Effects of a roughened femoral head and the locus of grafting on the wear resistance of the phospholipid polymer-grafted acetabular liner
Source abstract
Abstract
Although laboratory tests and mid-term clinical outcomes show the clinical safety and remarkable wear resistance of the highly cross-linked polyethylene (HXLPE) acetabular liner with a nanometer-scaled graft layer of poly(2-methacryloyloxyethyl phosphorylcholine) (PMPC), the wear resistance of the layer under severe abrasive conditions is concerning. We evaluated the effects of a roughened femoral head and the grafting locus on the wear resistance of the PMPC-grafted HXLPE liner and the effect of PMPC grafting on wear resistance of the HXLPE substrate by removing the PMPC-grafted layer using a severely roughened femoral head. Against a moderately roughened femoral head, the PMPC-grafted HXLPE liner showed negative wear, although an untreated HXLPE liner increased the wear by 154.1% compared with wear against a polished femoral head, confirming that PMPC grafts were unaffected. Against a severely roughened femoral head, the PMPC-grafted layer of the head contact area might be removed under severe conditions. However, the wear rate was reduced by 52.5% compared to that of untreated HXLPE liners. Moreover, the head non-contact area-modified PMPC-grafted HXLPE liner against a polished femoral head reduced the wear by 76.8% compared with untreated HXLPE liner; thus, this area may be also important in the development of fluid-film lubrication. STATEMENT OF SIGNIFICANCE: Here we describe effects of a roughened femoral head and the locus of grafting on the wear-resistance of the phospholipid polymer grafted highly cross-linked polyethylene (PMPC-HXLPE) liner. Against a moderately roughened femoral head, the PMPC-HXLPE liner showed negative wear, confirming that PMPC grafts were unaffected. After removing the PMPC layer of the head contact area using a severely roughened femoral head, the wear rate not only exceeded that of untreated HXLPE liners, but was reduced by 52.5%, confirming that PMPC grafting does not affect the wear-resistance of the HXLPE substrate. In addition, the head non-contact area-modified PMPC-HXLPE liner reduced the wear by 76.8%. Thus, this area may also may be important in the development of fluid-film lubrication.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Asian Studies and History · Linguistic, Cultural, and Literary Studies · Education and Islamic Studies
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (46)
Show source-provided references
1. The epidemiology of revision total hip arthroplasty in the United States10.2106/JBJS.H.00155 ↗
2. Wear and periprosthetic osteolysis - the problem10.1097/00003086-200112000-00007 ↗
3. R.D. Crowninshield, O.K. Muratoglu, G. Implant Wear Symposium Engineering Work, How have new sterilization techniques and new forms of polyethylene influenced wear in total joint replacement?, J. Am. Acad. Orthop. Surg. 16 Suppl 1 (2008) S80-5.10.5435/00124635-200800001-00017 ↗
4. S. Graves, D. Davidson, R. de Steiger, P. Lewis, Australian Orthopaedic Association National Joint Replacement Registry, 2017 Annual Report - Hip and Knee Arthroplasty; 2017. Available from URL:https://aoanjrr.sahmri.com/annual-reports-2017
5. Wear and osteolysis of highly crosslinked polyethylene at 10 to 14 years: the effect of femoral head size10.1007/s11999-015-4319-5 ↗
6. Analysis of products diffused into UHMWPE prosthetic components in vivo10.1016/S0142-9612(00)00182-4 ↗
7. Highly cross-linked polyethylene reduces wear and revision rates in total hip arthroplasty: a 10-year double-blinded randomized controlled trial10.2106/JBJS.16.00878 ↗
8. Evaluation of the response of primary human peripheral blood mononuclear phagocytes to challenge with in vitro generated clinically relevant UHMWPE particles of known size and dose10.1002/1097-4636(200011)52:2<296::AID-JBM8>3.0.CO;2-9 ↗
9. Comparative wear and wear debris under three different counterface conditions of crosslinked and non-crosslinked ultra high molecular weight polyethylene
10. Second-generation annealed highly cross-linked polyethylene exhibits low wear10.1007/s11999-011-2177-3 ↗
11. Vitamin-E diffused highly cross-linked polyethylene liner compared to standard liners in total hip arthroplasty. A randomized, controlled trial10.1007/s00264-015-2680-3 ↗
12. Clinical safety and wear resistance of the phospholipid polymer-grafted highly cross-linked polyethylene liner10.1002/jor.23473 ↗
13. Surface grafting of artificial joints with a biocompatible polymer for preventing periprosthetic osteolysis10.1038/nmat1233 ↗
14. The morphology and composition of the superficial zone of mammalian articular-cartilage
15. Frank Stinchfield Award: grafting of biocompatible polymer for longevity of artificial hip joints10.1097/01.blo.0000246553.33434.5f ↗
16. Wear resistance of artificial hip joints with poly(2-methacryloyloxyethyl phosphorylcholine) grafted polyethylene: comparisons with the effect of polyethylene cross-linking and ceramic femoral heads10.1016/j.biomaterials.2009.02.020 ↗
17. Effects of photo-induced graft polymerization of 2-methacryloyloxyethyl phosphorylcholine on physical properties of cross-linked polyethylene in artificial hip joints10.1007/s10856-007-3063-0 ↗
18. Long-term hip simulator testing of the artificial hip joint bearing surface grafted with biocompatible phospholipid polymer10.1002/jor.22526 ↗
19. Wear resistance of the biocompatible phospholipid polymer-grafted highly cross-linked polyethylene liner against larger femoral head10.1002/jor.22868 ↗
20. Effect of head surface roughness and sterilization on wear of UHMWPE acetabular cups10.1002/jbm.a.32161 ↗
21. Wear of gamma-crosslinked polyethylene acetabular cups against roughened femoral balls10.1097/00003086-199912000-00008 ↗
22. Wear of conventional and cross-linked ultra-high-molecular-weight polyethylene acetabular cups against polished and roughened CoCr femoral heads in a biaxial hip simulator10.1002/jbm.10471 ↗
23. In vivo femoral head damage and its effect on polyethylene wear10.1016/j.arth.2009.01.010 ↗
24. The effect of surface topography of retrieved femoral heads on the wear of UHMWPE sockets10.1016/S1350-4533(97)00032-5 ↗
25. Femoral prosthetic heads and their significance in polyethylene wear10.1007/s002640000129 ↗
26. Identification of scratch mechanisms on a retrieved metallic femoral head10.1016/j.wear.2004.09.035 ↗
27. Cartilage-mimicking, high-density brush structure improves wear resistance of crosslinked polyethylene: a pilot study10.1007/s11999-010-1718-5 ↗
28. Poor survival of cementless Biomet total hip: a report on 1,047 hips from the Finnish Arthroplasty Register10.3109/17453679909000975 ↗
29. Long-term results of the threaded Mecron cup in primary total hip arthroplasty : a 15–20-year follow-up study10.1007/s00264-009-0843-9 ↗
30. Acetabular screw rings and surface treatment10.1097/00003086-199707000-00017 ↗
31. Bipolar hip arthroplasty10.1016/j.arth.2011.03.003 ↗
32. The predictive power of surface profile parameters on the amount of wear measured in vitro on metal-on-polyethylene artificial hip joints10.1243/09544119JEIM95 ↗
33. Force actions transmitted by joints in the human body10.1098/rspb.1976.0004 ↗
34. Surface analysis of early retrieved acetabular polyethylene liners: a comparison of conventional and highly crosslinked polyethylenes10.1016/j.arth.2003.08.003 ↗
35. Isolation and characterization of UHMWPE wear particles down to ten nanometers in size from in vitro hip and knee joint simulators10.1002/jbm.a.30824 ↗
36. Surface grafting of biocompatible phospholipid polymer MPC provides wear resistance of tibial polyethylene insert in artificial knee joints10.1016/j.joca.2010.05.019 ↗
37. The lubricating ability of biomembrane models with dipalmitoyl phosphatidylcholine and gamma-globulin10.1243/0954411981534114 ↗
38. Influences of dehydration and rehydration on the lubrication properties of phospholipid polymer-grafted cross-linked polyethylene10.1177/0954411915588969 ↗
39. Study of the distribution of the molecular orientation in thick polyethylene samples by X-ray-diffraction, infrared dichroism and raman-spectroscopy10.1016/0032-3861(93)90245-6 ↗
40. Effect of counterface material on the characteristics of retrieved uncemented cobalt-chromium and titanium alloy total hip replacements10.1243/PIME_PROC_1996_210_412_02 ↗
41. A novel method of cross-linking ultra-high-molecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties. Recipient of the HAP Paul Award
42. The basis for a second-generation highly cross-linked UHMWPE10.1097/01.blo.0000238856.61862.7d ↗
43. Wear resistance and mechanical properties of highly cross-linked, ultrahigh-molecular weight polyethylene doped with vitamin E10.1016/j.arth.2005.07.009 ↗
44. Wear versus thickness and other features of 5-Mrad crosslinked UHMWPE acetabular liners10.1007/s11999-010-1555-6 ↗
45. A phospholipid polymer graft layer affords high resistance for wear and oxidation under load bearing conditions10.1016/j.jmbbm.2017.12.036 ↗
46. Early osteolysis with Hylamer acetabular liners10.1016/S0883-5403(98)90016-X ↗