Signal transmission analysis in implantable human body communication for abdominal medical devices
Source abstract
Abstract
Implantable medical devices, such as neurostimulators, need to be wirelessly controlled from outside the body. Many of these devices use high-frequency signals in the 400 MHz, 900 MHz, and 2.45 GHz bands to communicate with external devices. However, high-frequency bands can suffer from signal attenuation in biological tissues and from electromagnetic interference with surrounding devices. In contrast, human body communication (HBC) uses relatively low frequencies in the 3–30 MHz band, which can alleviate signal attenuation and prevent emissions outside the body during communication. In this study, we investigated the use of HBC for implantable medical devices in the abdomen using electromagnetic field simulations. The results showed that the transmission between the transmitter inside the body and the receiver outside the body was adequate for stable communication. However, when the receiver was detached from the skin surface, the transmission characteristics rapidly decreased. Moreover, when the receiver and skin were separated by 1 mm, the transmission degraded by 34 dB, indicating that electromagnetic interference between the implanted transmitter and surrounding devices (including other medical devices) is rare. Finally, we evaluated the proposed system from the perspective of medical electromagnetic compatibility and human safety. The simulation results demonstrated that the radiated emissions and human exposure of the HBC system meet international standards.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Wireless Body Area Networks · Molecular Communication and Nanonetworks · Neuroscience and Neural Engineering
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
The University of Electro-Communications , 1-5-1 Chofugaoka, Chofu, Tokyo 182-8585, Japan
Funding
Japan Society for the Promotion of Science · Takahashi Industrial and Economic Research Foundation · KDDI Foundation
References (47)
Show source-provided references
1. Recent advances in targeting ion channels to treat chronic pain: Editorial10.1111/bph.14215 ↗
2. Chronic pain as a symptom or a disease: The IASP classification of chronic pain for the international classification of diseases (ICD-11)10.1097/j.pain.0000000000001384 ↗
3. Spinal cord stimulation in chronic pain: Evidence and theory for mechanisms of action10.1186/s42234-019-0023-1 ↗
4. Spinal cord stimulation in chronic neuropathic pain: Mechanisms of action, new locations, new paradigms10.1097/j.pain.0000000000001854 ↗
5. Medtronic B. V. · 2016
6. Confirm Rx insertable cardiac monitor for primary atrial fibrillation detection in high-risk heart failure patients (Confirm-AF trial)10.1111/anec.13021 ↗
7. Wireless powering and telemetry of deep-body ingestible bioelectronic capsule10.1109/tap.2022.3184549 ↗
8. Compact triple-band implantable antenna for multitasking medical devices10.2478/jee-2022-0022 ↗
9. EM interaction of handset antennas and a human in personal communications10.1109/5.362755 ↗
10. 2005 · 7754
11. Modeling and characterization of biotelemetric radio channel from ingested implants considering organ contents10.1109/tap.2009.2014531 ↗
12. The attenuation characteristics of the body tissue on frequency function in WBAN channel10.12962/jaree.v5i2.185 ↗
13. Propagation-loss characterization for livestock implantables at (433, 868, 1400) MHz10.1109/tap.2021.3060501 ↗
14. Bluetooth signal attenuation analysis in human body tissue analogues10.1109/access.2021.3087780 ↗
15. American national standard recommended practice for an on-site, ad hoc test method for estimating electromagnetic immunity of medical devices to radiated radio-frequency (RF) emissions from RF transmitters
16. Testing the immunity of active implantable medical devices to CW magnetic fields up to 1 MHz by an immersion method10.1109/tbme.2007.893502 ↗
17. 2011 · 101
18. 2018 · 1
19. Emerging electromagnetic interferences between implantable cardioverter-defibrillators and left ventricular assist devices10.1093/europace/euaa006 ↗
20. 2022 · 124
21. Personal area networks: Near-field intrabody communication10.1147/sj.353.0609 ↗
22. Past results, present trends, and future challenges in intrabody communication10.1155/2018/9026847 ↗
23. Enabling covert body area network using electro-quasistatic human body communication10.1038/s41598-018-38303-x ↗
24. Wide band human body communication technology for wearable and implantable robot control10.1587/transcom.2019hmi0001 ↗
25. Intrabody communications between mobile device and wearable device at 26 MHz10.1109/lawp.2016.2588524 ↗
26. Design of cavity-backed bow-tie antenna with matching layer for human body application10.3390/s19184015 ↗
27. Transmission analysis in human body communication for head-mounted wearable devices10.3390/electronics10101213 ↗
28. 2022 · 1
29. Review of medical implant communication system (MICS) band and network10.1016/j.icte.2016.08.010 ↗
30. 2021 · 1
31. K. Makiko and M.Masaaki, Japanese 3-D body shape and dimensions data 2003, H18PRO-503, National Institute of Advanced Industrial Science and Technology, 2006.
32. Multilayered phantom for input impedance evaluation of human body communication electrodes10.4108/eai.28-9-2015.2261486 ↗
33. The thickness of the skin10.1097/00006534-195505000-00006 ↗
34. 2017 · 1
35. Comparison of passive 2-D and 3-D ring arrays for medical telemetry focusing10.1109/lawp.2019.2911928 ↗
36. St. Jude Medical, Proclaim implantable pulse generator clinician’s manual, 2017.
37. Electric field distributions of wearable devices using the human body as a transmission channel10.1109/tap.2007.900226 ↗
38. Analytical and experimental studies on human body communication between wristwatch and handheld devices using muscle homogenous phantom at 10 MHz10.18494/SAM.2014.1035 ↗
39. Equivalent circuit model viewed from receiver side in human body communication10.1109/tbcas.2019.2918323 ↗
40. Radio regulations, edition of 2020: Vol. 1: Articles, International Telecommunication Union, 2020.
41. 2019 · 5200
42. Transcutaneous optical telemetry system for a totally implantable artificial heart investigation of transcutaneous optical coupler consisting of three light emitting and receiving devices10.11392/jsao1972.29.315 ↗
43. 2021
44. 2019
45. R. F. Cleveland and J. L.Ulcek, Questions and answers about biological effects and potential hazards of radiofrequency electromagnetic fields (Federal Communications Commission, Washington, DC, USA, 1999).
46. IEEE standard for safety levels with respect to human exposure to radio frequency electromagnetic fields, 3 kHz to 300 GHz (IEEE C95.1-2005), IEEE, 2006.
47. 2018 · 417