Mechanical load-assisted dissolution of metallic implant surfaces: Influence of contact loads and surface stress state

Andrew Mitchell, Pranav Shrotriya

Acta Biomaterialia· Volume 4· Issue 2
Publisher page

Abstract

Mechanical load-assisted dissolution is identified as one of the key mechanisms governing material removal in fretting and crevice corrosion of biomedical implants. In the current study, material removal on a stressed surface of cobalt-chromium-molybdenum (CoCrMo) subjected to single asperity contact is investigated in order to identify the influence of contact loads and in-plane stress state on surface damage mechanisms. The tip of an atomic force microscope is used as a well-characterized "asperity" to apply controlled contact forces and mechanically stimulate the loaded specimen surface in different aqueous environments from passivating to corroding. The volume of the material removed is measured to determine the influence of contact loads, in-plane stresses and the environment on the material dissolution rate. Experimental results indicate that surface damage is initiated at all the contact loads studied and as expected in a wear situation, removal rate increases with increase in contact loads. Removal rates display a complex dependence on residual stresses and the environment. In a passivating environment, the material removal rate is linearly dependent on the stress state such that surface damage is accelerated under compressive stresses and suppressed under tensile stresses. In a corrosive environment, the dissolution rate demonstrates a quadratic dependence on stress, with both compressive and tensile stresses accelerating material dissolution. A surface damage mechanism based on stress-assisted dissolution is proposed to elucidate the experimental observations.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Hydrogen embrittlement and corrosion behaviors in metals · Cardiac Structural Anomalies and Repair

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (42)

Show source-provided references
1. Concentration- and composition-dependent effects of metal ions on human MG-63 osteoblasts10.1002/jbm.10106 ↗
2. Mechanisms of failure of total hip replacements – lessons learned from retrieval studies10.1097/00003086-200403000-00003 ↗
3. Nanotribology of CoCr-UHMWPE TJR prosthesis using atomic force microscopy10.1016/S0043-1648(02)00220-X ↗
4. Focal osteolysis at the junctions of a modular stainless-steel femoral intramedullary nail10.2106/00004623-200104000-00008 ↗
5. Osteolysis: a disease of access to fixation interfaces10.1097/00003086-200212000-00015 ↗
6. Synergistic effect of particles and cyclic pressure on cytokine production in human monocyte/macrophages: proposed role in periprosthetic osteolysis10.1016/S8756-3282(01)00658-5 ↗
7. Wear Particles of polyethylene in biologicalsystems10.1080/10402009608983603 ↗
8. Gigantic popliteal synovial cyst caused by wear particles after total knee arthroplasty
9. Wear reduction effect on ultra-high-molecular-weight polyethylene by application of hard coatings and ion implantation on cobalt chromium alloy, as measured in a knee wear simulation machine10.1016/S0257-8972(01)01074-X ↗
10. Correlation of wear debris-induced osteolysis and revision with volumetric wear-rates of polyethylene – a survey of 8 reports in the literature10.1080/000164701753606644 ↗
11. Osteolytic properties of the synovial-like tissue from aseptically failed joint prostheses10.1093/rheumatology/35.10.943 ↗
12. Aseptic loosening of implants
13. In vitro analysis of the wear, wear debris and biological activity of surface-engineered coatings for use in metal-on-metal total hip replacements10.1243/095441103765212659 ↗
14. The characterization of polyethylene and metal wear particles, retrieved from tissue adjacent to failed total joint replacements using a novel extraction technique
15. Early osteolysis following total hip arthroplasty with use of a hylamer liner in combination with a modular ceramic femoral head – a case report10.2106/00004623-199910000-00009 ↗
16. Dissemination of wear particles to the liver, spleen, and abdominal lymph nodes of patients with hip or knee replacement10.2106/00004623-200004000-00002 ↗
17. Orthopedic implant fretting corrosion10.1515/CORRREV.2003.21.2-3.183 ↗
18. Fretting corrosion accelerates crevice corrosion of modular hip tapers10.1002/jab.770060104 ↗
19. Size, shape, and composition of wear particles from metal–metal hip simulator testing: effects of alloy and number of loading cycles10.1002/jbm.a.10088 ↗
20. Differences in the fretting corrosion of metal–metal and ceramic–metal modular junctions of total hip replacements10.1016/S0736-0266(03)00186-4 ↗
21. The role of fretting damage in total hip arthoplasty with modular design hip joints – evaluation of retrieval studies and experimental simulation methods
22. Fretting corrosion in saline and serum10.1002/jbm.820150404 ↗
23. In vivo and In vitro considerations of corrosion testing10.3109/10731198109117601 ↗
24. Corrosion and other electrochemical aspects of biomaterials
25. Ph Shifts and precipitation associated with metal-ions in tissue-culture10.1016/0142-9612(85)90069-9 ↗
26. Stress-enhanced ion release – the effect of static loading10.1016/0142-9612(91)90108-M ↗
27. Fretting corrosion behaviour of ball-and-socket joint on dental implants with different prosthodontic alloys
28. Fretting wear in a modular neck hip prosthesis10.1002/(SICI)1097-4636(199705)35:2<207::AID-JBM9>3.0.CO;2-R ↗
29. Design-related fretting wear in modular neck hip prosthesis10.1002/(SICI)1097-4636(199602)30:2<181::AID-JBM7>3.0.CO;2-N ↗
30. The composition of the surface passive film formed on CoCrMo alloy in simulated physiological solution10.1016/S0013-4686(03)00396-7 ↗
31. Electrochemical response of CoCrMo to high-speed fracture of its metal oxide using an electrochemical scratch test method10.1002/(SICI)1097-4636(19971205)37:3<421::AID-JBM13>3.0.CO;2-E ↗
32. In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy coupling10.1002/jbm.820271210 ↗
33. Fraker AC, editor. ASM handbook. Corrosion of metallic implants and prosthetic devices. vol. 13. p. 1324–1335.
34. Characterization of application specific probes for SPMs10.1117/12.271229 ↗
35. Prediction of scratch resistance of cobalt chromium alloy bearing surface, articulating against ultra-high molecular weight polyethylene, due to third-body wear particles10.1243/095441104322807749 ↗
36. Johnson · 1994
37. Influences of stress on the measurement of mechanical properties using nanoindentation: Part I. Experimental studies in an aluminum alloy10.1557/JMR.1996.0091 ↗
38. Atomistic simulation of the influence of pre-existing stress on the interpretation of nanoindentation data10.1557/JMR.2004.0410 ↗
39. Passive film on orthopaedic TiAlV alloy formed in physiological solution investigated by X-ray photoelectron spectroscopy10.1016/S0142-9612(00)00145-9 ↗
40. Crack deflection at an interface between dissimilar elastic-materials – role of residual-stresses10.1016/0020-7683(94)90025-6 ↗
41. Effect of residual stress in surface layer on contact deformation of elastic–plastic layered media10.1115/1.1572516 ↗
42. Evolution of a surface-roughness spectrum caused by stress in nanometer-scale chemical etching10.1103/PhysRevLett.83.3872 ↗