Calibration of 4π NaI(Tl) detectors with coincidence summing correction using new numerical procedure and ANGLE4 software

Mohamed S. Badawi, Slobodan I. Jovanovic, Abouzeid A. Thabet, Ahmed M. El-Khatib, Aleksandar D. Dlabac, Bohaysa A. Salem, Mona M. Gouda, Nikola N. Mihaljevic, Kholud S. Almugren, Mahmoud I. Abbas

AIP Advances· Volume 7· Issue 3

Abstract

The 4π NaI(Tl) γ-ray detectors are consisted of the well cavity with cylindrical cross section, and the enclosing geometry of measurements with large detection angle. This leads to exceptionally high efficiency level and a significant coincidence summing effect, much more than a single cylindrical or coaxial detector especially in very low activity measurements. In the present work, the detection effective solid angle in addition to both full-energy peak and total efficiencies of well-type detectors, were mainly calculated by the new numerical simulation method (NSM) and ANGLE4 software. To obtain the coincidence summing correction factors through the previously mentioned methods, the simulation of the coincident emission of photons was modeled mathematically, based on the analytical equations and complex integrations over the radioactive volumetric sources including the self-attenuation factor. The measured full-energy peak efficiencies and correction factors were done by using 152Eu, where an exact adjustment is required for the detector efficiency curve, because neglecting the coincidence summing effect can make the results inconsistent with the whole. These phenomena, in general due to the efficiency calibration process and the coincidence summing corrections, appear jointly. The full-energy peak and the total efficiencies from the two methods typically agree with discrepancy 10%. The discrepancy between the simulation, ANGLE4 and measured full-energy peak after corrections for the coincidence summing effect was on the average, while not exceeding 14%. Therefore, this technique can be easily applied in establishing the efficiency calibration curves of well-type detectors.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Radioactivity and Radon Measurements · Radiation Detection and Scintillator Technologies · Nuclear Physics and Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Alexandria University 1 Physics Department, Faculty of Science, , 21511 Alexandria, Egypt

Beirut Arab University 2 Department of Physics, Faculty of Science, , Beirut, Lebanon

University of Montenegro 3 Centre for Nuclear Competence and Knowledge Management, , Podgorica, Montenegro

University of Montenegro 4 Department of Physics, Faculty of Mathematics and Natural Sciences, , Podgorica, Montenegro

Pharos University in Alexandria 5 Department of Medical Equipment Technology, Faculty of Allied Medical Sciences, , Alexandria, Egypt

Pharos University in Alexandria 6 Basic Science Department, Faculty of Physical Therapy, , Alexandria, Egypt

University of Montenegro 7 Department of Mathematics, Maritime Faculty, , Kotor, Montenegro

Princess Nourah Bint Abdulrahaman University 8 Physics Department, Faculty of Science, , 11544-55532 Riyadh, Saudi Arabia

References (37)

Show source-provided references
1. Transfer of detector efficiency calibration from a point source to other geometries using ETNA software
2. Testing the equivalence of several algorithms for calculation of coincidence summing corrections10.1016/j.apradiso.2009.07.002 ↗
3. Well Ge and semiplanar Ge (HP) detectors for low-level gamma spectrometry10.1016/0168-9002(94)01351-9 ↗
4. Monte Carlo calculation of the efficiency response of a low-background well-type HPGe detector10.1016/s0168-9002(01)00942-1 ↗
5. Accounting for incomplete charge collection in Monte Carlo simulations of the efficiency of well-type Ge-detectors10.1016/s0168-9002(02)02080-6 ↗
6. Calculation of the absolute peak efficiency of gamma-ray detectors for different counting geometries10.1016/0029-554x(81)90374-8 ↗
7. Coincidence summing corrections in Ge(Li)-spectrometry at low source-to-detector distances10.1016/s0029-554x(79)94845-6 ↗
8. Fast procedures for coincidence-summing correction in γ-ray spectrometry10.1016/s0969-8043(99)00239-0 ↗
9. Efficiency calibration and coincidence summing correction for a 4π NaI(Tl) detector array10.1016/j.nima.2004.06.174 ↗
10. Standardization of 67Ga by 4πγ (NaI) and 4πβ–γ coincidence methods10.1016/j.apradiso.2003.11.041 ↗
11. 1988
12. 1995
13. Coincidence summing corrections for volume samples using the PENELOPE/penEasy Monte Carlo code10.1016/j.apradiso.2013.11.057 ↗
14. Coincidence summing corrections for point and volume 152Eu sources10.1016/j.apradiso.2015.10.015 ↗
15. On the iteration of coincidence summing correction for determination of gamma-ray intensities10.1016/j.apradiso.2015.11.035 ↗
16. Coincidence summing corrections for a clover detector10.1016/j.nima.2014.06.043 ↗
17. Calculation of coincidence summing correction factors for an HPGe detector using GEANT410.1016/j.jenvrad.2016.04.008 ↗
18. Numerical expressions for the computation of coincidence-summing correction factors in γ-ray spectrometry with HPGe detectors10.1016/j.apradiso.2009.10.024 ↗
19. Experimental verification of gamma-efficiency calculations for scintillation detectors in ANGLE4 software10.2298/ntrp1501035t ↗
20. Calculation of relative full-energy peak efficiencies of well-type detectors10.1016/s0168-9002(01)01225-6 ↗
21. Analytical formulae for well-type NaI (Tl) and HPGe detectors efficiency computation10.1016/s0969-8043(01)00047-1 ↗
22. Calibration of the 4π gamma-ray spectrometer by a new numerical simulation approach10.1016/j.apradiso.2010.02.013 ↗
23. Analytical approach to calculate the efficiency of 4p NaI(Tl) gamma-ray detectors for extended sources10.1016/j.nima.2009.12.058 ↗
24. A new analytical method to calibrate cylindrical phoswich and LaBr3(Ce) scintillation detectors10.1016/j.nima.2010.05.025 ↗
25. Analytical formulae for borehole scintillation detectors efficiency calibration10.1016/j.nima.2010.06.241 ↗
26. New numerical simulation approach to calibrate the NaI(Tl) detectors array using non-axial extended spherical sources10.1088/1748-0221/8/11/p11005 ↗
27. Accurate calculation of well-type detector geometrical efficiency using sources with different shapes and geometries10.1088/1748-0221/10/10/p10017 ↗
28. Tables of X-ray mass attenuation coefficients and mass energy absorption coefficients 1 KeV to 20 MeV for elements Z=1 to 92 and 48 additional substances of dosimetric interest
29. Calibration of well-type NaI(Tl) detector using a point sources measured out the detector well at different axial distances10.1088/1748-0221/10/03/p03022 ↗
30. Well-type NaI(Tl) detector efficiency using analytical technique and ANGLE 4 software based on radioactive point sources located out the well cavity10.1016/j.cjph.2016.03.019 ↗
31. A. A. Thabet, Ph D. Thesis, “Calibration of Gamma-rays Detectors Using a New Trend to Calculate the Effective Solid Angle,” Faculty of science (Alexandria University, Egypt, 2015).
32. ETNA (Efficiency transfer for nuclide activity measurements) software for efficiency transfer and coincidence summing corrections in gamma-ray spectrometry
33. IAEA website (https://www-nds.iaea.org/).
34. Laboratoire National Henri Becquerel web page: http://www.nucleide.org.
35. Extraordinary improvement in scintillation detectors via post-processing with ASEDRA: solution to a 50-year-old problem
36. Table de Radionucléides
37. National Nuclear Data Center Web Page (https://www.nndc.bnl.gov/).