A method for estimating the contact area of a dual-mobility total hip prosthesis

Taufiq Hidayat, Muhammad Imam Ammarullah, Eko Saputra, M. Danny Pratama Lamura, Chethan K N, Rifky Ismail, Athanasius Priharyoto Bayuseno, J. Jamari

AIP Advances· Volume 14· Issue 1

Abstract

The term “contact area” refers to the total surface area of two entities in direct physical touch. When discussing an artificial hip joint, “contact area” refers to the surface area of contact between the components of the artificial hip joint (ball and cup) positioned inside the patient. Several methods can be used to figure out the contact area of an artificial hip joint, such as finite element analysis and traditional experiments on contact mechanics with hip joint simulators. The contact area in an artificial hip joint ensures load distribution. For optimal and long-term performance, the prosthetic hip joint’s contact area must be well understood for design, fitting, and monitoring. This study presented a novel method to estimate the liner surface contact area due to interaction contact in the artificial hip joint using a computer-aided design (CAD) program. This study also contrasted numerical approaches utilizing computer-aided engineering software and theoretical predictions based on Hertz theory with three-dimensional processes using CAD software to determine the contact area in the inner liner. There were no significant discrepancies in the outcomes of the three approaches.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Orthopedic Infections and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Mechanical Engineering, Faculty of Engineering, Universitas Muria Kudus 1 , Kudus 59324, West Java, Indonesia

Department of Mechanical Engineering, Faculty of Engineering, Universitas Diponegoro 2 , Semarang 50275, Central Java, Indonesia

Department of Mechanics and Aerospace Engineering, College of Engineering, Southern University of Science and Technology 3 , Shenzhen 518055, Guangdong, China

Undip Biomechanics Engineering and Research Centre (UBM-ERC), Universitas Diponegoro 4 , Semarang 50275, Central Java, Indonesia

Department of Mechanical Engineering, Politeknik Negeri Semarang 5 , Semarang 50275, Central Java, Indonesia

Department of Aeronautical and Automobile Engineering, Manipal Institute of Technology, Manipal Academy of Higher Education 6 , Manipal 576104, Karnataka, India

Center for Biomechanics Biomaterials Biomechatronics and Biosignal Processing (CBIOM3S), Universitas Diponegoro 7 , Semarang 50275, Central Java, Indonesia

Funding

World Class Research Universitas Diponegoro · Penelitian Fundamental - Reguler

References (58)

Show source-provided references
1. Evolution and new generation of dual mobility cups10.3928/01477447-20230804-01 ↗
2. The evolution of outcomes and indications for the dual-mobility cup: A systematic review10.1007/s00264-016-3377-y ↗
3. Computational investigation of various stem designs with different radial clearances in total hip arthroplasty10.1080/23311916.2023.2287296 ↗
4. Dual mobility in total hip arthroplasty: Biomechanics, indications and complications-current concepts10.1007/s43465-021-00471-w ↗
5. The performance of primary dual-mobility total hip arthroplasty in patients aged 55 years and younger: A systematic review10.1016/j.artd.2023.101241 ↗
6. The efficacy of dual-mobility cup in preventing dislocation after total hip arthroplasty: A systematic review and meta-analysis of comparative studies10.1007/s00264-018-4062-0 ↗
7. Comparison of dual-mobility cup and unipolar cup for prevention of dislocation after revision total hip arthroplasty10.1080/17453674.2016.1255482 ↗
8. Total hip arthroplasty using a cementless dual-mobility cup provides increased stability and favorable gait parameters at five years follow-up10.1016/j.otsr.2016.09.020 ↗
9. Wear estimation at the contact surfaces of oval shaped hip implants using finite element analysis10.1080/23311916.2023.2222985 ↗
10. Mechanical investigation for the use of polylactic acid in total hip arthroplasty using FEM analysis
11. Optimization of hip implant designs based on its mechanical behaviour
12. Static structural analysis of the effect of change in femoral head sizes used in Total Hip Arthroplasty using finite element method10.1080/23311916.2022.2027080 ↗
13. Evolution of different designs and wear studies in total hip prosthesis using finite element analysis: A review10.1080/23311916.2022.2027081 ↗
14. How to measure the real contact area? A simple marker and relocation foot-printing approach10.1016/j.triboint.2016.06.034 ↗
15. A novel test rig for in situ and real time optical measurement of the contact area evolution during pre-sliding of a spherical contact10.1007/s11249-006-9113-9 ↗
16. A new method for the measurement of real area of contact by the adhesive transfer of thin Au film10.1007/s11249-018-0982-5 ↗
17. A mathematical analysis for indentation tests of articular cartilage10.1016/0021-9290(72)90010-3 ↗
18. Axi-symmetric contact on thin compliant coatings10.1016/0022-5096(81)90018-1 ↗
19. Short communication: Running-in behavior on single-mobility total hip arthroplasty10.1016/j.medengphy.2022.103806 ↗
20. Static structural analysis of different stem designs used in total hip arthroplasty using finite element method10.1016/j.heliyon.2019.e01767 ↗
21. Fatigue life evaluation of different hip implant designs using finite element analysis10.5937/jaes0-44094 ↗
22. Three-dimensional finite analysis of acetabular contact pressure and contact area during normal walking10.1016/j.asjsur.2016.07.002 ↗
23. Optimization of the cycle time of robotics resistance spot welding for automotive applications10.1002/amp2.10084 ↗
24. Analysis of physical, mechanical and tribological behavior of Al7075-fly ash composite for lightweight applications10.1007/s12008-023-01583-3 ↗
25. HA-based coating by plasma spray techniques on titanium alloy for orthopedic applications10.1016/j.matpr.2021.03.165 ↗
26. Multiple-criteria decision-making and sensitivity analysis for selection of materials for knee implant femoral component10.3390/ma14082084 ↗
27. Environmental, economical and technological analysis of MQL-assisted machining of Al–Mg–Zr alloy using PCD tool10.3390/su13137321 ↗
28. Bibliometric analysis of entropy weights method for multi-objective optimization in machining operations10.1016/j.matpr.2021.08.132 ↗
29. Exploring the application sphere of electrical discharge machining in composite materials considering surface features: A content analysis10.1007/s12008-022-01060-3 ↗
30. Exploring the application sphere of the internet of things in industry 4.0: A review, bibliometric and content analysis10.3390/s22114276 ↗
31. Smart watches: A review of evolution in bio-medical sector10.1016/j.matpr.2021.07.460 ↗
32. An outline on modern day applications of solid lubricants10.1016/j.matpr.2020.05.558 ↗
33. Effect of cup abduction angle and head lateral microseparation on contact stresses in ceramic-on-ceramic total hip arthroplasty10.1016/j.jbiomech.2011.10.033 ↗
34. Relationship between deformation and running-in wear on hard-on-hard bearings from metal, ceramic, and diamond materials for total hip prosthesis
35. Optimized trapezoidal-shaped hip implant for total hip arthroplasty using finite element analysis10.1080/23311916.2020.1719575 ↗
36. Finite element analysis of elliptical shaped stem profile of hip prosthesis using dynamic loading conditions10.1088/2057-1976/acfe14 ↗
37. The transmission of load through the human hip joint10.1016/0021-9290(71)90041-8 ↗
38. Three-dimensional dynamic hip contact area and pressure distribution during activities of daily living10.1016/j.jbiomech.2005.06.026 ↗
39. Advanced biomaterials in hip joint arthroplasty. A review on polymer and ceramics composites as alternative bearings10.1016/j.compositesb.2015.07.019 ↗
40. Contact mechanics of modular metal-on-polyethylene total hip replacement under adverse edge loading conditions10.1016/j.jbiomech.2014.08.015 ↗
41. Materials for hip prostheses: A review of wear and loading considerations10.3390/ma12030495 ↗
42. An elasto-plastic finite element model for polyethylene wear in total hip arthroplasty10.1016/s0021-9290(01)00215-9 ↗
43. 3—Explicit finite element modelling of artificial hip and knee joints
44. Contact of dual mobility implants: Effects of cup wear and inclination10.1080/10255842.2014.936856 ↗
45. Predicting the wear of soft-on-hard bearing couples for human hip prosthesis using finite element concepts10.1142/s0219519416500202 ↗
46. Hip contact forces and gait patterns from routine activities10.1016/s0021-9290(01)00040-9 ↗
47. Standardized loads acting in hip implants10.1371/journal.pone.0155612 ↗
48. Determination of contact area in ‘cushion form’ bearings for artificial hip joints10.1243/pime_proc_1990_204_259_02 ↗
49. 2014
50. Contact mechanics and lubrication analyses of ceramic-on-metal total hip replacements10.1016/j.triboint.2012.02.012 ↗
51. Wear versus thickness and other features of 5-mrad crosslinked UHMWPE acetabular liners10.1007/s11999-010-1555-6 ↗
52. Experimental prediction of contact area in hip replacement and hemi-arthroplasty
53. The human hip stress analysis: A ball-socket elastic contact modelling10.4015/s1016237204000311 ↗
54. UHMWPE wear rate as a function of contact area and stress
55. A new formulation for the prediction of polyethylene wear in artificial hip joints10.1243/09544119jeim819 ↗
56. Effect of contact area on the wear and friction of UHMWPE in circular translation pin-on-disk tests10.1115/1.4036448 ↗
57. Effects of contact area and stress on the volumetric wear of ultrahigh molecular weight polyethylene10.1016/s0043-1648(03)00135-2 ↗
58. An analytical model for almost conformal spherical contact problems: Application to total hip arthroplasty with UHMWPE liner10.3390/app112311170 ↗