Hybrid analytical method for calibrating a standard NaI(Tl) gamma-ray scintillation detector using a lateral hexagonal radioactive source
Source abstract
Abstract
In various fields of radiation and nuclear physics, research and its applications help in the management and control of human activities, especially in the analysis of nuclear waste and measurement of low activity environmental samples. The current study uses a hybrid analytical technique and a 152Eu lateral hexagonal radioactive source to calibrate a standard cylindrical NaI(Tl) γ-ray scintillation detector. The technique is a combination of experimental measurements and theoretical calculations because for each environmental and/or radioactive waste sample, a separate reference calibration source equivalent to samples of different compositions is required, which needs additional funding. This method is suitable for any source–detector geometry. It can be used to calculate the geometric, total, and full-energy peak efficiency and the P/T ratio using the model probability of detecting γ-radiation photons, in particular from a lateral hexagonal radioactive source. The detection probability can be mainly influenced by the direction of the γ-quantum motion. This can help build more realistic and complex models than the typical point source case, where simple integration is used to numerically calculate the efficiency of the detection system. Using this method, the self-attenuation and coincidence-summing effect can be calculated for any gamma source matrices. The results obtained can show how useful this technique can be for calibrating complex gamma-ray detectors with any type of gamma-ray source.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Radioactivity and Radon Measurements · Radiation Detection and Scintillator Technologies · Nuclear Physics and Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Physics Department, Faculty of Science, Beirut Arab University 1 , Beirut, Lebanon
Physics Department, Faculty of Science, Alexandria University 2 , 21511 Alexandria, Egypt
Physics Department, Al-Jamoum University College, Umm Al-Qura University 3 , Al-Jamoum, Saudi Arabia
Department of Biomedical Equipment Technology, Faculty of Applied Health Sciences Technology, Pharos University in Alexandria 4 , Alexandria, Egypt
References (22)
Show source-provided references
1. 2010
2. A method to calculate the gamma ray detection efficiency of a cylindrical NaI(Tl) crystal10.1088/1748-0221/13/05/p05019 ↗
3. Standardization of 67Ga by 4πγ (NaI) and 4πβ–γ coincidence methods10.1016/j.apradiso.2003.11.041 ↗
4. Calibration of a single hexagonal NaI(Tl) detector using a new numerical method based on the efficiency transfer method10.1016/j.nima.2014.10.056 ↗
5. The determination of 125I activity using sum-peak method with a well-type HPGe-detector-based spectrometer10.1016/s0168-9002(00)01017-2 ↗
6. Validation of an efficiency calibration procedure for a coaxial n-type and a well-type HPGe detector used for the measurement of environmental radioactivity10.1016/j.nima.2016.02.039 ↗
7. Semiempirical efficiency calibration in semiconductor HPGe gamma-ray spectroscopy10.1155/2018/5392658 ↗
8. Efficiency calibration of a HPGe detector in the [46.54–2000] keV energy range for the measurement of environmental samples10.1016/s0168-9002(01)00798-7 ↗
9. 1995
10. Testing efficiency transfer codes for equivalence10.1016/j.apradiso.2009.10.012 ↗
11. On the Monte Carlo simulation of HPGe gamma-spectrometry systems10.1016/j.apradiso.2009.01.014 ↗
12. Geant4—A simulation toolkit10.1016/s0168-9002(03)01368-8 ↗
13. Calculation of HPGe efficiency for environmental samples: Comparison of EFFTRAN and GEANT410.1016/j.nima.2014.06.044 ↗
14. Full-energy peak efficiency of an NaI(Tl) detector with coincidence summing correction showing the effect of the source-to-detector distance10.1016/j.cjph.2016.11.013 ↗
15. Experimental verification of gamma-efficiency calculations for scintillation detectors in ANGLE 4 software10.2298/ntrp1501035t ↗
16. Characterization of the efficiency of a cubic NaI detector with rectangular cavity for axially positioned sources10.1088/1748-0221/15/02/p02013 ↗
17. Calibration of 4π NaI(Tl) detectors with coincidence summing correction using new numerical procedure and ANGLE4 software10.1063/1.4978214 ↗
18. Study on the effect of the self-attenuation coefficient on γ-ray detector efficiency calculated at low and high energy regions10.5516/net.04.2013.077 ↗
19. Mathematical method to calculate full-energy peak efficiency of detectors based on transfer technique10.1007/s12648-015-0737-1 ↗
20. Well-type NaI(Tl) detector efficiency using analytical technique and ANGLE 4 software based on radioactive point sources located out the well cavity10.1016/j.cjph.2016.03.019 ↗
21. Measurements correlated to efficiency studies of NaI(Tl) scintillator with vertical hexagonal radioactive source10.1088/1748-0221/16/07/p07011 ↗
22. Calibration of cylindrical NaI(Tl) gamma-ray detector intended for truncated conical radioactive source10.1016/j.net.2021.10.016 ↗