journal article·2021-06

Micromechanical measurement of adhesion of dehydrating silicone hydrogel contact lenses to corneal tissue

Dongkai Zhu, Yangping Liu, Jeremy L. Gilbert

Acta Biomaterialia· Volume 127
Publisher page

Abstract

Adhesion properties, which can vary with multiple factors, of silicone hydrogel contact lenses are important to their performance and comfort in the eye. In this study, we developed and used a simple, representative testing system and method to study the adhesive interactions of different silicone contact lenses (balafilcon A and senofilcon A) on polished titanium alloy and porcine whole eye cornea under dehydrating conditions. Adhesive interactions for senofilcon A varied by hydration state for both corneal and titanium adhesion, starting low, rising to a maximum and falling with dehydration time and dehydration state. Balafilcon A showed a rise and fall against titanium, but retained a relatively constant adhesive interaction with corneal tissue over dehydration time. Senofilcon A reached the highest adhesion forces (400 mN) within 5 to 10 min of testing against cornea, then dropped with time after that. Johnson-Kendall-Roberts (JKR) theory was applied to determine the surface energy of the lenses, and work of adhesion (WOA) was also determined for both lenses. Similar trends as observed with the force-hydration curves were seen with surface energy and work of adhesion as well (peak surface energy of 8 N/m and work of adhesion of 80 µJ for senofilcon A). Video imaging of the adhesive interactions showed significant corneal deformation taking place during testing, and post-test analysis shows damage to the corneal tissue. This method could be used to assess pre-clinical performance of long-lasting contact lenses and the role of hydration state. STATEMENT OF SIGNIFICANCE: Adhesion properties of contact lenses play significant roles in their performance and comfort in the eye. Adhesion is influenced by polymer chemistry, counterface materials and hydration state of the contact lenses. However, no test method has been developed to directly study the adhesion properties between contact lenses and corneal tissue during the dehydration process. Our work aims to fill this gap by developing testing and analysis methods for evaluating the adhesive interactions in vitro between contact lenses of different chemistries and properties and different counter surfaces under dehydrating conditions over time. Our study shows that adhesive interactions of contact lenses are highly dependent on polymer type, surface treatment, counterface material and hydration state.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Ocular Surface and Contact Lens · Corneal surgery and disorders · Adhesion, Friction, and Surface Interactions

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (51)

Show source-provided references
1. Silicone hydrogel contact lenses and the ocular surface.10.1016/S1542-0124(12)70262-8 ↗
2. Soft contact lens polymers: an evolution10.1016/S0142-9612(01)00165-X ↗
3. Water-dissolved-oxygen permeability coefficients of hydrogel contact lenses and boundary layer effects10.1016/S0376-7388(00)83317-7 ↗
4. Extended wear contact lenses
5. The impact of overnight wear on the risk of contact lens—associated ulcerative keratitis10.1001/archopht.1994.01090140062024 ↗
6. The relation between contact lens oxygen transmissibility and binding of Pseudomonas aeruginosa to the cornea after overnight wear10.1016/S0161-6420(94)31326-1 ↗
7. Corneal inflammatory events with daily silicone hydrogel lens wear10.1097/OPX.0000000000000105 ↗
8. Risk factors for corneal inflammatory and mechanical events with extended wear silicone hydrogel contact lenses10.1097/OPX.0b013e3181f6f97d ↗
9. Risk of corneal inflammatory events with silicone hydrogel and low dk hydrogel extended contact lens wear: a meta-analysis10.1097/OPX.0b013e3180421c47 ↗
10. Biocompatibility in the development of silicone-hydrogel lenses10.1097/ICL.0b013e31827dbb00 ↗
11. The impact of silicone hydrogel materials on overnight corneal swelling10.1097/01.icl.0000242166.09513.aa ↗
12. Biocompatibility in the development of silicone-hydrogel lenses10.1097/ICL.0b013e31827dbb00 ↗
13. Anatomy and physiology of the cornea10.1016/j.jcrs.2010.12.037 ↗
14. Influence of environmental factors in the in vitro dehydration of hydrogel and silicone hydrogel contact lenses10.1002/jbm.b.33057 ↗
15. In vivo dehydration of silicone hydrogel contact lenses10.1097/01.ICL.0000072825.23491.59 ↗
16. In vitro evaluation of the dehydration characteristics of silicone hydrogel and conventional hydrogel contact lens materials10.1016/S1367-0484(02)00033-4 ↗
17. Characterization and quantitation of PVP content in a silicone hydrogel contact lens produced by dual‐phase polymerization processing10.1002/jbm.b.33904 ↗
18. Smart contact lenses with graphene coating for electromagnetic interference shielding and dehydration protection10.1021/acsnano.7b00370 ↗
19. Dynamic in vitro dehydration patterns of unworn and worn silicone hydrogel contact lenses10.1002/jbm.b.31279 ↗
20. Lubricity of surface hydrogel layers10.1007/s11249-012-0076-8 ↗
21. Interfacial polymerization of a thin film on contact lenses for improving lubricity10.1016/j.jcis.2020.03.060 ↗
22. Friction coefficient of soft contact lenses: measurements and modeling10.1007/s11249-005-3610-0 ↗
23. Improving wettability and lubricity of commercial contact lenses by polymerizing a thin film of dimethylacryamide10.1016/j.colsurfa.2019.123974 ↗
24. In vitro friction testing of contact lenses and human ocular tissues: Effect of proteoglycan 4 (PRG4)10.1016/j.triboint.2014.11.022 ↗
25. Transcription, translation, and function of lubricin, a boundary lubricant, at the ocular surface10.1001/jamaophthalmol.2013.2385 ↗
26. Effect of hyperosmolarity on PRG4’s ocular surface boundary lubricating ability at a human cornea-eyelid biointerface
27. Proteoglycan 4 and hyaluronan as boundary lubricants for model contact lens hydrogels10.1002/jbm.b.33895 ↗
28. Ethylene oxide-block-butylene oxide copolymer uptake by silicone hydrogel contact lens materials10.1016/j.apsusc.2013.02.064 ↗
29. Friction measurements on contact lenses in a physiologically relevant environment: effect of testing conditions on friction10.1167/iovs.16-19713 ↗
30. Effect of multipurpose care solutions upon physical dimensions of silicone hydrogel contact lenses10.1002/jbm.b.34533 ↗
31. Controlling adhesion with surface hole patterns10.1080/00218460600646610 ↗
32. Enhanced adhesion of elastic materials to small-scale wrinkles10.1021/la302314z ↗
33. Microbial adhesion to silicone hydrogel lenses: a review10.1097/ICL.0b013e318275e284 ↗
34. A laboratory assessment of factors that affect bacterial adhesion to contact lenses10.3390/biology2041268 ↗
35. Corneal cell adhesion to contact lens hydrogel materials enhanced via tear film protein deposition10.1371/journal.pone.0105512 ↗
36. The effect of temperature on soft contact lens modulus and diameter10.1097/ICL.0b013e31822e8c3b ↗
37. Overnight corneal swelling with high and low powered silicone hydrogel lenses10.1016/j.optom.2014.03.002 ↗
38. Effect of lens care systems on silicone hydrogel contact lens hydrophobicity10.1097/ICL.0000000000000247 ↗
39. Limits in measurements of contact lens surface profile using atomic force microscopy10.1016/j.colsurfb.2018.02.018 ↗
40. Johnson · 1971 · 8324 · 301
41. A comparison of JKR-based methods to analyze quasi-static and dynamic indentation force curves10.1016/j.jcis.2005.12.062 ↗
42. Enhanced adhesion of elastic materials to small-scale wrinkles10.1021/la302314z ↗
43. Silicone hydrogel contact lenses and the ocular surface10.1016/S1542-0124(12)70262-8 ↗
44. Imaging protein deposits on contact lens materials10.1097/OPX.0b013e31818e8ad6 ↗
45. Adhesion of Pseudomonas aeruginosa and Staphylococcus epidermidis to silicone–hydrogel contact lenses10.1097/01.opx.0000168585.53845.64 ↗
46. The adhesion strength of Candida parapsilosis to glass and silicone as a function of hydrophobicity, roughness and cell morphology10.1016/j.colsurfa.2004.08.058 ↗
47. Contact lens care solutions and their influence on contact lens hydrophobicity10.1016/j.clae.2018.04.124 ↗
48. The thickness of the tear film10.1080/02713680490516099 ↗
49. The parameters of the porcine eyeball10.1007/s00417-011-1617-9 ↗
50. The effect of polymer surface on the wetting and adhesion of liquid systems10.1163/156856107780684567 ↗
51. Effect of different cleaning regimens on the adhesion of resin to saliva‐contaminated ceramics10.1111/jopr.12170 ↗