journal article·2026-07-27

Soft, Multi‐Wavelength Photoplethysmography Enables Reliable Neonatal Blood Pressure Monitoring Via Error Stratification

Wenqi Shi, Lanlan Mi, Jiarong Chen, Biao Liu, Min Wang, Zisheng Dai, Bin Liu, Ruize Gao, Yunfeng Lin, Weiwei Guo, Guoxing Wang, Wenyi Luo, Cheng Chen

Publisher page

Abstract

ABSTRACT Continuous noninvasive blood pressure monitoring is important for neonatal hemodynamic management. However, cuffless photoplethysmography (PPG) is still difficult to use clinically because acquisition‐related errors and their effects on blood pressure estimation are not well understood. Most studies report only overall error, which can hide error patterns under different signal‐quality and motion conditions. Here, we developed a soft multi‐wavelength wearable that combines reflective and transmissive optical paths to record complementary PPG signals. We also built NEO‐BP, a prospective clinical dataset containing synchronized multi‐wavelength PPG and invasive arterial blood pressure waveforms from 42 neonates. In a held‐out segment‐level test set without excluding samples based on signal quality (n = 9378), the green‐red‐infrared (G+R+IR) model achieved mean absolute errors of 9.67 mmHg for systolic blood pressure and 6.17 mmHg for diastolic blood pressure. Using this dataset, we established a retrospective descriptive subgroup error profile by stratifying estimation deviations across measurable acquisition and physiological conditions. Estimation errors increased under lower signal quality and stronger motion. These results identify error‐prone acquisition states, support quality‐aware interpretation of segment‐level cuffless blood pressure estimates, and provide a basis for future models that estimate the error risk of individual readings.
Source: crossref · Quality: baseline

Affiliations

National Key Laboratory of Advanced Micro and Nano Manufacture Technology Shanghai Jiao Tong University Shanghai China

School of Integrated Circuits (School of Information Science and Electronic Engineering) Shanghai Jiao Tong University Shanghai China

Department of Neonatology, Shanghai Children's Medical Center, School of Medicine Shanghai Jiao Tong University Shanghai China

Department of Neonatology Fujian Children's Hospital (Fujian Branch of Shanghai Children's Medical Center) Fujian China

Shenzhen Ninenovo Technology Limited Shenzhen China

Department of Nursing, Shanghai Children's Medical Center, School of Medicine Shanghai Jiao Tong University Shanghai China

Funding

Shanghai Jiao Tong University · Fujian Provincial Department of Science and Technology

References (42)

Show source-provided references
1. Neonatal Blood Pressure by Birth Weight, Gestational Age, and Postnatal Age: A Systematic Review10.1186/s40748-024-00180-w ↗
2. Baker S. · 2023 · 3107
3. Wearable Biosensors in Pediatric Cardiovascular Disease: Promises and Pitfalls Toward Generating Actionable Insights10.1161/CIRCULATIONAHA.119.038483 ↗
4. Hemodynamic Bedside Monitoring Instrument with Pressure and Optical Sensors: Validation and Modality Comparison10.1002/advs.202307718 ↗
5. A Review of Wearable Multi‐Wavelength Photoplethysmography10.1109/RBME.2021.3121476 ↗
6. W.Shi L.Mi Z.Dai C.Chen W.Luo W.Guo andG.Wang “Live Demonstration: Yupei: A Flexible Wireless Multi‐Physiological Parameters Wearable Device for NICU ” in2024 IEEE Biomedical Circuits and Systems Conference (BioCAS)(IEEE 2024) 1–2.10.1109/BioCAS61083.2024.10798205 ↗
7. A Benchmark for Machine‐Learning‐Based Non‐Invasive Blood Pressure Estimation Using Photoplethysmogram10.1038/s41597-023-02020-6 ↗
8. 10.1136/bmjinnov‐2016‐00014510.1136/bmjinnov‐2016‐000145 ↗
9. Binodal, Wireless Epidermal Electronic Systems with In‐Sensor Analytics for Neonatal Intensive Care10.1126/science.aau0780 ↗
10. Skin‐Interfaced Biosensors for Advanced Wireless Physiological Monitoring in Neonatal and Pediatric Intensive‐Care Units10.1038/s41591-020-0792-9 ↗
11. Advancing Wearable Biosensors for Congenital Heart Disease: Patient and Clinician Perspectives: A Science Advisory from the American Heart Association10.1161/CIR.0000000000001225 ↗
12. Synchronized Wearables for the Detection of Haemodynamic States via Electrocardiography and Multispectral Photoplethysmography10.1038/s41551-023-01098-y ↗
13. Wireless, Skin‐Interfaced Devices for Pediatric Critical Care: Application to Continuous, Noninvasive Blood Pressure Monitoring10.1002/adhm.202100383 ↗
14. Neonatal Wearable Device for Colorimetry‐Based Real‐Time Detection of Jaundice with Simultaneous Sensing of Vitals10.1126/sciadv.abe3793 ↗
15. Skin Sensors Are the Future of Health Care10.1038/d41586-019-02143-0 ↗
16. Cuffless Blood Pressure Measurement10.1146/annurev-bioeng-110220-014644 ↗
17. The Use of Photoplethysmography for Assessing Hypertension10.1038/s41746-019-0136-7 ↗
18. Wearable Photoplethysmography for Cardiovascular Monitoring10.1109/JPROC.2022.3149785 ↗
19. G.Nie J.Zhu G.Tang et al. “A Review of Deep Learning Methods for Photoplethysmography Data ”arXivpreprint arXiv:2401.12783 (2024).
20. PCA‐Based Multi‐Wavelength Photoplethysmography Algorithm for Cuffless Blood Pressure Measurement on Elderly Subjects10.1109/JBHI.2020.3004032 ↗
21. Z.Hou D.Huang Y.Huang et al. “Exploiting Multi‐Wavelength Morphological Features of Camera‐PPG for Blood Pressure Estimation ”IEEE Transactions on Instrumentation and Measurement(2025).10.1109/TIM.2025.3551005 ↗
22. The Past, Current, and Future of Neonatal Intensive Care Units with Artificial Intelligence: A Systematic Review10.1038/s41746-023-00941-5 ↗
23. Feasibility and Measurement Stability of Smartwatch‐Based Cuffless Blood Pressure Monitoring: A Real‐World Prospective Observational Study10.1038/s41440-023-01215-z ↗
24. Advances in Non‐Invasive Blood Pressure Measurement Techniques10.1109/RBME.2022.3141877 ↗
25. Quantification of the Phenomena Affecting Reflective Arterial Photoplethysmography10.3390/bioengineering10040460 ↗
26. Feasibility Study for the Non‐Invasive Blood Pressure Estimation Based on PPG Morphology: Normotensive Subject Study10.1186/s12938-016-0302-y ↗
27. ArvaNet: Deep Recurrent Architecture for PPG‐Based Negative Mental‐State Monitoring10.1109/TCSS.2020.2977715 ↗
28. NeuroKit2: A Python Toolbox for Neurophysiological Signal Processing10.3758/s13428-020-01516-y ↗
29. Multi‐Wavelength Photoplethysmography Enabling Continuous Blood Pressure Measurement with Compact Wearable Electronics10.1109/TBME.2018.2874957 ↗
30. Signal‐Quality Indices for the Electrocardiogram and Photoplethysmogram: Derivation and Applications to Wireless Monitoring
31. “Non‐Invasive Sphygmomanometers—Part 2: Clinical Investigation of Intermittent Automated Measurement Type ”International Organization for Standardization Standard ISO 81060‐2:2018 (2018) amendments: 2020 2024.
32. “IEEE Standard for Wearable Cuffless Blood Pressure Measuring Devices ”Institute of Electrical and Electronics Engineers Standard IEEE Std 1708‐2014 (2014).
33. “IEEE Standard for Wearable Cuffless Blood Pressure Measuring Devices—Amendment 1 ”Institute of Electrical and Electronics Engineers Standard IEEE Std 1708a‐2019 (2019) clarifications on cohort selection posture/activity applicability and motion artifacts.
34. The British Hypertension Society Protocol for the Evaluation of Blood Pressure Measuring Devices
35. PulseDB: A Large, Cleaned Dataset Based on MIMIC‐III and VitalDB for Benchmarking Cuff‐Less Blood Pressure Estimation Methods10.3389/fdgth.2022.1090854 ↗
36. Q.Hong C.Bian X.Zhou X.Li Y.Li andZ.Zeng “Lost in Time? A Meta‐Learning Framework for Time‐Shift‐Tolerant Physiological Signal Transformation ” inProceedings of the AAAI Conference on Artificial Intelligence40(2026) 21761–21769.10.1609/aaai.v40i26.39327 ↗
37. Photoplethysmographic Sensors, Potential and Limitations: Is It Time for Regulation? A Comprehensive Review10.1016/j.measurement.2023.113150 ↗
38. B.Liu H.Wu G.Wang J.Chen C.Chen andG. K. H.Pang “Ring‐BP: Using a Wearable Smart Ring to Cufflessly Estimate Blood Pressure with Mobile and Efficient Net ” in2025 IEEE 7th International Conference on Artificial Intelligence Circuits and Systems (AICAS)(2025) 1–5.10.1109/AICAS64808.2025.11173101 ↗
39. S.Abbaspourazad O.Elachqar A. C.Miller S.Emrani U.Nallasamy andI.Shapiro “Large‐Scale Training of Foundation Models for Wearable Biosignals ”arXivpreprint arXiv:2312.05409 (2023).
40. Multi‐Channel Masked Autoencoder and Comprehensive Evaluations for Reconstructing 12‐Lead ECG from Arbitrary Single‐Lead ECG10.1038/s44325-024-00036-4 ↗
41. Does Measurement of Four‐Limb Blood Pressures at Birth Improve Detection of Aortic Arch Anomalies?10.1038/jp.2015.203 ↗
42. Comparison of Calf and Brachial Blood Pressures in Infants: Is There a Difference Between Calf and Brachial Blood Pressures?10.1016/j.jvn.2014.03.003 ↗