Effects of a roughened femoral head and the locus of grafting on the wear resistance of the phospholipid polymer-grafted acetabular liner

Toru Moro, Kazuhiko Ishihara, Yoshio Takatori, Sakae Tanaka, Masayuki Kyomoto, Masami Hashimoto, Hisatoshi Ishikura, Ryo Hidaka, Takeyuki Tanaka, Hiroshi Kawaguchi, Kozo Nakamura

Acta Biomaterialia· Volume 86
Publisher page

Abstract

Although laboratory tests and mid-term clinical outcomes show the clinical safety and remarkable wear resistance of the highly cross-linked polyethylene (HXLPE) acetabular liner with a nanometer-scaled graft layer of poly(2-methacryloyloxyethyl phosphorylcholine) (PMPC), the wear resistance of the layer under severe abrasive conditions is concerning. We evaluated the effects of a roughened femoral head and the grafting locus on the wear resistance of the PMPC-grafted HXLPE liner and the effect of PMPC grafting on wear resistance of the HXLPE substrate by removing the PMPC-grafted layer using a severely roughened femoral head. Against a moderately roughened femoral head, the PMPC-grafted HXLPE liner showed negative wear, although an untreated HXLPE liner increased the wear by 154.1% compared with wear against a polished femoral head, confirming that PMPC grafts were unaffected. Against a severely roughened femoral head, the PMPC-grafted layer of the head contact area might be removed under severe conditions. However, the wear rate was reduced by 52.5% compared to that of untreated HXLPE liners. Moreover, the head non-contact area-modified PMPC-grafted HXLPE liner against a polished femoral head reduced the wear by 76.8% compared with untreated HXLPE liner; thus, this area may be also important in the development of fluid-film lubrication. STATEMENT OF SIGNIFICANCE: Here we describe effects of a roughened femoral head and the locus of grafting on the wear-resistance of the phospholipid polymer grafted highly cross-linked polyethylene (PMPC-HXLPE) liner. Against a moderately roughened femoral head, the PMPC-HXLPE liner showed negative wear, confirming that PMPC grafts were unaffected. After removing the PMPC layer of the head contact area using a severely roughened femoral head, the wear rate not only exceeded that of untreated HXLPE liners, but was reduced by 52.5%, confirming that PMPC grafting does not affect the wear-resistance of the HXLPE substrate. In addition, the head non-contact area-modified PMPC-HXLPE liner reduced the wear by 76.8%. Thus, this area may also may be important in the development of fluid-film lubrication.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Asian Studies and History · Linguistic, Cultural, and Literary Studies · Education and Islamic Studies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (46)

Show source-provided references
1. The epidemiology of revision total hip arthroplasty in the United States10.2106/JBJS.H.00155 ↗
2. Wear and periprosthetic osteolysis - the problem10.1097/00003086-200112000-00007 ↗
3. R.D. Crowninshield, O.K. Muratoglu, G. Implant Wear Symposium Engineering Work, How have new sterilization techniques and new forms of polyethylene influenced wear in total joint replacement?, J. Am. Acad. Orthop. Surg. 16 Suppl 1 (2008) S80-5.10.5435/00124635-200800001-00017 ↗
4. S. Graves, D. Davidson, R. de Steiger, P. Lewis, Australian Orthopaedic Association National Joint Replacement Registry, 2017 Annual Report - Hip and Knee Arthroplasty; 2017. Available from URL:https://aoanjrr.sahmri.com/annual-reports-2017
5. Wear and osteolysis of highly crosslinked polyethylene at 10 to 14 years: the effect of femoral head size10.1007/s11999-015-4319-5 ↗
6. Analysis of products diffused into UHMWPE prosthetic components in vivo10.1016/S0142-9612(00)00182-4 ↗
7. Highly cross-linked polyethylene reduces wear and revision rates in total hip arthroplasty: a 10-year double-blinded randomized controlled trial10.2106/JBJS.16.00878 ↗
8. Evaluation of the response of primary human peripheral blood mononuclear phagocytes to challenge with in vitro generated clinically relevant UHMWPE particles of known size and dose10.1002/1097-4636(200011)52:2<296::AID-JBM8>3.0.CO;2-9 ↗
9. Comparative wear and wear debris under three different counterface conditions of crosslinked and non-crosslinked ultra high molecular weight polyethylene
10. Second-generation annealed highly cross-linked polyethylene exhibits low wear10.1007/s11999-011-2177-3 ↗
11. Vitamin-E diffused highly cross-linked polyethylene liner compared to standard liners in total hip arthroplasty. A randomized, controlled trial10.1007/s00264-015-2680-3 ↗
12. Clinical safety and wear resistance of the phospholipid polymer-grafted highly cross-linked polyethylene liner10.1002/jor.23473 ↗
13. Surface grafting of artificial joints with a biocompatible polymer for preventing periprosthetic osteolysis10.1038/nmat1233 ↗
14. The morphology and composition of the superficial zone of mammalian articular-cartilage
15. Frank Stinchfield Award: grafting of biocompatible polymer for longevity of artificial hip joints10.1097/01.blo.0000246553.33434.5f ↗
16. Wear resistance of artificial hip joints with poly(2-methacryloyloxyethyl phosphorylcholine) grafted polyethylene: comparisons with the effect of polyethylene cross-linking and ceramic femoral heads10.1016/j.biomaterials.2009.02.020 ↗
17. Effects of photo-induced graft polymerization of 2-methacryloyloxyethyl phosphorylcholine on physical properties of cross-linked polyethylene in artificial hip joints10.1007/s10856-007-3063-0 ↗
18. Long-term hip simulator testing of the artificial hip joint bearing surface grafted with biocompatible phospholipid polymer10.1002/jor.22526 ↗
19. Wear resistance of the biocompatible phospholipid polymer-grafted highly cross-linked polyethylene liner against larger femoral head10.1002/jor.22868 ↗
20. Effect of head surface roughness and sterilization on wear of UHMWPE acetabular cups10.1002/jbm.a.32161 ↗
21. Wear of gamma-crosslinked polyethylene acetabular cups against roughened femoral balls10.1097/00003086-199912000-00008 ↗
22. Wear of conventional and cross-linked ultra-high-molecular-weight polyethylene acetabular cups against polished and roughened CoCr femoral heads in a biaxial hip simulator10.1002/jbm.10471 ↗
23. In vivo femoral head damage and its effect on polyethylene wear10.1016/j.arth.2009.01.010 ↗
24. The effect of surface topography of retrieved femoral heads on the wear of UHMWPE sockets10.1016/S1350-4533(97)00032-5 ↗
25. Femoral prosthetic heads and their significance in polyethylene wear10.1007/s002640000129 ↗
26. Identification of scratch mechanisms on a retrieved metallic femoral head10.1016/j.wear.2004.09.035 ↗
27. Cartilage-mimicking, high-density brush structure improves wear resistance of crosslinked polyethylene: a pilot study10.1007/s11999-010-1718-5 ↗
28. Poor survival of cementless Biomet total hip: a report on 1,047 hips from the Finnish Arthroplasty Register10.3109/17453679909000975 ↗
29. Long-term results of the threaded Mecron cup in primary total hip arthroplasty : a 15–20-year follow-up study10.1007/s00264-009-0843-9 ↗
30. Acetabular screw rings and surface treatment10.1097/00003086-199707000-00017 ↗
31. Bipolar hip arthroplasty10.1016/j.arth.2011.03.003 ↗
32. The predictive power of surface profile parameters on the amount of wear measured in vitro on metal-on-polyethylene artificial hip joints10.1243/09544119JEIM95 ↗
33. Force actions transmitted by joints in the human body10.1098/rspb.1976.0004 ↗
34. Surface analysis of early retrieved acetabular polyethylene liners: a comparison of conventional and highly crosslinked polyethylenes10.1016/j.arth.2003.08.003 ↗
35. Isolation and characterization of UHMWPE wear particles down to ten nanometers in size from in vitro hip and knee joint simulators10.1002/jbm.a.30824 ↗
36. Surface grafting of biocompatible phospholipid polymer MPC provides wear resistance of tibial polyethylene insert in artificial knee joints10.1016/j.joca.2010.05.019 ↗
37. The lubricating ability of biomembrane models with dipalmitoyl phosphatidylcholine and gamma-globulin10.1243/0954411981534114 ↗
38. Influences of dehydration and rehydration on the lubrication properties of phospholipid polymer-grafted cross-linked polyethylene10.1177/0954411915588969 ↗
39. Study of the distribution of the molecular orientation in thick polyethylene samples by X-ray-diffraction, infrared dichroism and raman-spectroscopy10.1016/0032-3861(93)90245-6 ↗
40. Effect of counterface material on the characteristics of retrieved uncemented cobalt-chromium and titanium alloy total hip replacements10.1243/PIME_PROC_1996_210_412_02 ↗
41. A novel method of cross-linking ultra-high-molecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties. Recipient of the HAP Paul Award
42. The basis for a second-generation highly cross-linked UHMWPE10.1097/01.blo.0000238856.61862.7d ↗
43. Wear resistance and mechanical properties of highly cross-linked, ultrahigh-molecular weight polyethylene doped with vitamin E10.1016/j.arth.2005.07.009 ↗
44. Wear versus thickness and other features of 5-Mrad crosslinked UHMWPE acetabular liners10.1007/s11999-010-1555-6 ↗
45. A phospholipid polymer graft layer affords high resistance for wear and oxidation under load bearing conditions10.1016/j.jmbbm.2017.12.036 ↗
46. Early osteolysis with Hylamer acetabular liners10.1016/S0883-5403(98)90016-X ↗