Effect of e-beam sterilization on the in vivo performance of conventional UHMWPE tibial plates for total knee arthroplasty

Shine Tone, Masahiro Hasegawa, Giuseppe Pezzotti, Leonardo Puppulin, Akihiro Sudo

Acta Biomaterialia· Volume 55

Abstract

Although the introduction of highly cross-linked polyethylene is effective in reducing the amount of wear, there are still major concerns regarding the use of this material in total knee arthroplasty (TKA), essentially due to the reduction of fatigue resistance and toughness. Monitoring the in vivo performance of different types of UHMWPE is a much needed task to tackle the lack of information on which should be the most reliable choice for TKA. The present study was aimed at investigating the mid-term degradation of electron beam sterilized conventional UHMWPE tibial plates. Visual inspection enabled to grade the surface damage of 12 retrievals according to the Hood's score: the total wear damage correlates to the in vivo time (Spearman's ρ=0.681, p<0.05) and BMI (ρ=0.834, p<0.001). Surface degradation was less severe than that quantified in similar studies on γ-sterilized UHMWPE. Raman and infra-red spectroscopies were utilized to unfold the microstructural modifications. In the load zone, polyethylene whitened damage regions were noticed in the inserts implanted longer than 1year, in which oxidation index (OI) is clearly higher than 1 (max 8). The maximum OI (ρ=0.802, p<0.005) and αc (ρ=0.816, p<0.005) correlate to the implantation time in the load zone. The crystallinity increased along with the extent of oxidation. Concentration of absorbed species from synovial fluid is higher in the contact zone and correlates to maximum OI (Spearman's ρ=0.699, p=0.011). Absorption was promoted in the contact area by the mechanical action of the femoral counterpart and it exacerbated the oxidative degradation in retrievals with high concentration of absorbed species. In the non-load zone, mild but detectable oxidation was observed, probably due to free radicals trapped after sterilization. Statement of significance Although several clinical studies on retrieved tibial bearings have been published so far, monitoring and comparing the in vivo performance of different types of UHMWPE is still a much needed task. The present study reports for the first time results on the effect of sterilization by electron beam on the mid-term in vivo performance of conventional UHMWPE tibial plates. In the present investigation, visual inspection of wear damage based on the Hood's scoring method, Raman micro-spectroscopy and Fourier-transformed infrared spectroscopy were utilized to unveil the damage, the microstructural modifications and the oxidation occurred during implantation. The findings of this investigation have been discussed and compared to previous clinical studies on γ-air sterilized, γ-inert sterilized tibial bearings.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Orthopedic Infections and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (67)

Show source-provided references
1. The Anatomical Graduated Component total knee replacement A LONG-TERM EVALUATION WITH 20-YEAR SURVIVAL ANALYSIS10.1302/0301-620X.91B6.21854 ↗
2. Long-term results of total condylar knee arthroplasty10.1016/j.arth.2005.04.006 ↗
3. Kutz · 2011
4. Why are total knee arthroplasties failing today—has anything changed after 10 years?10.1016/j.arth.2013.07.024 ↗
5. Why are total knee arthroplasties failing today?10.1097/00003086-200211000-00003 ↗
6. Why knee replacements fail in 2013 patient, surgeon, or implant?10.1302/0301-620X.96B11.34350 ↗
7. Osteolysis around total knee arthroplasty: a review of pathogenetic mechanisms10.1016/j.actbio.2013.05.005 ↗
8. The clinical performance of UHMWPE in knee replacements
9. Scuderi · 2014
10. Wear of high-dose gamma-irradiated polyethylene in total hip replacements10.1007/s007760050155 ↗
11. The optimum dose of gamma radiation–heavy doses to low wear polyethylene in total hip prostheses10.1023/A:1018582027349 ↗
12. Current sterilization and packaging methods for polyethylene10.1097/00003086-199912000-00006 ↗
13. Development of an extremely wear-resistant ultra high molecular weight polythylene for total hip replacements10.1002/jor.1100170203 ↗
14. A novel method of cross-linking ultra-high-molecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties: recipient of the 1999 HAP Paul Award10.1054/arth.2001.20540 ↗
15. Galvin · 2005 · 617
16. Metal-on-conventional polyethylene total hip arthroplasty bearing surfaces have a higher risk of revision than metal-on-highly crosslinked polyethylene: results from a US registry10.1007/s11999-014-4105-9 ↗
17. Highly cross-linked polyethylene the debate is over—in opposition10.1016/j.arth.2005.03.016 ↗
18. Highly cross-linked polyethylene: the debate is over—in the affirmative10.1016/j.arth.2005.03.015 ↗
19. Is there a difference in total knee arthroplasty risk of revision in highly crosslinked versus conventional polyethylene?10.1007/s11999-014-4046-3 ↗
20. Is there a rationale to use highly cross-linked polyethylene in posterior—stabilized total knee arthroplasty?
21. Highly cross-linked polyethylene may not have an advantage in total knee arthroplasty10.1007/s11420-013-9352-x ↗
22. Highly crosslinked polyethylene is safe for use in total knee arthroplasty10.1007/s11999-008-0472-4 ↗
23. The impact of sterilization method on wear in knee arthroplasty10.1097/00003086-199811000-00024 ↗
24. Physicochemical and mechanical properties of UHMWPE 45 years’ experience10.1007/s11610-007-0052-4 ↗
25. Ex vivo stability loss of irradiated and melted ultra-high molecular weight polyethylene10.2106/JBJS.I.01017 ↗
26. In vivo oxidative stability changes of highly cross-linked polyethylene bearings: an ex vivo investigation10.1016/j.arth.2015.05.006 ↗
27. A new mechanism of oxidation in ultrahigh molecular weight polyethylene caused by squalene absorption10.1002/jbm.b.32507 ↗
28. Oxidation and other property changes of retrieved sequentially annealed UHMWPE acetabular and tibial bearings10.1002/jbm.b.33240 ↗
29. Degradation of mechanical behavior in UHMWPE after natural and accelerated aging10.1016/S0142-9612(00)00021-1 ↗
30. Villarraga · 2004
31. Micromechanics of shelf-aged and retrieved UHMWPE tibial inserts: indentation testing, oxidative profiling, and thickness effects10.1002/jbm.b.30285 ↗
32. Study of fatigue resistance of chemical and radiation crosslinked medical grade ultrahigh molecular weight polyethylene10.1002/(SICI)1097-4636(19990915)46:4<573::AID-JBM16>3.0.CO;2-A ↗
33. The effects of degree of crosslinking on the fatigue crack initiation and propagation resistance of orthopedic-grade polyethylene10.1002/jbm.a.10606 ↗
34. History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene10.1007/s11999-011-1872-4 ↗
35. Verschleißbeständigkeit von UHMWPE-Artikulationen in der Hüftgelenksendoprothetik
36. Update on UHMWPE research from the bench to the bedside10.1080/17453670810016939 ↗
37. Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses10.1002/jbm.820170510 ↗
38. Raman spectroscopic method for determining the crystallinity of polyethylene
39. Estimation of crystallinity in polyethylene by Raman spectroscopy10.1002/jrs.1250240813 ↗
40. Raman spectroscopy employed for the determination of the intermediate phase in polyethylene10.1021/ma00112a050 ↗
41. A. Standard, F2102-06 · 2006
42. Validation of a protocol based on Raman and infrared spectroscopies to nondestructively estimate the oxidative degradation of UHMWPE used in total joint arthroplasty10.1016/j.actbio.2016.04.040 ↗
43. Wear damage in mobile-bearing TKA is as severe as that in fixed-bearing TKA10.1007/s11999-010-1557-4 ↗
44. Analysis of products diffused into UHMWPE prosthetic components in vivo10.1016/S0142-9612(00)00182-4 ↗
45. Topographical analysis of the femoral components of ex vivo total knee replacements10.1007/s10856-012-4815-z ↗
46. Optical analysis of surface changes on early retrievals of highly cross-linked and conventional polyethylene tibial inserts10.1016/S0883-5403(03)00293-6 ↗
47. Multiaxial fatigue behavior of oxidized and unoxidized UHMWPE during cyclic small punch testing at body temperature10.1520/JAI11218 ↗
48. Knee-simulator testing of conventional and cross-linked polyethylene tibial inserts10.1016/j.arth.2004.03.019 ↗
49. A study of the wear resistance of three types of clinically applied UHMWPE for total replacement hip prostheses10.1016/S0142-9612(99)00096-4 ↗
50. Wear of polyethylene acetabular cups against zirconia femoral heads studied with a hip joint simulator10.1016/0043-1648(94)90148-1 ↗
51. Evolution of morphology in UHMWPE following accelerated aging: the effect of heating rates10.1002/(SICI)1097-4636(199907)46:1<112::AID-JBM13>3.0.CO;2-U ↗
52. The relationship between the clinical performance and large deformation mechanical behavior of retrieved UHMWPE tibial inserts10.1016/S0142-9612(99)00178-7 ↗
53. Radiation induced crystallinity changes in linear polyethylenes: long term aging effects10.1295/polymj.21.739 ↗
54. In vivo oxidation of retrieved cross-linked ultra–high-molecular-weight polyethylene acetabular components with residual free radicals10.1016/j.arth.2005.07.019 ↗
55. Accelerated ageing experiments with crosslinked and conventional ultra-high molecular weight polyethylene (UHMW-PE) stabilised with α-tocopherol for total joint arthroplasty10.1007/s10856-006-0608-6 ↗
56. Time resolved simultaneous small-and wide-angle X-ray scattering during polyethylene deformation—II. Cold drawing of linear polyethylene10.1016/S0032-3861(97)00226-7 ↗
57. Double yield points in polyethylene: structural changes under tensile deformation10.1080/00222349508219486 ↗
58. A real-time simultaneous small-and wide-angle X-ray scattering study of in-situ deformation of isotropic polyethylene10.1021/ma00123a001 ↗
59. Large deformation compression induced crystallinity degradation of conventional and highly crosslinked UHMWPEs10.1016/j.biomaterials.2005.04.008 ↗
60. In vivo oxidation of γ-barrier–sterilized ultra–high-molecular-weight polyethylene bearings10.1016/j.arth.2006.07.006 ↗
61. Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups10.2106/00004623-200709000-00019 ↗
62. In vivo oxidation contributes to delamination but not pitting in polyethylene components for total knee arthroplasty10.1016/j.arth.2010.07.010 ↗
63. Chapter 22: In Vivo oxidation of UHMWPE
64. Can accelerated aqueous aging simulate in vivo oxidation of gamma-sterilized UHMWPE?10.1002/jbm.b.30514 ↗
65. Accelerated aqueous aging simulation of in vivo oxidation for gamma-sterilized UHMWPE10.1002/jbm.b.31294 ↗
66. Otto Aufranc Award Paper: significance of in vivo degradation for polyethylene in total hip arthroplasty10.1097/01.blo.0000246547.18187.0b ↗
67. Bower · 1992