Backside wear, particle migration and effectiveness of screw hole plugs in acetabular hip joint replacement with cross-linked polyethylene

Steffen Braun, Shirin Vardag, Ulrike Mueller, Stefan Schroeder, Robert Sonntag, Therese Bormann, Tobias Gotterbarm, Jan Philippe Kretzer

Acta Biomaterialia· Volume 97
Publisher page

Abstract

In total hip arthroplasty, osteolysis of the acetabulum often occurs at the backside of cups in the area of screw holes, indicating a clinically relevant amount of polyethylene (PE) wear particles in this area. In order to avoid a possible migration of wear particles to the acetabulum-bone, screw hole plugs are provided for some implant systems. The aims of this study were to quantitatively determine backside wear and to investigate the migration behaviour of articulation-related wear particles in a cup system with open and closed screw holes by plugs. Titanium cup systems with backside holes for screw fixations were sinusoidally loaded with 2.7 kN. The articulation area was separated from the backside area of the cup. A defined amount of articulation-generated particles was added to the fluid of the articulation chamber. The fluids in the two chambers were separately filtered after 2 × 106 cycles for a particle analysis. Backside wear with noticeably small (65.6 ± 4.2 nm) and round PE particles was identified. With both open and closed screw holes, a migration of the articulating wear particles from the articulation area behind the cup could be observed. Backside wear was estimated to be below 1% of the articulated wear. Screw hole plugs did not effectively prevent the migration of PE wear particles behind the investigated cups. STATEMENT OF SIGNIFICANCE: Backside wear occurs in a proven cup-system. Furthermore, it was quantitatively observed that articulation-generated wear products could migrate from the articulating area along the cup/liner-interface through the screw holes behind the cup. An almost unimpeded particle migration to the acetabulum-bone, in conjunction with very small backside wear particles, could produce a clinically relevant amount of PE with respect to pelvic lysis. These findings highlight the importance of management to avoid particle migration in artificial hip cups. Therefore, primarily the use of screw hole plugs, as far as available for the respective cup-system, is recommended. The aim of avoiding particle migration by plugs, but also by using a sophisticated anchoring mechanism between cup and PE liner should continue in future.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Hip disorders and treatments · Advanced materials and composites

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (44)

Show source-provided references
1. ISO 14242-1, Implants for surgery – Wear of total hip-joint prosthese – Part1: Loading and displacement parameters for wear-testing machines and corresponding environment conditions for test: Geneva, Switzerland: International Organization of Standardization., (2014).
2. ISO 14242-2, Implants for surgery – Wear of total hip-joint protheses – Part 2: Methods of measurent. Geneva, Switzerland: International Organisation of Standardization (2016).
3. Wear performance of neat and vitamin E blended highly cross-linked PE under severe conditions: the combined effect of accelerated ageing and third body particles during wear test10.1016/j.jmbbm.2016.08.003 ↗
4. Low wear rates seen in THAs with highly crosslinked polyethylene at 9 to 14 years in patients younger than age 50 years10.1007/s11999-015-4422-7 ↗
5. Wear comparison between conventional and highly cross-linked polyethylene against a zirconia head: a concise follow-up, at an average 10 years, of a previous report10.1016/j.arth.2012.12.020 ↗
6. Biologic effects of implant debris
7. The cellular and molecular biology of periprosthetic osteolysis10.1097/01.blo.0000238813.95035.1b ↗
8. Periprosthetic osteolysis after total hip replacement: molecular pathology and clinical management10.1007/s10787-013-0192-6 ↗
9. A literature review of the association between wear rate and osteolysis in total hip arthroplasty10.1054/arth.2002.33664 ↗
10. Wear and periprosthetic osteolysis: the problem
11. J. Kaerrholm, M. Mohaddes, D. Odin, C. Rogmark, O. Rolfson, Swedish Hip Arthroplasty Register – Annual Report 2017, (2017).
12. A. Grimberg, V. Jansson, T. Liebs, O. Melsheimer, A. Siebert, Endoprothesenregister Deutschland (EPRD) – Jahresbericht 2017, (2017).
13. S. Graves, R. de Steiger, L. Lewis, I. Harris, Hip, Knee & Shoulder Arthroplasty – Annual Report 2018, Australian Orthopaedic Association National Joint Replacement Registry. (2018).
14. Polyethylene and metal debris generated by non-articulating surfaces of modular acetabular components10.1302/0301-620X.76B4.8027142 ↗
15. Articulation and backside wear analysis after long-term in vitro wear simulation of vitamin e stabilized polyethylene acetabular liners with a press-fit locking mechanism10.21823/2311-2905-2018-24-2-29-40 ↗
16. Backside wear is not dependent on the acetabular socket design in crosslinked polyethylene liners
17. Minimal backside surface changes observed in retrieved acetabular liners10.1016/j.arth.2010.07.012 ↗
18. Backside wear is low in retrieved modern, modular, and nonmodular acetabular liners10.1097/01.blo.0000177719.16146.33 ↗
19. The acetabular insert-metal backing interface: an additional source of polyethylene wear debris10.1016/j.arth.2005.04.033 ↗
20. Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses10.1002/jbm.820170510 ↗
21. Backside wear in acetabular hip joint replacement10.1016/j.actbio.2018.10.045 ↗
22. Backside volumetric change in the polyethylene of uncemented acetabular components10.1302/0301-620X.91B8.21850 ↗
23. Simulation of initial frontside and backside wear rates in a modular acetabular component with multiple screw holes10.1016/S0021-9290(99)00043-3 ↗
24. Periprosthetic wear particle migration and distribution modelling and the implication for osteolysis in cementless total hip replacement10.1016/j.jmbbm.2014.01.005 ↗
25. The relationship between the design, position, and articular wear of acetabular components inserted without cement and the development of pelvic osteolysis10.2106/00004623-199405000-00008 ↗
26. Periprosthetic bone loss in total hip arthroplasty. Polyethylene wear debris and the concept of the effective joint space10.2106/00004623-199274060-00006 ↗
27. Extensive osteolysis caused by polyethylene particle migration in an anatomical hydroxyapatite-coated hip prosthesis: 10 years' follow-up10.1016/j.arth.2009.08.021 ↗
28. Effect of locking mechanism on fluid and particle flow through modular acetabular components10.1016/S0883-5403(98)90169-3 ↗
29. Osteolysis around uncemented acetabular components of cobalt-chrome surface replacement hip arthroplasty
30. Highly Cross-Linked Polyethylene Reduces Osteolysis Incidence and Wear-Related Reoperation Rate in Cementless Total Hip Arthroplasty Compared With Conventional Polyethylene at a Mean 12-Year Follow-Up10.1016/j.arth.2017.06.047 ↗
31. Osteoclasts in Periprosthetic Osteolysis: The Charnley Arthroplasty Revisited10.1016/j.arth.2017.05.030 ↗
32. Successful long-term fixation and progression of osteolysis associated with first-generation cementless acetabular components retrieved post mortem10.2106/JBJS.J.01507 ↗
33. Patient activity after total hip arthroplasty declines with advancing age10.1007/s11999-009-0756-3 ↗
34. ASTM F1877-16, Standard Practice for Characterization of Particles. West Conshohocken, USA: American Society for Testing and Materials, (2010).
35. Size, shape, and composition of wear particles from metal-metal hip simulator testing: effects of alloy and number of loading cycles10.1002/jbm.a.10088 ↗
36. How do gait frequency and serum-replacement interval affect polyethylene wear in knee-wear simulator tests?10.1007/s10856-014-5271-8 ↗
37. Wear testing and particle characterisation of sequentially crosslinked polyethylene acetabular liners using different femoral head sizes10.1007/s10856-013-4936-z ↗
38. Measurement outcomes from hip simulators10.1177/0954411916628559 ↗
39. New polyethylenes in total hip replacement: A ten- to 12-year follow-up study10.1302/0301-620X.95B3.29456 ↗
40. Highly crosslinked vs conventional polyethylene particles–an in vitro comparison of biologic activities10.1016/j.arth.2007.05.043 ↗
41. Highly crosslinked vs conventional polyethylene particles: relative in vivo inflammatory response10.1016/j.arth.2008.01.134 ↗
42. Pressure-induced periprosthetic osteolysis: a rat model10.1002/jor.1100180322 ↗
43. Fluid pressure causes bone resorption in a rabbit model of prosthetic loosening
44. Particle analysis for the determination of UHMWPE wear10.1002/jbm.b.30213 ↗