Distinguishing shear and tensile myocardial wall stiffness using ex vivo anisotropic Magnetic Resonance Elastography

Cyril Tous, Guillaume Flé, Stanislas Rapacchi, Matthew McGarry, Philip Bayly, Keith Paulsen, Curtis L. Johnson, Elijah Van Houten

Acta Biomaterialia· Volume 202

Abstract

ABSTRACT The organized myofiber structure within the myocardium indicates its mechanical anisotropy. By projecting the MR Elastography (MRE) stiffness matrix along either the myocardial fiber or sheet orientations determined by Diffusion Tensor Imaging (DTI), anisotropic MRE (aMRE) maps axial and transverse shear and Young’s moduli into three tensile and six shear deformation modes. Ten healthy ex vivo swine hearts were imaged three times at 3T using MRE and DTI sequences. aMRE results showed a within-subject coefficient of variation at 19% for the fiber model and 28% for the sheet model across specimens and metrics, with coefficients lower than 15% for seven of the ten specimens across models. Repeatability coefficient of ±0.5 kPa for Young’s moduli and ±0.17 kPa for shear’s moduli, demonstrating repeatability within the 95% agreement limit. Isotropic MRE underestimated stiffnesses by 31% compared to aMRE, where anisotropic moduli aligned more closely with prior finite element studies and some mechanical loading tests. The myocardium’s anisotropic elasticity reflects with its helicoidal myofiber microstructure, with mid-wall circumferential fibers requiring twice the force to deform as longitudinal fibers at the epicardium or endocardium. At the mid-wall, fiber model values were μ ax = 1.9 ± 0.1 kPa, μ tra = 1.3 ± 0.1 kPa, E ax = 5.6 ± 0.4 kPa, and E tra = 3.8 ± 0.3 kPa. Identified deformation modes included: (FF), (NN), (FF or SS), (NN or SS), (SN or NS), (FN or FS), (SF or FS), and (SN or NF), where N is normal to both fiber (F) and sheet (S) orientations. By aligning elasticity matrices more accurately with myocardial architecture than isotropic MRE, aMRE was able to reliably measure shear and Young’s moduli in ex vivo swine hearts. These mappings of deformation modes may bring myocardial stiffness assessment closer to clinical application, providing a foundation for a non-invasive methodology capable of true mechanical characterization of the cardiac wall using MR imaging. Statement of Significance The myocardium’s anisotropic elasticity, due to its helicoidal myofiber structure, is revealed through anisotropic MR elastography, using fiber and sheet elastic models. Mid-wall circumferential fibers require twice the force to deform equally compared to epicardial or endocardial fibers. Characterizing shear and Young’s moduli across cardiac modes offers noninvasive measures of ventricular compliance, comparable to pressure-volume relationships. This could enhance early diagnosis of “stiff heart syndrome” and clarify its underlying mechanisms. Additionally, it aids understanding of myocardial pathologies, including amyloidosis, hypertrophic and dilated cardiomyopathies, and ischemic damage. By characterizing tensile and shear interactions, it may inform diagnosis and treatment of conduction issues and arrhythmia, where tissue has lost its normal mechanical behavior, while patient-specific models could optimize surgical and therapeutic outcomes.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Elasticity and Material Modeling · Advanced Neuroimaging Techniques and Applications · Cardiovascular Function and Risk Factors

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (79)

Show source-provided references
1. Biomechanical and hemodynamic measures of right ventricular diastolic function: translating tissue biomechanics to clinical relevance10.1161/JAHA.117.006084 ↗
2. Biomechanical properties and mechanobiology of cardiac ECM10.1007/978-3-319-97421-7_1 ↗
3. J.W. Holmes, T.K. Borg, J.W. Covell, Structure and mechanics of healing myocardial infarcts, 10.1146/Annurev.Bioeng.7.060804.100453. 7 (2005) 223–253. https://doi.org/10.1146/ANNUREV.BIOENG.7.060804.100453.10.1146/annurev.bioeng.7.060804.100453 ↗
4. A machine learning model to estimate myocardial stiffness from EDPVR
5. Comparison of three machine learning methods to estimate myocardial stiffness10.1016/B978-0-32-389967-3.00025-1 ↗
6. Passive myocardial mechanical properties: meaning, measurement, models10.1007/s12551-021-00838-1 ↗
7. In vivo assessment of MR elastography-derived effective end-diastolic myocardial stiffness under different loading conditions10.1002/jmri.22531 ↗
8. T. Elgeti, F. Knebel, R. Hättasch, B. Hamm, J. Braun, I. Sack, Shear-wave amplitudes measured with cardiac MR elastography for diagnosis of diastolic dysfunction, 10.1148/Radiol.13131605. 271 (2014) 681–687. https://doi.org/10.1148/RADIOL.13131605.10.1148/radiol.13131605 ↗
9. In vivo waveguide cardiac magnetic resonance elastography10.1186/1532-429X-17-S1-P35 ↗
10. In vivo quantification of myocardial stiffness in hypertensive porcine hearts using MR elastography10.1002/jmri.25423 ↗
11. Cardiac MR elastography for quantitative assessment of elevated myocardial stiffness in cardiac amyloidosis10.1002/jmri.25678 ↗
12. Measuring age-dependent myocardial stiffness across the cardiac cycle using MR elastography: a reproducibility study10.1002/mrm.25760 ↗
13. Regional assessment of in vivo myocardial stiffness using 3D magnetic resonance elastography in a porcine model of myocardial infarction10.1002/mrm.26695 ↗
14. Laminar structure of the heart: ventricular myocyte arrangement and connective tissue architecture in the dog
15. The presence of two local myocardial sheet populations confirmed by diffusion tensor MRI and histological validation10.1002/jmri.22725 ↗
16. Shear properties of passive ventricular myocardium10.1152/ajpheart.00111.2002 ↗
17. A viscoelastic model for human myocardium10.1016/j.actbio.2021.08.036 ↗
18. Observation of direction-dependent mechanical properties in the human brain with multi-excitation MR elastography10.1016/j.jmbbm.2016.03.005 ↗
19. Mapping heterogenous anisotropic tissue mechanical properties with transverse isotropic nonlinear inversion MR elastography10.1016/j.media.2022.102432 ↗
20. Anisotropic mechanical properties in the healthy human brain estimated with multi-excitation transversely isotropic MR elastography10.1016/j.brain.2022.100051 ↗
21. Quantifying stability of parameter estimates for in vivo nearly incompressible transversely-isotropic brain MR elastography10.1088/2057-1976/ac5ebe ↗
22. Impact of anesthesia and euthanasia on metabolomics of mammalian tissues: studies in a C57BL/6J mouse model10.1371/journal.pone.0117232 ↗
23. Rigor mortis at the myocardium investigated by post-mortem magnetic resonance imaging10.1016/j.forsciint.2015.07.039 ↗
24. Effects of fixation and preservation on tissue elastic properties measured by quantitative optical coherence elastography (OCE)10.1016/j.jbiomech.2016.02.013 ↗
25. B.S.C.L. Johnson, US11033199B2 - Echo-planar imaging magnetic resonance elastography pulse sequence - Google patents, (2016). https://patents.google.com/patent/US11033199B2/en.
26. Ex vivo cardiovascular magnetic resonance diffusion weighted imaging in congenital heart disease, an insight into the microstructures of tetralogy of Fallot, biventricular and univentricular systemic right ventricle10.1186/s12968-020-00662-8 ↗
27. The effect of gradient sampling schemes on measures derived from diffusion tensor MRI: a Monte Carlo study10.1002/mrm.20033 ↗
28. Optimal strategies for measuring diffusion in anisotropic systems by magnetic resonance imaging10.1002/(SICI)1522-2594(199909)42:3<515::AID-MRM14>3.0.CO;2-Q ↗
29. Condition number as a measure of noise performance of diffusion tensor data acquisition schemes with MRI10.1006/jmre.2000.2209 ↗
30. High-resolution diffusion tensor mr imaging for the evaluation of myocardial anisotropy and fiber tracking at 3 T: effect of the numbers of diffusion-sensitizing gradient directions10.1186/1532-429X-11-S1-P175 ↗
31. Optimal diffusion weighting for in vivo cardiac diffusion tensor imaging10.1002/mrm.25418 ↗
32. An in vivo comparison of stimulated-echo and motion compensated spin-echo sequences for 3 T diffusion tensor cardiovascular magnetic resonance at multiple cardiac phases10.1186/s12968-017-0425-8 ↗
33. Convex optimized diffusion encoding (CODE) gradient waveforms for minimum echo time and bulk motion-compensated diffusion-weighted MRI10.1002/mrm.26166 ↗
34. Spin echo versus stimulated echo diffusion tensor imaging of the in vivo human heart
35. A heterogenous, time harmonic, nearly incompressible transverse isotropic finite element brain simulation platform for MR elastography
36. An overlapping subzone technique for MR-based elastic property reconstruction10.1002/(SICI)1522-2594(199910)42:4<779::AID-MRM21>3.0.CO;2-Z ↗
37. Mechanically anisotropic phantoms for magnetic resonance elastography
38. General guidelines for the performance of viscoelastic property identification in elastography: a Monte-Carlo analysis from a closed-form solution10.1002/cnm.3741 ↗
39. Multiresolution MR elastography using nonlinear inversion10.1118/1.4754649 ↗
40. An octahedral shear strain-based measure of SNR for 3D MR elastography10.1088/0031-9155/56/13/N02 ↗
41. Evaluation of the standard deviation from duplicate results10.1007/s00769-008-0390-x ↗
42. On the AIC-based model reduction for the general Holzapfel–Ogden myocardial constitutive law10.1007/s10237-019-01140-6 ↗
43. S.P. Arunachalam, A. Arani, K.J. Glaser, R.L. Ehman, P.A. Araoz, In vivo assessment of systolic and diastolic myocardial stiffness in pigs using 3d cardiac magnetic resonance elastography, (2017).
44. Magnetic resonance elastography as a method for the assessment of effective myocardial stiffness throughout the cardiac cycle10.1002/mrm.22467 ↗
45. Biomechanical properties and microstructure of neonatal porcine ventricles10.1016/j.jmbbm.2018.07.038 ↗
46. Estimation of transversely isotropic material properties from magnetic resonance elastography using the optimised virtual fields method10.1002/cnm.2979 ↗
47. Left ventricular diastolic myocardial stiffness and end-diastolic myofibre stress in Human heart failure using personalised biomechanical analysis10.1007/s12265-018-9816-y ↗
48. Insights into the passive mechanical behavior of left ventricular myocardium using a robust constitutive model based on full 3D kinematics10.1016/j.jmbbm.2019.103508 ↗
49. In vivo estimation of normal left ventricular stiffness and contractility based on routine cine MR acquisition10.1016/j.medengphy.2020.09.003 ↗
50. Effects of dispersed fibres in myocardial mechanics, part I: passive response10.3934/mbe.2022183 ↗
51. Kolipaka, A feasibility study: MR elastography as a method to compare stiffness estimates in hypertrophic obstructive cardiomyopathy & in normal volunteers, (2011). https://archive.ismrm.org/2011/0274.html (accessed January 5, 2023).
52. Biomechanical properties and microstructure of human ventricular myocardium10.1016/j.actbio.2015.06.031 ↗
53. Right ventricular myocardial mechanics: multi-modal deformation, microstructure, modeling, and comparison to the left ventricle10.1016/j.actbio.2020.12.006 ↗
54. Finite element stress analysis of left ventricular mechanics in the beating dog heart10.1016/0021-9290(94)00174-3 ↗
55. Structure-based finite strain modelling of the human left ventricle in diastole10.1002/cnm.2497 ↗
56. Influence of myocardial fiber/sheet orientations on left ventricular mechanical contraction10.1177/1081286513485779 ↗
57. Modeling the dispersion in electromechanically coupled myocardium10.1002/cnm.2575 ↗
58. An orthotropic viscoelastic model for the passive myocardium: continuum basis and numerical treatment10.1080/10255842.2016.1176155 ↗
59. Relative identifiability of anisotropic properties from magnetic resonance elastography10.1002/nbm.3848 ↗
60. Comparison of biomechanical properties and microstructure of trabeculae carneae, papillary muscles, and myocardium in the Human heart10.1115/1.4041966 ↗
61. Oomens · 2017
62. In vivo, high-frequency three-dimensional cardiac MR elastography: feasibility in normal volunteers10.1002/mrm.26101 ↗
63. Measuring age-dependent myocardial stiffness across the cardiac cycle using MR elastography: a reproducibility study10.1002/mrm.25760 ↗
64. Quantitative 3D magnetic resonance elastography: comparison with dynamic mechanical analysis10.1002/mrm.26207 ↗
65. Magnetic resonance imaging assessment of myocardial elastic modulus and viscosity using displacement imaging and phase-contrast velocity mapping10.1002/mrm.20589 ↗
66. Quantification of the mechanical properties of noncontracting canine myocardium under simultaneous biaxial loading10.1016/0021-9290(87)90279-X ↗
67. Regional mechanical properties of passive myocardium10.1016/0021-9290(94)90016-7 ↗
68. Quantitative comparison of myocardial fiber structure between mice, rabbit, and sheep using diffusion tensor cardiovascular magnetic resonance10.1186/1532-429X-13-74 ↗
69. In vivo estimation of passive biomechanical properties of human myocardium10.1007/s11517-017-1768-x ↗
70. Robust and efficient fixed-point algorithm for the inverse elastostatic problem to identify myocardial passive material parameters and the unloaded reference configuration10.1016/j.jcp.2022.111266 ↗
71. Extraction of the 3D local orientation of myocytes in human cardiac tissue using X-ray phase-contrast micro-tomography and multi-scale analysis10.1016/j.media.2017.02.006 ↗
72. Measurement of local orientation of cardiomyocyte aggregates in human left ventricle free wall samples using X-ray phase-contrast microtomography10.1016/j.media.2021.102269 ↗
73. Probing cardiomyocyte mobility with multi-phase cardiac diffusion tensor MRI10.1371/journal.pone.0241996 ↗
74. Viscoelastic polyacrylamide MR elastography phantoms with tunable damping ratio independent of shear stiffness10.1016/j.jmbbm.2024.106522 ↗
75. Phantom evaluations of low frequency MR elastography10.1088/1361-6560/ab0290 ↗
76. Phantom evaluations of nonlinear inversion MR elastography10.1088/1361-6560/aacb08 ↗
77. MR elastography at 1 hz of gelatin phantoms using 3D or 4D acquisition10.1016/j.jmr.2018.08.012 ↗
78. Shear wave speeds in nearly-incompressible fibrous materials with two fiber families10.1121/10.0003528 ↗
79. Mechanical properties of porcine brain tissue in vivo and ex vivo estimated by MR elastography10.1016/j.jbiomech.2018.01.016 ↗