journal article·2023-03

Damage analysis of retrieved BioloxⓇdelta components used in hard and soft bearings

Joachim Döring, Adrian Buchholz, Maria Herbster, Jennifer Gehring, Ulf Betke, Paweł Chodór, Jan Zabrzyński, Jessica Bertrand, Christoph H. Lohmann, Łukasz Łapaj

Acta Biomaterialia· Volume 158
Publisher page

Abstract

This retrieval study included 43 Biolox delta explants (18 CoC, 25 CoP). Implants were examined macroscopically, whereby damage was evaluated using a semi quantitative scoring system. Confocal microscopy was used to examine wear related damage patterns of the articulating surfaces. Scanning electron microscopy (SEM) with energy-dispersive X-ray spectroscopy (EDS) was used to analyze wear marks on the implant surface and wear debris in periprosthetic tissue samples. Raman spectroscopy and X-ray diffraction (XRD) were used to quantify monoclinic zirconia fractions. On all components, in vivo wear resulted predominantly in different damage patterns caused by metal transfer. In CoC bearings stripe wear was additionally detected, and some implants underwent severe damage due to component breakage. The wear scores were higher for CoC components, with no differences between the scores for CoC heads and liners. Wear features caused comparable roughening on implants from CoC and CoP bearings. SEM studies demonstrated that most wear marks were caused by metal debris released from implant components. Grain pull-out was observed in stripe wear regions. Monoclinic phase shift was observed in a similar quantity on components from CoP and CoC bearings. The increase of monoclinic zirconia content around metal deposits was minimal and was more pronounced in areas of stripe wear. The results of this study indicate, that ZTA components in general undergo minimal wear in both, CoC and CoP bearings, however, it is more pronounced in the former. Metal deposits, as the most common wear feature, have no significant effect on monoclinic phase transition. STATEMENT OF SIGNIFICANCE: In this paper, we classify all damage patterns macroscopically according to an established scoring system and assess them regarding surface roughness (confocal microscopy) and monoclinic phase content (Raman spectroscopy) in order to derive the severity for patients. We compare hard-hard and hard-soft bearings and relate damage patterns with metal transfer based on SEM/EDS examinations. Furthermore, we work out correlations between patient-specific data, cause of revision and the physical condition of each individual sample Our cohort consists of 43 Biolox delta retrievals, a comparatively large quantity. In addition, we address current topics such as metal transfer and, based on the classification of damage patterns, provide incentives and/or meaningful focal points for further research.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Bone Tissue Engineering Materials · Dental materials and restorations

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (67)

Show source-provided references
1. The operation of the century: total hip replacement10.1016/S0140-6736(07)60457-7 ↗
2. EPRD Deutsche Endoprothesenregister gGmbH, EPRD-Jahresbericht 2021, 2021.
3. Retrieval analysis of three generations of Biolox® femoral heads: spectroscopic and SEM characterisation10.1016/j.jmbbm.2012.04.003 ↗
4. Wear mechanisms in ceramic hip implants
5. Raman and fluorescence investigations on retrieved Biolox® delta femoral heads10.1002/jrs.4105 ↗
6. Zirconia as a ceramic biomaterial10.1016/S0142-9612(98)00010-6 ↗
7. Transformation toughening: part 1: size effects associated with the thermodynamics of constrained transformations10.1007/BF00809057 ↗
8. Transformation toughening: part 2: contribution to fracture toughness10.1007/BF00809058 ↗
9. Transformation toughening: part 4: fabrication, fracture toughness and strength10.1007/BF00809060 ↗
10. Medical Products Division, BIOLOX®delta - Nanocomposite for Arthroplasty: The Fourth Generation of Ceramics. Scientific Information and Performance Data, Germany, 2011.
11. Zirconia-toughened alumina (ZTA) ceramics10.1007/BF02385721 ↗
12. Does metal transfer affect the tribological behaviour of femoral heads? Roughness and phase transformation analyses on retrieved zirconia and Biolox® Delta composites10.1016/j.compositesb.2016.02.020 ↗
13. The significance of metal staining on alumina femoral heads in total hip arthroplasty10.1016/j.arth.2006.02.155 ↗
14. Zirconia phase transformation, metal transfer, and surface roughness in retrieved ceramic composite femoral heads in total hip arthroplasty10.1016/j.arth.2014.08.011 ↗
15. On the tribological behavior of retrieved hip femoral heads affected by metallic debris. A comparative investigation by stylus and optical profilometer for a new roughness measurement protocol10.1016/j.measurement.2016.05.003 ↗
16. Raman and photoemission spectroscopic analyses of explanted biolox® delta femoral heads showing metal transfer
17. Retrieval analysis of squeaking ceramic implants: are there related specific features?10.1016/j.otsr.2011.12.003 ↗
18. Raman spectroscopic analysis of advanced ceramic composite for hip prosthesis10.1111/j.1551-2916.2007.01507.x ↗
19. Detailed inspection of metal implants10.5301/hipint.5000215 ↗
20. Goldstein · 1975
21. Direct determination by Raman microprobe of the transformation zone size in Y2O3 containing tetragonal ZrO2 polycrystals
22. Measurement of the crystallographically transformed zone produced by fracture in ceramics containing tetragonal zirconia10.1111/j.1151-2916.1982.tb10445.x ↗
23. J.A. Muñoz Tabares, M.J. Anglada, Quantitative Analysis of Monoclinic Phase in 3Y-TZP by Raman Spectroscopy, 2020.
24. Is surface metastability of today's ceramic bearings a clinical issue?10.3390/jcs5100273 ↗
25. 2004 · 491
26. L. Spieß, G. Teichert, R. Schwarzer, H. Behnken, C. Genzel, Moderne Röntgenbeugung, Springer Fachmedien Wiesbaden, Wiesbaden, 2019.10.1007/978-3-8348-8232-5 ↗
27. A profile refinement method for nuclear and magnetic structures10.1107/S0021889869006558 ↗
28. TOPAS and TOPAS-Academic an optimization program integrating computer algebra and crystallographic objects written in C++10.1107/S1600576718000183 ↗
29. Crystallography open database - an open-access collection of crystal structures10.1107/S0021889809016690 ↗
30. Zirconia phase transformation in zirconia-toughened alumina ceramic femoral heads: an implant retrieval analysis10.1016/j.arth.2019.07.014 ↗
31. Fracture of a BIOLOX delta ceramic femoral head: a case report and update of the literature10.2106/JBJS.CC.18.00336 ↗
32. Reconciling in vivo and in vitro kinetics of the polymorphic transformation in zirconia-toughened alumina for hip joints: I. Phenomenology10.1016/j.msec.2016.11.059 ↗
33. Reconciling in vivo and in vitro kinetics of the polymorphic transformation in zirconia-toughened alumina for hip joints: III. Molecular scale mechanisms10.1016/j.msec.2016.11.007 ↗
34. Reconciling in vivo and in vitro kinetics of the polymorphic transformation in zirconia-toughened alumina for hip joints: II. Theory10.1016/j.msec.2016.10.031 ↗
35. Metal transfer evaluation on ceramic biocomponents: a protocol based on 3D scanners10.1016/j.measurement.2020.108574 ↗
36. ISO - International Organization for Standardization, 7206-2 Implants for surgery — Partial and total hip joint prostheses —: Part 2: Articulating surfaces made of metallic, ceramic and plastics materials, 2011.
37. Does metal transfer differ on retrieved Biolox ® Delta composites femoral heads? Surface investigation on three Biolox ® generations from a biotribological point of view10.1016/j.compositesb.2017.01.026 ↗
38. Phase transformation and roughening in artificially aged and retrieved zirconia-toughened alumina femoral heads10.1016/j.arth.2018.12.025 ↗
39. Influence of metallic deposition on ceramic femoral heads on the wear behavior of artificial hip joints: a simulator study
40. Experimental investigation of material transfer on bearings for total hip arthroplasty—a retrieval study on ceramic and metallic femoral heads10.3390/jcm11143946 ↗
41. The noisy ceramic hip: is component malpositioning the cause?10.1016/j.arth.2008.04.001 ↗
42. Metal transfer on a ceramic head with a single rim contact10.1016/j.arth.2011.02.025 ↗
43. DIN Deutsches Institut für Normung e.V., Metallische Werkstoffe: Teil 3: Titan aluminium-6 Vanadium-4 Knetlegierung, 2000, pp. 111–138.
44. Combined wear and ageing of ceramic-on-ceramic bearings in total hip replacement under edge loading conditions10.1016/j.jmbbm.2019.05.011 ↗
45. Stripe wear and squeaking in ceramic total hip bearings10.1053/j.sart.2006.09.015 ↗
46. Hip-simulator wear studies of an alumina-matrix composite (AMC) ceramic compared to retrieval studies of AMC balls with 1–7 years follow-up10.1016/j.wear.2009.02.018 ↗
47. Chemically driven tetragonal-to-monoclinic polymorphic transformation in retrieved ZTA femoral heads from dual mobility hip implants10.1016/j.jmbbm.2015.11.032 ↗
48. In vitro versus in vivo phase instability of zirconia-toughened alumina femoral heads: a critical comparative assessment10.3390/ma10050466 ↗
49. Effects of in vitro shocks and hydrothermal degradation on wear of ceramic hip joints: towards better experimental simulation of in vivo ageing10.1016/j.triboint.2016.05.010 ↗
50. Delta ceramic-on-alumina ceramic articulation in primary THA: prospective, randomized FDA-IDE study and retrieval analysis10.1007/s11999-009-1143-9 ↗
51. Effect of the composition and sintering process on mechanical properties and residual stresses in zirconia–alumina composites10.1016/j.jeurceramsoc.2004.09.025 ↗
52. DIELECTRIC Manufacturing, UHWM PE Characteristics. https://dielectricmfg.com/knowledge-base/uhmw/(accessed 15.05.22).
53. Bearing surfaces in primary total hip arthroplasty10.1302/2058-5241.3.180300 ↗
54. Ceramic-on-ceramic vs ceramic-on-polyethylene, a comparative study with 10-year follow-up10.5312/wjo.v12.i1.14 ↗
55. The ideal total hip replacement bearing surface in the young patient: a prospective randomized trial comparing alumina ceramic-on-ceramic with ceramic-on-conventional polyethylene: 15-year follow-up10.1016/j.arth.2017.11.066 ↗
56. Does Biolox Delta ceramic reduce the rate of component fractures in total hip replacement?10.1016/j.otsr.2014.05.010 ↗
57. Catastrophic intraoperative failure of a ceramic femoral head10.1016/j.artd.2021.08.017 ↗
58. A focused update on preventing ceramic fractures in hip arthroplasty: is the ‘cup’ half full?10.1302/0301-620X.101B8.BJJ-2019-0309.R1 ↗
59. Fracture of a BIOLOX delta ceramic femoral head: a case report and update of the literature10.2106/JBJS.CC.18.00336 ↗
60. Incidence, risk factors, and outcome of ceramic-on-ceramic bearing breakage in total hip arthroplasty10.1016/j.arth.2021.03.021 ↗
61. Fracture of ceramic bearing surfaces following total hip replacement: a systematic review10.1155/2013/157247 ↗
62. Preventing ceramic liner fracture after Delta ceramic-on-ceramic total hip arthroplasty10.1007/s00402-020-03515-2 ↗
63. Effect of inner taper angle of acetabular metal shell on the malseating and dissociation force of ceramic liner10.1016/j.arth.2016.11.002 ↗
64. Clinical outcomes and bearing-specific complications following fourth-generation alumina ceramic-on-ceramic total hip arthroplasty: a single-surgeon series of 749 hips at a minimum of 5-year follow-up10.1016/j.arth.2018.02.045 ↗
65. Impingement between the metal stem and the ceramic liner in total hip arthroplasty: the neck is not the only place where impingement occurs10.2106/JBJS.20.00354 ↗
66. Surgeons' awareness and impaction technique of a ceramic liner into a metal shell10.4055/cios21102 ↗
67. Mid-term results of the BIOLOX delta ceramic-on-ceramic total hip arthroplasty10.1302/0301-620X.99B6.BJJ-2016-0486.R3 ↗