaodong Han

AIP Advances· Volume 8· Issue 10

Abstract

An ionization vacuum gauge with carbon nanotube (CNT) cathode was designed for measurement of ultra high vacuum (UHV). The sensor was constructed based on Bayard-Alpert type gauge. The key index of the control unit is determined by theoretical analysis and experimental results. A customized detection circuit was designed considering the characteristics of the output ion current of the sensor, the performance of which, such as linearity, noise, bandwidth and gain, were determined by a series of tests and calibrations. At last, the vacuum measurement range and uncertainty of this gauge were determined. The test results showed that the lower limit of the CNT-cathode ionization vacuum gauge enters 10-8Pa, The measurement uncertainty is about 4.0% in the pressure range of 10-4Pa-10-7Pa. This work can provide a new solution for UHV measurement.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Mechanical and Optical Resonators · Carbon Nanotubes in Composites · Advanced MEMS and NEMS Technologies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Science and Technology on Vacuum Technology and Physics Laboratory, Lanzhou Institute of Physics , Lanzhou 730000, China

Funding

National Natural Science Foundation of China

References (10)

Show source-provided references
1. Latest progress in measurement technology of ultra high and extremely high vacuum
2. An overview of ionization gauges with carbon nanotube cathodes10.1088/0022-3727/48/47/473001 ↗
3. Measurement of vacuum:1950-200310.1116/1.1599870 ↗
4. Extension of the low pressure range of the ionization gauge10.1063/1.1745653 ↗
5. Design of micro current detection circuit for FAIMS10.3724/sp.j.1187.2011.00711 ↗
6. Weak current signal detection and reception of quasi-electrostatic field
7. 2003
8. Methods of improving performance of weak current integrated amplifier
9. An accurate low current measurement circuit for heavy iron beam current monitor10.1016/j.nimb.2012.01.033 ↗
10. Vacuum-calibration apparatus with pressure down to 10−10 Pa10.1116/1.3457934 ↗