Surface modifications and oxidative degradation in MPC-grafted highly cross-linked polyethylene liners retrieved from short-term total hip arthroplasty
Source abstract
Abstract
Highly cross-linked polyethylene (HXLPE) hip liners grafted with 2-methacryloyloxyethyl phosphorylcholine (MPC) on their bearing surfaces have recently been commercialized as components of a new generation of artificial hip joints, while improvements in wear resistance and biocompatibility were reported based on in vitro studies. The present study aimed at evaluating the surface modification and oxidative degradation in short-term retrieved MPC-grafted liners by X-ray photoelectron spectroscopy (XPS), Fourier-transform infrared spectroscopy (FT-IR) with attenuated total reflection (ATR) equipment and Raman spectroscopy. In none of 3 samples of retrieved MPC-grafted liners, detectable MPC graft remained on the bearing surfaces although 2 samples yet contained remains of MPC polymer in their rim zone. These results revealed that the MPC polymer might have quickly disappeared from the bearing surface under in vivo loading, which is more severe than the in vitro one. Furthermore, a detectable oxidation index (OI) value (>0.1) was not only observed in any zone of any sample investigated, but also in the rim zones of Samples 1 and 2, which surprisingly experienced the most remarkable increase in OI value detected in this study. We thus confirmed that: (i) annealing of HXLPE cannot completely remove free radicals; (ii) the MPC graft has no beneficial effect in protecting HXLPE against oxidation and wear; and, (iii) lipid absorption occurred even in the rim zone where the MPC layer remained. Based on these evidences we consider that the declaimed advanced MPC technology is not a suitable one to elongate the in vivo lifetime of hip joints. Statement of significance Several studies reported that highly crosslinked polyethylene (HXLPE) have resulted in reduced wear in total hip arthroplasty. Beyond those studies, HXLPE hip liners grafted with 2-methacryloyloxyethyl phosphorylcholine (MPC) on their bearing surface were extensively studied in vitro and then commercialized as a new generation of artificial hip joints. The present study reports for the first time results about the evaluation of surface modification and oxidative degradation in retrieved the MPC grafted liners. The findings of this investigation clearly show that the MPC layer has been peeled off on the bearing surface of the liner main wear zone although the MPC layer remained on the surface of the rim zones. Furthermore, we assessed the microstructural modifications and the oxidation drifts that occurred in vivo in the hip joints despite the presence of the MPC layer.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Polymer Surface Interaction Studies
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (26)
Show source-provided references
1. Five-year experience with Crossfire highly cross-linked polyethylene10.1097/00003086-200512000-00024 ↗
2. Clinical performance of a highly cross-linked polyethylene at two years in total hip arthroplasty: a randomized prospective trial10.1016/S0883-5403(03)00341-3 ↗
3. Five-year comparative study of highly crosslinked (crossfire) and traditional polyethylene10.1016/j.arth.2007.09.015 ↗
4. Does highly cross linked polyethylene wear less than conventional polyethylene in total hip arthroplasty10.1016/j.arth.2006.12.117 ↗
5. The Otto Aufranc Award. Highly cross-linked polyethylene in total hip arthroplasty: randomized evaluation of penetration rate in cemented and uncemented sockets using radiostereometric analysis10.1097/01.blo.0000150314.70919.e3 ↗
6. In vivo oxidation in remelted highly cross-linked retrievals10.2106/JBJS.I.01006 ↗
7. Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups10.2106/00004623-200709000-00019 ↗
8. Investigation of surgically retrieved, vitamin E-stabilized, crosslinked UHMWPE implants after short-term in vivo service10.1002/jbm.b.33465 ↗
9. Retrieval analysis of sequentially annealed highly crosslinked polyethylene used in total hip arthroplasty10.1007/s11999-014-4113-9 ↗
10. Surface grafting of artificial joints with a biocompatible polymer for preventing periprosthetic osteolysis10.1038/nmat1233 ↗
11. Long-term hip simulator testing of the artificial hip joint bearing surface grafted with biocompatible phospholipid polymer10.1002/jor.22526 ↗
12. Preparation of phospholipid polymers and their properties as polymer hydrogel membranes10.1295/polymj.22.355 ↗
13. A hydrated phospholipid polymer-grafted layer prevents lipid-related oxidative degradation of cross-linked polyethylene10.1016/j.biomaterials.2016.10.021 ↗
14. Enhanced wear resistance of modified cross-linked polyethylene by grafting with poly (2-methacryloyloxyethyl phosphorylcholine)10.1002/jbm.a.31134 ↗
15. ASTM Standard, F2102-13. Standard guide for evaluating the extent of oxidation in ultra-high-molecular-weight polyethylene fabricated forms intended for surgical implants. West Conshohocken, PA, USA: ASTM International; 2013.
16. Raman spectroscopic method for determining the crystallinity of polyethylene10.1002/pol.1978.180160704 ↗
17. Estimation of crystallinity in polyethylene by Raman spectroscopy10.1002/jrs.1250240813 ↗
18. Raman spectroscopy employed for the determination of the intermediate phase in polyethylene10.1021/ma00112a050 ↗
19. Wear resistance of the biocompatible phospholipid polymer-grafted highly cross-linked polyethylene liner against larger femoral head10.1002/jor.22868 ↗
20. Wear testing of total hip replacements under severe conditions10.1586/17434440.2015.1050378 ↗
21. History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene10.1007/s11999-011-1872-4 ↗
22. Oxidative properties and surface damage mechanisms of remelted highly crosslinked polyethylenes in total knee arthroplasty10.1007/s00264-013-1796-6 ↗
23. In vivo oxidation of retrieved cross-linked ultra-high-molecular-weight polyethylene acetabular components with residual free radicals10.1016/j.arth.2005.07.019 ↗
24. In vivo oxidation of gamma-barrier-sterilized ultra-high-molecular-weight polyethylene bearings10.1016/j.arth.2006.07.006 ↗
25. Ex vivo stability loss of irradiated and melted ultra-high molecular weight polyethylene10.2106/JBJS.I.01017 ↗
26. Large deformation compression induced crystallinity degradation of conventional and highly crosslinked UHMWPEs10.1016/j.biomaterials.2005.04.008 ↗