Demonstration of a scalable frequency-domain readout of metallic magnetic calorimeters by means of a microwave SQUID multiplexer
Source abstract
Abstract
We report on the first demonstration of a scalable GHz frequency-domain readout of metallic magnetic calorimeters (MMCs) using a 64 pixel detector array that is read out by an integrated, on-chip microwave SQUID multiplexer. The detector array is optimized for detecting soft X-ray photons and the multiplexer is designed to provide a signal rise time τrise<400ns and an intrinsic energy sensitivity ϵ<30h. This results in an expected energy resolution ΔEFWHM<10eV. We measured a signal rise time τrise as low as 90ns and an energy resolution ΔEFWHM as low as 50eV for 5.9keV photons. The rise time is about an order of magnitude faster compared to other multiplexed low-temperature microcalorimeters and close to the intrinsic value set by the coupling between electron and spins. The energy resolution is degraded with respect to our design value due to a rather low intrinsic quality factor of the microwave resonators that is caused by the quality of the Josephson junction of the associated rf-SQUID as well as an elevated chip temperature as compared to the heat bath. Though the achieved energy resolution is not yet compatible with state-of-the-art single-channel MMCs, this demonstration of a scalable readout approach for MMCs in combination with the full understanding of the device performance showing ways how to improve represents an important milestone for the development of future large-scale MMC detector arrays.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Superconducting and THz Device Technology · Physics of Superconductivity and Magnetism · Terahertz technology and applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Heidelberg University Kirchhoff-Institute for Physics, , Im Neuenheimer Feld 227, 69120 Heidelberg, Germany
Funding
Deutsche Forschungsgemeinschaft
References (38)
Show source-provided references
1. Metallic magnetic calorimeters10.1007/b12169 ↗
2. Magnetically coupled microcalorimeters10.1007/s10909-012-0544-4 ↗
3. maxs: Microcalorimeter arrays for high-resolution x-ray spectroscopy at gsi/fair10.1007/s10909-012-0557-z ↗
4. Towards fair: First measurements of metallic magnetic calorimeters for high-resolution x-ray spectroscopy at gsi10.1088/0031-8949/2015/T166/014054 ↗
5. Development of mmc gamma detectors for nuclear analysis10.1007/s10909-013-1063-7 ↗
6. Reproducibility and calibration of mmc-based high-resolution gamma detectors10.1063/1.4958699 ↗
7. Metallic magnetic calorimeters for absolute activity measurement10.1007/s10909-008-9793-7 ↗
8. Beta spectrometry with magnetic calorimeters10.1007/s10909-008-9787-5 ↗
9. First measurement of the beta spectrum of 241pu with a cryogenic detector10.1016/j.apradiso.2009.11.054 ↗
10. Development of cryogenic alpha spectrometers using metallic magnetic calorimeters10.1016/j.nima.2010.11.066 ↗
11. Prospects for measuring the 229th isomer energy using a metallic magnetic microcalorimeter10.1016/j.nima.2013.09.012 ↗
12. Evidence for the exchange effect down to very low energy in the beta decays of 63ni and 241pu10.1007/s10909-014-1190-9 ↗
13. The electron capture 163ho experiment echo10.1007/s10909-014-1187-4 ↗
14. Concept of metallic magnetic calorimeters for rare event search in the lumineu project10.1007/s10909-013-1023-2 ↗
15. The first tests of a large-area light detector equipped with metallic magnetic calorimeters for scintillating bolometers for the lumineu neutrinoless double beta decay search10.1007/s10909-016-1535-7 ↗
16. Cryogenic micro-calorimeters for mass spectrometric identification of neutral molecules and molecular fragments10.1063/1.4930036 ↗
17. 2016
18. Squid electronics10.1002/3527603646 ↗
19. Time domain multiplexed readout of magnetically coupled calorimeters10.1109/tasc.2013.2243792 ↗
20. Time-division squid multiplexers10.1063/1.1457650 ↗
21. Squid multiplexers for transition-edge sensors10.1016/s0921-4534(01)01167-4 ↗
22. Microwave squid multiplexer10.1063/1.1791733 ↗
23. Microwave squid multiplexers for low-temperature detectors10.1016/j.nima.2005.12.144 ↗
24. Demonstration of a multiplexer of dissipationless superconducting quantum interference devices10.1063/1.2803852 ↗
25. Shannon limits for low-temperature detector readout10.1063/1.3292320 ↗
26. Superconducting single-junction interferometers with small critical currents10.1063/1.1662916 ↗
27. rf squid’s in the nonhysteretic mode: Detailed comparison of theory and experiment10.1063/1.322984 ↗
28. Microwave kinetic inductance detectors: The first decade10.1063/1.3292300 ↗
29. Superconducting microresonators: Physics and applications10.1146/annurev-conmatphys-020911-125022 ↗
30. Three-dimensional multi-terminal superconductive integrated circuit inductance extraction10.1088/0953-2048/24/12/125015 ↗
31. Microwave squid multiplexer for the readout of metallic magnetic calorimeters10.1007/s10909-014-1153-1 ↗
32. 2016
33. Characterization of the reliability and uniformity of an anodization-free fabrication process for high-quality nb/al-alox/nb josephson junctions10.1088/0953-2048/26/6/065012 ↗
34. Evaluation of a microwave squid multiplexer prototype10.1109/tasc.2007.898118 ↗
35. Flux-ramp modulation for squid multiplexing10.1007/s10909-012-0518-6 ↗
36. Noise properties of superconducting coplanar waveguide microwave resonators10.1063/1.2711770 ↗
37. Contribution of dielectrics to frequency and noise of nbtin superconducting resonators10.1063/1.2937837 ↗
38. Experimental evidence for a surface distribution of two-level systems in superconducting lithographed microwave resonators10.1063/1.2906373 ↗