journal article·2014-12-01Open access

An unshielded radio-frequency atomic magnetometer with sub-femtoTesla sensitivity

David A. Keder, David W. Prescott, Adam W. Conovaloff, Karen L. Sauer

AIP Advances· Volume 4· Issue 12

Abstract

We demonstrate a radio-frequency potassium-vapor magnetometer operating with sensitivities of 0.3 fT/Hz at 0.5 MHz and 0.9 fT/Hz at 1.31 MHz in the absence of radio-frequency and mu-metal or magnetic shielding. The use of spatially separated magnetometers, two voxels within the same cell, permits for the subtraction of common mode noise and the retention of a gradient signal, as from a local source. At 0.5 MHz the common mode noise was white and measured to be 3.4 fT/Hz; upon subtraction the noise returned to the values observed when the magnetometer was shielded. At 1.31 MHz, the common mode noise was from a nearby radio station and was reduced by a factor of 33 upon subtraction, limited only by the radio signal picked up by receiver electronics. Potential applications include in-the-field low-field magnetic resonance, such as the use of nuclear quadrupole resonance for the detection of explosives.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Atomic and Subatomic Physics Research · Advanced MRI Techniques and Applications · Quantum optics and atomic interactions

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

School of Physics, Astronomy, and Computational Sciences, George Mason University , Fairfax, Virginia, 22030, USA

Funding

National Science Foundation · Defense Advanced Research Projects Agency

References (27)

Show source-provided references
1. Nat. Phys. · 2007 · 3 · 22710.1038/nphys566 ↗
2. Magnetics, IEEE Transactions on · 1979 · 15 · 47810.1109/TMAG.1979.1060155 ↗
3. Review of Scientific Instruments · 1996 · 6710.1063/1.1147122 ↗
4. Applied Superconductivity, IEEE Transactions on · 1997 · 7 · 321310.1109/77.622015 ↗
5. Superconductor Science and Technology · 2008 · 21 · 01502310.1088/0953-2048/21/01/015023 ↗
6. Journal of Magnetic Resonance · 2007 · 185 · 21410.1016/j.jmr.2006.12.012 ↗
7. Journal of Magnetic Resonance · 2009 · 201 · 22210.1016/j.jmr.2009.09.013 ↗
8. Proceedings of the National Academy of Sciences · 2006 · 103 · 1266810.1073/pnas.0605396103 ↗
9. Applied Physics Letters · 2013 · 103 · 04370310.1063/1.4816433 ↗
10. Applied Physics Letters · 2006 · 89 · 21110410.1063/1.2392722 ↗
11. J. Opt. Soc. Am. B · 2007 · 24 · 235710.1364/JOSAB.24.002357 ↗
12. NeuroImage · 2014 · 89 · 14310.1016/j.neuroimage.2013.10.040 ↗
13. Applied Physics Letters · 2004 · 85 · 640910.1063/1.1839274 ↗
14. Applied Physics Letters · 2012 · 100 · 24240110.1063/1.4729016 ↗
15. Phys. Rev. Lett. · 2005 · 95 · 06300410.1103/PhysRevLett.95.063004 ↗
16. Journal of Magnetic Resonance · 2009 · 199 · 18810.1016/j.jmr.2009.04.012 ↗
17. Applied Physics Letters · 2006 · 89 · 21410610.1063/1.2390643 ↗
18. Geoscience and Remote Sensing, IEEE Transactions on · 2001 · 39 · 110810.1109/36.927420 ↗
19. American Scientist · 2005 · 93 · pp. 5010.1511/2005.51.953 ↗
20. Applied Physics B · 2002 · 75 · 60510.1007/s00340-002-0959-8 ↗
21. Nature · 2003 · 422 · 59610.1038/nature01484 ↗
22. Applied Physics Letters · 2004 · 8510.1063/1.1814434 ↗
23. Phys. Rev. A · 2013 · 87 · 01341310.1103/PhysRevA.87.013413 ↗
24. Phys. Rev. Lett. · 2013 · 110 · 16080210.1103/PhysRevLett.110.160802 ↗
25. 2013
26. Radio Science Bulletin · 2010 · 49
27. Phys. Rev. Lett. · 2010 · 104 · 01360110.1103/PhysRevLett.104.013601 ↗