journal article·2023-03Open access

Isolation and characterisation of wear debris surrounding failed total ankle replacements

Ashley A Stratton-Powell, Sophie Williams, Joanne L Tipper, Anthony C Redmond, Claire L Brockett

Acta Biomaterialia· Volume 159

Abstract

Aseptic loosening and osteolysis continue to be a short- to mid-term problem for total ankle replacement (TAR) devices. The production of wear particles may contribute to poor performance, but their characteristics are not well understood. This study aimed to determine the chemical composition, size and morphology of wear particles surrounding failed TARs. A recently developed wear particle isolation method capable of isolating both high- and low-density materials was applied to 20 retrieved periprosthetic tissue samples from 15 failed TARs of three different brands. Isolated particles were imaged using ultra-high-resolution imaging and characterised manually to determine their chemical composition, size, and morphology. Six different materials were identified, which included: UHMWPE, calcium phosphate (CaP), cobalt chromium alloy (CoCr), commercially pure titanium, titanium alloy and stainless steel. Eighteen of the 20 samples contained three or more different wear particle material types. In addition to sub-micron UHMWPE particles, which were present in all samples, elongated micron-sized shards of CaP and flakes of CoCr were commonly isolated from tissues surrounding AES TARs. The mixed particles identified in this study demonstrate the existence of a complex periprosthetic environment surrounding TAR devices. The presence of such particles suggests that early failure of devices may be due in part to the multifaceted biological cascade that ensues after particle release. This study could be used to support the validation of clinically-relevant wear simulator testing, pre-clinical assessment of fixation wear and biological response studies to improve the performance of next generation ankle replacement devices. STATEMENT OF SIGNIFICANCE: Total ankle replacement devices do not perform as well as total hip and knee replacements, which is in part due to the relatively poor scientific understanding of how they fail. The excessive production of certain types of wear debris is known to contribute to joint replacement failure. This is the first study to successfully isolate and characterise high- and low-density wear particles from tissues collected from patients with a failed total ankle replacement. This article includes the chemical composition and characteristics of the wear debris generated by ankle devices, all of which may affect their performance. This research provides clinically relevant reference values and images to support the development of pre-clinical testing for future total ankle replacement designs.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Orthopedic Infections and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (69)

Show source-provided references
1. Effect of bearing surface on survival of cementless and hybrid total hip arthroplasty: study of data in the national joint registry for England, Wales, Northern Ireland and the Isle of Man10.2106/JBJS.OA.19.00075 ↗
2. Wear characteristics of conventional ultrahigh-molecular-weight polyethylene versus highly cross-linked polyethylene in total ankle arthroplasty10.1177/1071100718786501 ↗
3. The clinical biomechanics award 2013 – presented by the International Society of Biomechanics: new observations on the morphology of the talar dome and its relationship to ankle kinematics10.1016/j.clinbiomech.2013.10.009 ↗
4. Outcome after total ankle arthroplasty with a minimum of five years follow-up: a systematic review and meta-analysis10.1016/j.fas.2019.07.006 ↗
5. AKILE™ total ankle arthroplasty: Clinical and CT scan analysis of periprosthetic cysts10.1016/j.otsr.2014.09.019 ↗
6. Peri-prosthetic bone cysts after total ankle replacement. A systematic review and meta-analysis10.1016/j.fas.2017.11.002 ↗
7. Long-term functional and radiographic outcome of a mobile bearing ankle prosthesis10.1177/1071100716661477 ↗
8. How successful are current ankle replacements?: a systematic review of the literature10.1007/s11999-009-0987-3 ↗
9. Medium to long-term results of 130 Ankle Evolutive System total ankle replacements—Inferior survival due to peri-implant osteolysis10.1016/j.fas.2017.03.016 ↗
10. Particle release from dental implants immediately after placement – An ex vivo comparison of different implant systems10.1016/j.dental.2022.04.003 ↗
11. Nano-analyses of wear particles from metal-on-metal and non-metal-on-metal dual modular neck hip arthroplasty10.1016/j.nano.2016.11.003 ↗
12. The lexicon of polyethylene wear in artificial joints10.1016/j.biomaterials.2007.07.040 ↗
13. Systemic and local toxicity of metal debris released from hip prostheses: a review of experimental approaches10.1016/j.nano.2018.01.001 ↗
14. Influence of crosslinking on the wear performance of polyethylene within total ankle arthroplasty10.1177/1071100714558507 ↗
15. Influence of kinematics on the wear of a total ankle replacement10.1016/j.jbiomech.2017.01.001 ↗
16. Wear behaviour in total ankle replacement: a comparison between an in vitro simulation and retrieved prostheses10.1016/j.clinbiomech.2009.06.006 ↗
17. Osteolytic cysts with total ankle replacement: Frequency and causes?10.1016/j.fas.2015.03.005 ↗
18. British Standards Institute, BS ISO 22622:2019 Implants for surgery. Wear of total ankle-joint prostheses. Loading and displacement parameters for wear-testing machines with load or displacement control and corresponding environmental conditions for test, 2019.
19. RANKL in the osteolysis of AES total ankle replacement implants10.1016/j.bone.2012.05.007 ↗
20. Histological analysis of early osteolysis in total ankle arthroplasty10.1177/1071100716682333 ↗
21. Medium term follow-up of the AES ankle prosthesis: high rate of asymptomatic osteolysis, Foot and ankle surgery : official journal of the European Society of Foot and10.1016/j.fas.2009.05.013 ↗
22. Severe periprosthetic osteolytic lesions after the Ankle Evolutive System total ankle replacement10.1302/0301-620X.91B7.22434 ↗
23. Clinical evaluation and radiographic assessment of bone lysis of the AES total ankle replacement10.3113/FAI.2009.0964 ↗
24. Ballooning osteolysis in 71 failed total ankle arthroplasties: Is hydroxyapatite a risk factor?10.1080/17453674.2016.1188346 ↗
25. Periprosthetic osteolysis after total ankle arthroplasty10.1177/1071100713509247 ↗
26. Bone cysts and osteolysis in ankle replacement10.1177/1071100720955155 ↗
27. Scandinavian total ankle replacement: a 3.7-year average followup of 65 patients10.1097/01.blo.0000132245.18548.09 ↗
28. Histologic study of periprosthetic osteolytic lesions after AES total ankle replacement. A 22 case series10.1016/j.otsr.2013.07.009 ↗
29. High rate of osteolytic lesions in medium-term followup after the AES total ankle replacement10.3113/FAI.2011.0168 ↗
30. A wear debris cyst following STAR Total Ankle Replacement—surgical management10.1016/j.fas.2008.06.002 ↗
31. Periprosthetic osteolysis after total ankle arthroplasty
32. Etiopathogenesis of osteolytic cysts associated with total ankle arthroplasty, a histological study10.1016/j.fas.2015.02.004 ↗
33. Management of failed total ankle replacement with a retrograde short femoral nail: a case report10.1097/01.bot.0000171880.03581.4a ↗
34. Ballooning osteolysis in 71 failed total ankle arthroplasties: Is hydroxyapatite a risk factor?
35. British Standard ISO 17853, Wear of implant materials — Polymer and metal wear particles — Isolation and characterization, BS ISO 17853:2011, British Standards Institute, London, UK, 2011.
36. Biologic effects of implant debris
37. Ankle arthroplasties generate wear particles similar to knee arthroplasties10.1097/01.blo.0000132182.54217.da ↗
38. Mixed material wear particle isolation from periprosthetic tissue surrounding total joint replacements10.1002/jbm.b.35076 ↗
39. Identification of nanometre-sized ultra-high molecular weight polyethylene wear particles in samples retrieved in vivo10.1302/0301-620X.90B8.20737 ↗
40. A novel method for isolation and recovery of ceramic nanoparticles and metal wear debris from serum lubricants at ultra-low wear rates10.1016/j.actbio.2016.07.004 ↗
41. Wear debris from hip prostheses characterized by electron imaging
42. ASTM International, F1877–16 Standard practice for characterization of particles, ASTM International,, 2016.
43. Fiji: an open-source platform for biological-image analysis10.1038/nmeth.2019 ↗
44. Histologic and quantitative wear particle analyses of tissue around cementless ceramic total knee prostheses10.1054/arth.2002.29330 ↗
45. Composition and morphology of wear debris in failed uncemented total hip replacement10.1302/0301-620X.76B1.8300684 ↗
46. Isolation and characterization of debris in membranes around total joint prostheses10.2106/00004623-199411000-00010 ↗
47. Isolation of predominantly submicron-sized UHMWPE wear particles from periprosthetic tissues10.1002/jbm.820290118 ↗
48. Isolation and characterization of wear debris generated in patients wearing polyethylene Hylamer inserts, gamma irradiated in air10.1177/0885328205049407 ↗
49. Highly crosslinked polyethylene does not reduce the wear in total knee arthroplasty: in vivo study of particles in synovial fluid10.1016/j.arth.2013.01.026 ↗
50. Quantitative analysis of polyethylene wear debris, wear rate and head damage in retrieved Charnley hip prostheses10.1023/A:1008901302646 ↗
51. Presence of corrosion products and hypersensitivity-associated reactions in periprosthetic tissue after aseptic loosening of total hip replacements with metal bearing surfaces10.1016/j.actbio.2008.07.032 ↗
52. Morphological characteristics of total joint arthroplasty-derived ultra-high molecular weight polyethylene (UHMWPE) wear debris that provoke inflammation in a murine model of inflammation10.1002/jbm.a.10368 ↗
53. Polyethylene wear in retrieved reverse total shoulder components10.1016/j.jse.2011.03.012 ↗
54. Correlation between sintering temperature of hydroxyapatite particles and the production of inflammatory cytokines by human monocytes10.1016/S0927-7765(03)00083-3 ↗
55. Importance of hydroxyapatite particles characteristics on cytokines production by human monocytes in vitro10.1016/S0142-9612(03)00089-9 ↗
56. Periprosthetic osteolysis: mechanisms, prevention and treatment10.3390/jcm8122091 ↗
57. RANKL in the osteolysis of AES total ankle replacement implants10.1016/j.bone.2012.05.007 ↗
58. Histologic study of periprosthetic osteolytic lesions after AES total ankle replacement. A 22 case series10.1016/j.otsr.2013.07.009 ↗
59. Etiopathogenesis of osteolytic cysts associated with total ankle arthroplasty, a histological study10.1016/j.fas.2015.02.004 ↗
60. Elemental and morphological identification of third-body particulate and calcium stearate inclusions in polyethylene components10.1002/(SICI)1097-4636(2000)53:2<137::AID-JBM2>3.0.CO;2-Y ↗
61. Human monocyte response to particulate biomaterials generated in vivo and in vitro10.1002/jor.1100130520 ↗
62. Polyethylene and metal wear particles: characteristics and biological effects10.1007/s00281-011-0242-3 ↗
63. Local cellular responses to titanium dioxide from orthopedic implants10.1089/biores.2017.0017 ↗
64. Characterisation of wear particles produced by metal on metal and ceramic on metal hip prostheses under standard and microseparation simulation10.1007/s10856-006-0015-z ↗
65. Isolation and characterization of wear particles generated in patients who have had failure of a hip arthroplasty without cement10.2106/00004623-199509000-00002 ↗
66. Reproduction of fretting wear at the stem—cement interface in total hip replacement10.1243/09544119JEIM333 ↗
67. Wear study of Total Ankle Replacement explants by microstructural analysis10.1016/j.jmbbm.2015.12.035 ↗
68. Influence of crosslinking on the wear performance of polyethylene within total ankle arthroplasty10.1177/1071100714558507 ↗
69. Postmortem retrieval analysis of metallosis and periprosthetic tissue metal concentrations in total knee arthroplasty10.1016/j.arth.2019.08.038 ↗