journal article·2023-03-01Open access

Sedimentation of magnetorheological fluids measured using an automated vertical axis inductance monitoring system

Ran Ma, Norman M. Wereley

AIP Advances· Volume 13· Issue 3

Abstract

Stable magnetorheological fluids (MRFs) are inevitably required to develop reliable MRF-based applications. Suspension stability of MRFs can be studied using an automated vertical axis inductance monitoring system (AVAIMS), where an inductance sensor is translated vertically along a column of MRF to measure particle concentration as a function of column height and time. The AVAIMS allows for tracking the particle concentration discontinuities due to the sedimentation of MRFs including the mud-line, the gel-line, and the cake-line. The sedimentation progress of MRFs can be evaluated by analyzing these boundaries. In this study, the AVAIMS design was improved to better identify the sedimentation zone boundaries by using an optimized inductance sensor that measures the concentration gradient as a function of time. The inductance sensor design, material selection, and geometry were optimized to improve the sensor sensitivity. The calibration curve of the sensor was obtained by measuring samples with different particle volume fractions. The sensor design was validated by examining an MRF sample with a particle concentration of 20% volume fraction for seven consecutive days. These tests show the effectiveness of the AVAIMS optimized to better localize the sedimentation zone boundaries using the concentration gradient profile method.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Vibration Control and Rheological Fluids · Characterization and Applications of Magnetic Nanoparticles · Structural Health Monitoring Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

School of Mechanical and Electrical Engineering, Jiangsu Normal University 1 , Xuzhou 221116, Jiang Su, People’s Republic of China

Department of Aerospace Engineering, University of Maryland 2 , College Park, Maryland 20742, USA

References (26)

Show source-provided references
1. New control logic based on mechanical energy conservation for aircraft landing gear system with magnetorheological dampers10.1088/1361-665x/ab9e11 ↗
2. A novel passive adaptive magnetorheological energy absorber10.1088/1361-665x/abc149 ↗
3. Effects of magnetic core parameters on landing stability and efficiency of magnetorheological damper-based landing gear system10.1177/1045389x19862639 ↗
4. A wearable elbow exoskeleton for tremor suppression equipped with rotational semi-active actuator10.1016/j.ymssp.2021.107674 ↗
5. A magnetorheological fluid-filled soft crawling robot with magnetic actuation10.1109/tmech.2020.2988049 ↗
6. Magnetorheological energy absorber with dual concentric annular valves10.1177/1045389x15577659 ↗
7. Experimental study on shock control of a vehicle semi-active suspension with magneto-rheological damper10.1088/1361-665x/ab859e ↗
8. A novel semi-active control strategy based on the cascade quantitative feedback theory for a vehicle suspension system10.1177/0954407018790608 ↗
9. A review on the magnetorheological fluid preparation and stabilization10.1016/j.jmmm.2014.09.020 ↗
10. A review of challenges and solutions in the preparation and use of magnetorheological fluids10.1186/s40712-019-0109-2 ↗
11. Magnetorheological fluid based on thixotropic PTFE-oil organogel10.1016/j.jmmm.2017.11.005 ↗
12. Viscometric and sedimentation characterization of bidisperse magnetorheological fluids10.1109/tmag.2007.893867 ↗
13. Testing and analysis of magnetorheological fluid sedimentation in a column using a vertical axis inductance monitoring system10.1088/0964-1726/25/4/04lt01 ↗
14. Magnetorheological fluids based on ionic liquids10.1002/adma.200700302 ↗
15. Preparation and characterization of carbonyl iron/strontium hexaferrite magnetorheological fluids10.1016/j.partic.2014.04.021 ↗
16. Study of sedimentation stability of magnetorheological fluid10.11648/j.am.20150401.11 ↗
17. Piezo-capacitive behavior of a magnetically structured particle-based conductive polymer with high sensitivity and a wide working range10.1039/c8tc00302e ↗
18. In-situ capacitance sensing for the settlement of magnetorheological fluid: simulation and experiments10.3389/fmats.2021.663925 ↗
19. The sedimentation rate measurement system with thermal probe and its application
20. Effect of compound surfactants modified carbonyl iron on magnetorheological fluids10.1007/s10948-021-05813-1 ↗
21. Permalloy inductor based instrument that measures the sedimentation constant of magnetorheological fluids10.1063/1.1581356 ↗
22. Description and performance of a fully automatic device for the study of the sedimentation of magnetic suspensions10.1063/1.3609228 ↗
23. Vertical axis inductance monitoring system to measure stratification in a column of magnetorheological fluid10.1109/tmag.2016.2606345 ↗
24. Monitoring sedimentation of magnetorheological fluids using a vertical axis monitoring system with a low aspect ratio sensor coil10.1088/1361-665x/aaf757 ↗
25. Magnetorheological fluids