Extrusion-based 3D printed biodegradable porous iron
Source abstract
Abstract
Extrusion-based 3D printing followed by debinding and sintering is a powerful approach that allows for the fabrication of porous scaffolds from materials (or material combinations) that are otherwise very challenging to process using other additive manufacturing techniques. Iron is one of the materials that have been recently shown to be amenable to processing using this approach. Indeed, a fully interconnected porous design has the potential of resolving the fundamental issue regarding bulk iron, namely a very low rate of biodegradation. However, no extensive evaluation of the biodegradation behavior and properties of porous iron scaffolds made by extrusion-based 3D printing has been reported. Therefore, the in vitro biodegradation behavior, electrochemical response, evolution of mechanical properties along with biodegradation, and responses of an osteoblastic cell line to the 3D printed iron scaffolds were studied. An ink formulation, as well as matching 3D printing, debinding and sintering conditions, was developed to create iron scaffolds with a porosity of 67%, a pore interconnectivity of 96%, and a strut density of 89% after sintering. X-ray diffracometry confirmed the presence of the α-iron phase in the scaffolds without any residuals from the rest of the ink. Owing to the presence of geometrically designed macropores and random micropores in the struts, the in vitro corrosion rate of the scaffolds was much improved as compared to the bulk counterpart, with 7% mass loss after 28 days. The mechanical properties of the scaffolds remained in the range of those of trabecular bone despite 28 days of in vitro biodegradation. The direct culture of MC3T3-E1 preosteoblasts on the scaffolds led to a substantial reduction in living cell count, caused by a high concentration of iron ions, as revealed by the indirect assays. On the other hand, the ability of the cells to spread and form filopodia indicated the cytocompatibility of the corrosion products. Taken together, this study shows the great potential of extrusion-based 3D printed porous iron to be further developed as a biodegradable bone substituting biomaterial.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Bone Tissue Engineering Materials · Magnesium Alloys: Properties and Applications · Orthopaedic implants and arthroplasty
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (97)
Show source-provided references
1. Updates on the research and development of absorbable metals for biomedical applications10.1007/s40204-018-0091-4 ↗
2. Current status and outlook on the clinical translation of biodegradable metals10.1016/j.mattod.2018.05.018 ↗
3. Biodegradable metals10.1016/j.mser.2014.01.001 ↗
4. Design strategy for biodegradable Fe-based alloys for medical applications10.1016/j.actbio.2009.07.039 ↗
5. Advances in Fe-based biodegradable metallic materials10.1039/C6RA20594A ↗
6. Recent advancements in Fe-based biodegradable materials for bone repair10.1007/s10853-018-3011-z ↗
7. Review on iron and its importance for human health
8. Influence of iron on bone homeostasis10.3390/ph11040107 ↗
9. The role of reactive oxygen species in mesenchymal stem cell adipogenic and osteogenic differentiation: a review10.1089/scd.2014.0484 ↗
10. Iron restriction negatively affects bone in female rats and mineralization of hFOB osteoblast cells10.1177/153537020623100403 ↗
11. Iron chelator deferoxamine alters iron-regulatory genes and proteins and suppresses osteoblast phenotype in fetal rat calvaria cells10.1016/j.bone.2010.01.376 ↗
12. Iron-dependent human platelet activation and hydroxyl radical formation: involvement of protein kinase C10.1161/01.CIR.99.24.3118 ↗
13. Biocorrosion properties and blood and cell compatibility of pure iron as a biodegradable biomaterial10.1007/s10856-010-4070-0 ↗
14. Biodegradable Fe-based alloys for use in osteosynthesis: outcome of an in vivo study after 52 weeks10.1016/j.actbio.2014.04.007 ↗
15. Strontium and bisphosphonate coated iron foam scaffolds for osteoporotic fracture defect healing10.1016/j.biomaterials.2017.11.049 ↗
16. Magnesium, iron and zinc alloys, the trifecta of bioresorbable orthopaedic and vascular implantation - a review10.4172/2155-952X.1000178 ↗
17. Antiferromagnetism of γ-FeMn alloys10.1063/1.1656274 ↗
18. Novel Fe-Mn-Si-Pd alloys: insights into mechanical, magnetic, corrosion resistance and biocompatibility performances, R
19. Novel biodegradable Fe-Mn-C-S alloy with superior mechanical and corrosion properties10.1016/j.matlet.2016.10.037 ↗
20. Accelerated degradation behavior and cytocompatibility of pure iron treated with sandblasting10.1021/acsami.6b07068 ↗
21. Controlling the degradation kinetics of porous iron by poly(lactic-co-glycolic acid) infiltration for use as temporary medical implants10.1038/srep11194 ↗
22. An in vitro corrosion study of open cell iron structures with PEG coating for bone replacement applications10.3390/met8070499 ↗
23. In vitro corrosion behavior of biodegradable iron foams with polymeric coating10.3390/ma13010184 ↗
24. Development and characterization of silver containing calcium phosphate coatings on pure iron foam intended for bone scaffold applications10.1016/j.matdes.2018.03.061 ↗
25. Cancellous-bone-like porous iron scaffold coated with strontium incorporated octacalcium phosphate nanowhiskers for bone regeneration10.1021/acsbiomaterials.8b01188 ↗
26. Investigation on mechanical behavior of biodegradable iron foams under different compression test conditions10.3390/met7060202 ↗
27. Microstructural and mechanical characteristics of porous iron prepared by powder metallurgy10.1016/j.msec.2014.06.046 ↗
28. Microstructural and mechanical properties of biodegradable iron foam prepared by powder metallurgy10.1016/j.matdes.2015.06.022 ↗
29. Mechanical properties, corrosion performance and cell viability studies on newly developed porous Fe-Mn-Si-Pd alloys10.1016/j.jallcom.2017.07.112 ↗
30. Bioresorbable Fe–Mn and Fe–Mn–HA m for orthopedic implantation: enhancing degradation through porosity control10.1002/adhm.201700120 ↗
31. Investigation of porosity on mechanical properties, degradation and in-vitro cytotoxicity limit of Fe30Mn using space holder technique10.1016/j.msec.2019.02.055 ↗
32. In vitro degradation behaviour, cytocompatibility and hemocompatibility of topologically ordered porous iron scaffold prepared using 3D printing and pressureless microwave sintering10.1016/j.msec.2019.110247 ↗
33. Novel processing of iron-manganese alloy-based biomaterials by inkjet 3-D printing10.1016/j.actbio.2013.04.016 ↗
34. Binder-jetting 3D printing and alloy development of new biodegradable Fe-Mn-Ca/Mg alloys10.1016/j.actbio.2016.08.032 ↗
35. 3D printed Fe scaffolds with HA nanocoating for bone regeneration10.1021/acsbiomaterials.7b00885 ↗
36. 3D printing of high-strength bioscaffolds for the synergistic treatment of bone cancer10.1038/s41427-018-0015-8 ↗
37. Experimental investigation into the fabrication of green body developed by micro-extrusion-based 3D printing process10.1002/pc.25514 ↗
38. Effect of sintering parameters on the microstructure and compressive mechanical properties of porous Fe scaffold fabricated using 3D printing and pressure less microwave sintering
39. Mechanical behaviour of 3D printed ordered pore topological iron scaffold10.1016/j.msea.2020.139293 ↗
40. Additively manufactured biodegradable porous iron10.1016/j.actbio.2018.07.011 ↗
41. Additively manufactured functionally graded biodegradable porous iron10.1016/j.actbio.2019.07.013 ↗
42. Selective laser melted Fe-Mn bone scaffold: microstructure, corrosion behavior and cell response
43. Additively manufactured iron-manganese for biodegradable porous load-bearing bone scaffold applications10.1016/j.actbio.2019.12.018 ↗
44. Additive manufacturing of biomaterials, tissues, and organs10.1007/s10439-016-1719-y ↗
45. Additive manufacturing of biodegradable metals: current research status and future perspectives10.1016/j.actbio.2019.04.046 ↗
46. Multi-material additive manufacturing technologies for Ti-, Mg-, and Fe-based biomaterials for bone substitution10.1016/j.actbio.2020.03.037 ↗
47. A strawberry-like Ag-decorated barium titanate enhances piezoelectric and antibacterial activities of polymer scaffold10.1016/j.nanoen.2020.104825 ↗
48. Laser additive manufacturing of Mg-based composite with improved degradation behaviour10.1080/17452759.2020.1748381 ↗
49. 2017
50. Preparation and assessment of revised simulated body fluids10.1002/jbm.a.10482 ↗
51. Biocorrosion behavior of magnesium alloy in different simulated fluids for biomedical application10.1016/j.msec.2009.01.014 ↗
52. 2017
53. 2004
54. 2015
55. 2011
56. 2012
57. The use of hypromellose in oral drug delivery10.1211/0022357055957 ↗
58. Tailoring green and sintered density of pure iron parts using binder jetting additive manufacturing
59. Fabrication of 3D plotted scaffold with microporous strands for bone tissue engineering10.1039/C9TB02360G ↗
60. In vitro degradation and biocompatibility of Fe-Pd and Fe-Pt composites fabricated by spark plasma sintering10.1016/j.msec.2013.10.023 ↗
61. A novel porous Fe/Fe-W alloy scaffold with a double-layer structured skeleton: preparation, in vitro degradability and biocompatibility10.1016/j.colsurfb.2016.03.002 ↗
62. Influence of pore characteristics on microstructure, mechanical properties and corrosion resistance of selective laser sintered porous Ti-Mo alloys for biomedical applications10.1016/j.electacta.2013.04.105 ↗
63. Effects of porosity on corrosion resistance of Mg alloy foam produced by powder metallurgy technology10.1016/j.matchar.2014.07.012 ↗
64. Corrosion behaviour of the porous iron scaffold in simulated body fluid for biodegradable implant application10.1016/j.msec.2019.01.114 ↗
65. Effect of capillary action on bone regeneration in micro-channeled ceramic scaffolds10.1016/j.ceramint.2014.02.033 ↗
66. Electroformed pure iron as a new biomaterial for degradable stents: in vitro degradation and preliminary cell viability studies10.1016/j.actbio.2010.01.008 ↗
67. Influence of cross-rolling on the micro-texture and biodegradation of pure iron as biodegradable material for medical implants10.1016/j.actbio.2015.01.024 ↗
68. Strong corrosion induced by carbon nanotubes to accelerate Fe biodegradation10.1016/j.msec.2019.109935 ↗
69. Highly biodegradable and bioactive Fe-Pd-bredigite biocomposites prepared by selective laser melting10.1016/j.jare.2019.06.001 ↗
70. Electrochemical aspects in biomedical alloy characterization: electrochemical Impedance Spectrosopy10.5772/13039 ↗
71. Influence of preparation conditions of Layered Double Hydroxide conversion films on corrosion protection10.1016/j.electacta.2013.11.111 ↗
72. The evaluation of coating performance by analyzing the intersection of Bode plots10.1016/S1452-3981(23)08155-5 ↗
73. Another approach in analysis of paint coatings with EIS measurement: phase angle at high frequencies10.1016/j.corsci.2006.03.012 ↗
74. The evaluation of coating performance by the variations of phase angles in middle and high frequency domains of EIS10.1016/j.corsci.2008.08.049 ↗
75. Protection of iron corrosion by stearic acid and stearic imidazoline self-assembled monolayers
76. An investigation of the corrosion of pure iron by electrochemical techniques and in situ observations10.1016/0010-938X(93)90254-E ↗
77. Physical Interpretations of Nyquist Plots for EDLC Electrodes and Devices10.1021/acs.jpcc.7b10582 ↗
78. In vitro and in vivo corrosion properties of new iron-manganese alloys designed for cardiovascular applications10.1002/jbm.b.33234 ↗
79. Mechanical properties of cortical and trabecular bone
80. Biomechanics of trabecular bone10.1146/annurev.bioeng.3.1.307 ↗
81. Additively manufactured open-cell porous biomaterials made from six different space-filling unit cells: the mechanical and morphological properties10.3390/ma8041871 ↗
82. Additively manufactured metallic porous biomaterials based on minimal surfaces: a unique combination of topological, mechanical, and mass transport properties10.1016/j.actbio.2017.02.024 ↗
83. Investigation of the structure and mechanical properties of additively manufactured Ti-6Al-4 V biomedical scaffolds designed with a Schwartz primitive unit-cell10.1016/j.msea.2018.12.104 ↗
84. D. Carluccio, C. Xu, J. Venezuela, Y. Cao, D. Kent, M. Bermingham, A.G. Demir, B. Previtali, Q. Ye, M. Dargusch, Additively manufactured iron-manganese for biodegradable porous load-bearing bone scaffold applications, (2019) 1–15. 10.1016/j.actbio.2019.12.018.10.1016/j.actbio.2019.12.018 ↗
85. Strain hardening of porous iron under uniaxial compression10.1007/BF02677449 ↗
86. The manufacturing of high porosity iron with an ultra-fine microstructure via free pressureless spark plasma sintering10.3390/ma9060495 ↗
87. Effects of particle size, particle/matrix interface adhesion and particle loading on mechanical properties of particulate-polymer composites10.1016/j.compositesb.2008.01.002 ↗
88. Microstructure, mechanical property, biodegradation behavior, and biocompatibility of biodegradable Fe-Fe2O3 composites10.1002/jbm.a.34882 ↗
89. 2009
90. Iron overload inhibits osteoblast biological activity through oxidative stress10.1007/s12011-013-9605-z ↗
91. Exploring the role of manganese on the microstructure, mechanical properties, biodegradability, and biocompatibility of porous iron-based scaffolds10.1021/acsbiomaterials.8b01497 ↗
92. Protein interactions with corroding metal surfaces: comparison of Mg and Fe10.1039/C4FD00253A ↗
93. Influence of albumin interaction on corrosion resistance of sintered iron biomaterials with polyethyleneimine coating10.1016/j.apsusc.2020.145379 ↗
94. A new Fe-Mn-Si alloplastic biomaterial as bone grafting material: in vivo study10.1016/j.apsusc.2015.04.197 ↗
95. Characterization and in vivo evaluation of a bio-corrodible nitrided iron stent10.1007/s10856-012-4823-z ↗
96. In vivo evaluation of biodegradability and biocompatibility of Fe30Mn alloy10.3415/VCOT-17-06-0080 ↗
97. Are Fe-based stenting materials biocompatible? A critical review of in vitro and in vivo studies