Recovery of low volumes of wear debris from rat stifle joint tissues using a novel particle isolation method

J. Patel, S. Lal, K. Nuss, S.P. Wilshaw, B. von Rechenberg, R.M. Hall, J.L. Tipper

Acta Biomaterialia· Volume 71

Abstract

Less than optimal particle isolation techniques have impeded analysis of orthopaedic wear debris in vivo. The purpose of this research was to develop and test an improved method for particle isolation from tissue. A volume of 0.018 mm3 of clinically relevant CoCrMo, Ti-6Al-4V or Si3N4 particles was injected into rat stifle joints for seven days of in vivo exposure. Following sacrifice, particles were located within tissues using histology. The particles were recovered by enzymatic digestion of periarticular tissue with papain and proteinase K, followed by ultracentrifugation using a sodium polytungstate density gradient. Particles were recovered from all samples, observed using SEM and the particle composition was verified using EDX, which demonstrated that all isolated particles were free from contamination. Particle size, aspect ratio and circularity were measured using image analysis software. There were no significant changes to the measured parameters of CoCrMo or Si3N4 particles before and after the recovery process (KS tests, p > 0.05). Titanium particles were too few before and after isolation to analyse statistically, though size and morphologies were similar. Overall the method demonstrated a significant improvement to current particle isolation methods from tissue in terms of sensitivity and efficacy at removal of protein, and has the potential to be used for the isolation of ultra-low wearing total joint replacement materials from periprosthetic tissues. This research presents a novel method for the isolation of wear particles from tissue. Methodology outlined in this work would be a valuable resource for future researchers wishing to isolate particles from tissues, either as part of preclinical testing, or from explants from patients for diagnostic purposes. It is increasingly recognised that analysis of wear particles is critical to evaluating the safety of an orthopaedic device.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Orthopedic Infections and Treatments · Polymer Surface Interaction Studies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (52)

Show source-provided references
1. NJR. 13th Annual Report. 2016.
2. The OPG/RANKL/RANK system10.1210/endo.142.12.8536 ↗
3. Effects of polyethylene and TiAlV wear particles on expression of RANK, RANKL and OPG mRNA10.1080/00016470410001222 ↗
4. High incidence of pseudotumour formation after large-diameter metal-on-metal total hip replacement a prospective cohort study10.1302/0301-620X.94B6.28373 ↗
5. FDA, Use of International Standard ISO 10993-1, “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process”, Guidance for Industry and Food and Drug Administration Staff, in: FDA (Ed.), 2016.
6. Wear Debris characterization and corresponding biological response: artificial hip and knee joints10.3390/ma7020980 ↗
7. Isolation of predominantly submicron-sized Uhmwpe wear particles from periprosthetic tissues10.1002/jbm.820290118 ↗
8. The size and shape of particulate polyethylene wear debris in total joint replacements10.1243/0954411971534638 ↗
9. Standardized analysis of UHMWPE wear particles from failed total joint arthroplasties10.1002/jbm.10302 ↗
10. Analysis of polyethylene wear debris using micro-Raman spectroscopy: a report on the presence of beta-carotene10.1002/(SICI)1097-4636(199704)35:1<31::AID-JBM4>3.0.CO;2-N ↗
11. Separation and characterization of polyethylene wear debris from synovial fluid and tissue samples of revised knee replacements10.1002/(SICI)1097-4636(199701)34:1<57::AID-JBM8>3.0.CO;2-M ↗
12. Composition and morphology of wear debris in failed uncemented total hip-replacement10.1302/0301-620X.76B1.8300684 ↗
13. Quantitative analysis of polyethylene wear debris, wear rate and head damage in retrieved Charnley hip prostheses10.1023/A:1008901302646 ↗
14. Quantitative characterization of polyethylene debris isolated from periprosthetic tissue in early failure knee implants and early and late failure Charnley hip implants10.1002/jbm.1036 ↗
15. Isolation and characterization of debris in membranes around total joint prostheses10.2106/00004623-199411000-00010 ↗
16. Isolation, characterization and quantification of polyethylene wear debris from periprosthetic tissues around total joint replacements10.1016/j.wear.2006.11.005 ↗
17. Characterization and comparison of wear debris from failed total hip implants of different types10.2106/00004623-199608000-00014 ↗
18. Characterization of in vivo wear debris from ceramic-ceramic total hip arthroplasties10.1002/(SICI)1097-4636(199612)32:4<627::AID-JBM16>3.0.CO;2-A ↗
19. Wear debris from two different alumina-on-alumina total hip arthroplasties10.1302/0301-620X.82B6.0820901 ↗
20. Debris from failed ceramic-on-ceramic and ceramic-on-polyethylene hip prostheses10.1097/00003086-200108000-00017 ↗
21. Comparison of periprosthetic tissue digestion methods for ultra-high molecular weight polyethylene wear debris extraction10.1002/jbm.b.31416 ↗
22. Effects of digestion protocols on the isolation and characterization of metal-metal wear particles. II. analysis of ion release and particle composition10.1002/1097-4636(20010605)55:3<330::AID-JBM1021>3.0.CO;2-Y ↗
23. Metal wear particle characterization from metal on metal total hip replacements: transmission electron microscopy study of periprosthetic tissues and isolated particles10.1002/(SICI)1097-4636(199810)42:1<103::AID-JBM13>3.0.CO;2-M ↗
24. Characterisation of wear particles produced by metal on metal and ceramic on metal hip prostheses under standard and microseparation simulation10.1007/s10856-006-0015-z ↗
25. In vivo production of nanosized metal wear debris formed by tribochemical reaction as confirmed by high-resolution TEM and XPS analyses10.1002/jbm.a.32301 ↗
26. ISO 17853 · 2011
27. The John Charnley award: an accurate and extremely sensitive method to separate, display, and characterize wear debris part 2: metal and ceramic particles10.1007/s11999-011-2058-9 ↗
28. A novel method for isolation and recovery of ceramic nanoparticles and metal wear debris from serum lubricants at ultra-low wear rates10.1016/j.actbio.2016.07.004 ↗
29. Comparison of the cytotoxicity of clinically relevant cobalt-chromium and alumina ceramic wear particles in vitro10.1016/S0142-9612(02)00360-5 ↗
30. Directive 2010/63/EU of The European Parliament and of the council of 22 on the protection of animals used for scientific purposes
31. ASTM F1537 · 2011
32. J. Patel, S. Lal, S.P. Wilshaw, R.M. Hall, J.L. Tipper, Recovery rate of particle isolation using sodium polytungstate density gradients, Data in Brief, 2017, Under review.10.1016/j.dib.2018.06.019 ↗
33. J. Patel, S. Lal, S.P. Wilshaw, R.M. Hall, J.L. Tipper, Development and optimisation of a tissue digestion method for the isolation of orthopaedic wear particles, Data in Brief, 2017, Under review.10.1016/j.dib.2018.07.060 ↗
34. J. Patel, S. Lal, S.P. Wilshaw, R.M. Hall, J.L. Tipper, Validation of a novel particle isolation procedure using particle doped tissue samples, Data in Brief, 2017, Under review.10.1016/j.dib.2018.04.096 ↗
35. ASTM F1877 · 2016
36. Wear of historical polyethylenes in hip prostheses. Biomechanical success and a biological failure10.1177/112070000301302s05 ↗
37. Analysis of retrieved alumina ceramic components from Mittelmeier total hip prostheses10.1016/S0142-9612(99)00081-2 ↗
38. Effects of clinically relevant alumina ceramic wear particles on TNF-alpha production by human peripheral blood mononuclear phagocytes10.1016/S0142-9612(02)00510-0 ↗
39. Consequences of exposure to peri-articular injections of micro- and nano-particulate cobalt-chromium alloy10.1016/j.biomaterials.2013.07.073 ↗
40. Periprosthetic tissues from third generation alumina-on-alumina total hip arthroplasties10.1016/j.arth.2012.10.021 ↗
41. Detection of diatom in formalin-fixed tissue by proteinase K digestion10.1016/j.forsciint.2009.05.005 ↗
42. Size, shape, and composition of wear particles from metal-metal hip simulator testing: effects of alloy and number of loading cycles10.1002/jbm.a.10088 ↗
43. Nanoparticle interaction with plasma proteins as it relates to particle biodistribution, biocompatibility and therapeutic efficacy10.1016/j.addr.2009.03.009 ↗
44. Characterisation of wear debris from UHMWPE on zirconia ceramic, metal-on-metal and alumina ceramic-on-ceramic hip prostheses generated in a physiological anatomical hip joint simulator10.1016/S0043-1648(01)00653-6 ↗
45. Quantitative analysis of wear and wear debris from metal-on-metal hip prostheses tested In a physiological hip joint simulator
46. Wear particles and ions from cemented and uncemented titanium-based hip prostheses-a histological and chemical analysis of retrieval material10.1002/jbm.b.33243 ↗
47. Morphology and dissolution rate of wear debris from silicon nitride coatings10.1021/acsbiomaterials.6b00133 ↗
48. The biological response to nanometre-sized polymer particles10.1016/j.actbio.2015.05.016 ↗
49. Determining the size and shape dependence of gold nanoparticle uptake into mammalian cells10.1021/nl052396o ↗
50. Diverse cellular and apoptotic responses to variant shapes of UHMWPE particles in a murine model of inflammation10.1016/S0142-9612(02)00032-7 ↗
51. Unambiguous observation of shape effects on cellular fate of nanoparticles10.1038/srep04495 ↗
52. Time evolution of the nanoparticle protein corona10.1021/nn901372t ↗