journal article·2014-06

Potential release of in vivo trace metals from metallic medical implants in the human body: From ions to nanoparticles – A systematic analytical review

Henryk Matusiewicz

Acta Biomaterialia· Volume 10· Issue 6
Publisher page

Abstract

Metal ion release from metallic materials, e.g. metallic alloys and pure metals, implanted into the human body in dental and orthopedic surgery is becoming a major cause for concern. This review briefly provides an overview of both metallic alloys and pure metals used in implant materials in dental and orthopedic surgery. Additionally, a short section is dedicated to important biomaterials and their corrosive behavior in both real solutions and various types of media that model human biological fluids and tissues. The present review gives an overview of analytical methods, techniques and different approaches applied to the measurement of in vivo trace metals released into body fluids and tissues from patients carrying metal-on-metal prostheses and metal dental implants. Reference levels of ion concentrations in body fluids and tissues that have been determined by a host of studies are compiled, reviewed and presented in this paper. Finally, a collection of published clinical data on in vivo released trace metals from metallic medical implants is included.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Extraction and Separation Processes · Titanium Alloys Microstructure and Properties

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (100+)

Show source-provided references
1. Special modes of corrosion under physiological and simulated physiological conditions10.1016/j.actbio.2007.12.003 ↗
2. Corrosion of bio implants
3. Metal ion release from metal implants10.1016/j.msec.2004.08.018 ↗
4. Metal is not inert: role of metal ions released by biocorrosion in aseptic loosening – current concepts10.1002/jbm.a.32625 ↗
5. Corrosion degradation and prevention by surface modification of biometallic materials10.1007/s10856-006-0016-y ↗
6. Assessing the corrosion of biodegradable magnesium implants: a critical review of current methodologies and their limitations10.1016/j.actbio.2011.11.014 ↗
7. The issue of corrosion in dental implants: a review
8. Corrosion in titanium dental implants/prostheses – a review
9. Corrosion fatigue of biomedical metallic alloys: mechanisms and mitigation10.1016/j.actbio.2011.09.012 ↗
10. Wear induced by motion between bone and titanium or cobalt–chrome alloys10.1302/0301-620X.76B5.8083256 ↗
11. Current concepts review. Corrosion of metal orthopaedic implants10.2106/00004623-199802000-00015 ↗
12. Migration of corrosion products from modular hip prostheses10.2106/00004623-199409000-00009 ↗
13. Wear debris from hip or knee replacements causes chromosomal damage in human cells in tissue culture10.1302/0301-620X.86B4.14368 ↗
14. In vivo production of nanosized metal wear debris formed by tribochemical reaction as confirmed by high-resolution TEM and XPS analyses
15. The study of metal ion release and cytotoxicity in Co–Cr–Mo and Ti–Al–V alloy in total knee prosthesis – scanning electron microscopic observation
16. Dissemination of wear particles to the liver, spleen, and abdominal lymph nodes of patients with hip or knee replacement10.2106/00004623-200004000-00002 ↗
17. Detection of metallic elements migration around a prosthesis by neutron activation analysis and by the PIXE method
18. Metal wear particle characterization from metal on metal total hip replacements: transmission electron microscopy study of periprosthetic tissues and isolated particles10.1002/(SICI)1097-4636(199810)42:1<103::AID-JBM13>3.0.CO;2-M ↗
19. Migration of metal and polyethylene particles from articular prostheses may generate lymphadenopathy with histiocytosis10.1002/(SICI)1097-4636(199602)30:2<157::AID-JBM4>3.0.CO;2-Q ↗
20. Autopsy analysis thirty years after metal-on-metal total hip replacement10.2106/00004623-200311000-00027 ↗
21. Presence of corrosion products and hypersensitivity-associated reactions in periprosthetic tissue after aseptic loosening of total hip replacements with metal bearing surfaces10.1016/j.actbio.2008.07.032 ↗
22. Effect of femoral head size on the wear of metal on metal bearings in total hip replacements under adverse edge-loading conditions10.1002/jbm.b.32824 ↗
23. Effects of digestion protocols on the isolation and characterization of metal–metal wear particles. I. Analysis of particle size and shape10.1002/1097-4636(20010605)55:3<320::AID-JBM1020>3.0.CO;2-3 ↗
24. Effects of digestion protocols on the isolation and characterization of metal–metal wear particles. II. Analysis of ion release and particle composition10.1002/1097-4636(20010605)55:3<330::AID-JBM1021>3.0.CO;2-Y ↗
25. Widespread dissemination of metal debris from implants10.1302/0301-620X.76B5.8083255 ↗
26. PIXE micro-beam mapping of metals in human peri-implant tissues10.1023/A:1020309108950 ↗
27. Development of new metallic alloys for biomedical applications10.1016/j.actbio.2012.06.037 ↗
28. Titanium alloys in total joint replacement – a materials science perspective10.1016/S0142-9612(97)00146-4 ↗
29. Magnesium and its alloys as orthopedic biomaterials: A review10.1016/j.biomaterials.2005.10.003 ↗
30. The history of biodegradable magnesium implants: a review10.1016/j.actbio.2010.02.028 ↗
31. Porous NiTi for bone implants: a review10.1016/j.actbio.2008.02.009 ↗
32. Introduction to corrosion of bioimplants10.1134/S2070205111030105 ↗
33. Review of wear mechanisms in hip implants. Paper I. General10.1016/j.matdes.2003.11.010 ↗
34. Metallic materials for biomedical applications: laboratory and clinical studies10.1351/PAC-CON-10-07-09 ↗
35. Cutaneous metal sensitivity in patients with orthopaedic injuries10.1097/00005131-200102000-00002 ↗
36. Hip implants. Paper VI. Ion concentrations10.1016/j.matdes.2005.05.018 ↗
37. Release of chromium from orthopaedic arthroplasties10.2174/1874325000802010010 ↗
38. Cobalt-base alloys used in bone surgery10.1016/S0020-1383(00)80018-2 ↗
39. Comparison of metal release from various metallic biomaterials in vitro10.1016/j.biomaterials.2004.02.005 ↗
40. Cobalt, chromium, and nickel concentrations in body fluids of patients with porous-coated knee or hip prostheses10.1002/jor.1100070302 ↗
41. Determination of chromium, cobalt and molybdenum in synovial fluid by GFAAS
42. Serum titanium level for diagnosis of a failed, metal-backed patellar component10.1054/arth.2000.6632 ↗
43. Study rationale and protocol: prospective randomized comparison of metal ion concentrations in the patient’s plasma after implantation of coated and uncoated total knee prostheses10.1186/1471-2474-10-128 ↗
44. Serum metal ion exposure after total knee arthroplasty10.1097/BLO.0b013e31806450ef ↗
45. Serum metal ion levels after rotating-hinge knee arthroplasty: comparison between a standard device and a megaprosthesis10.1007/s00264-011-1317-4 ↗
46. High-resolution ICP-MS determination of Ti, V, Cr Co, Ni, and Mo in human blood and urine of patients implanted with a hip or knee prosthesis10.1007/s00216-008-2188-4 ↗
47. Concentration of metal elements in the blood and urine in the patients with cementless total knee arthroplasty10.1620/tjem.185.253 ↗
48. A dried urine spot test to simultaneously monitor Mo and Ti levels using solid sampling high-resolution continuum source graphite furnace atomic absorption spectrometry10.1016/j.sab.2012.12.001 ↗
49. Concentration of wear products in hair, blood, and urine after total hip replacement10.1136/bmj.1.5852.527 ↗
50. Metal-on-metal total hip replacement. What does the literature say?10.1016/j.arth.2004.08.011 ↗
51. Metal determination in organic fluids of patients with stainless steel hip arthroplasty10.3109/17453678308992891 ↗
52. Can a safe level for metal ions in patients with metal-on-metal total hip arthroplasties be determined?10.1016/j.arth.2004.09.008 ↗
53. Increased blood cobalt and chromium after total hip replacement10.1081/CLT-100102463 ↗
54. Four-year study of cobalt and chromium blood levels in patients managed with two different metal-on-metal total hip replacements10.1016/S0736-0266(02)00152-3 ↗
55. Ion release and chromosomal damage from total hip prostheses with metal-on-metal articulation10.1002/jbm.b.10070 ↗
56. Changes in metal levels and chromosome aberrations in the peripheral blood of patients after metal-on-metal hip arthroplasty10.1016/j.arth.2004.09.010 ↗
57. Metal ion levels after metal-on-metal proximal femoral replacements. A 30-year follow-up10.1302/0301-620X.87B5.15384 ↗
58. The association between metal ions from hip resurfacing and reduced T-cell counts10.1302/0301-620X.88B4.17216 ↗
59. Metal ion levels in the blood of patients after hip resurfacing: a comparison between twenty-eight and thirty-six-millimeter-head metal-on-metal prostheses10.2106/JBJS.H.00442 ↗
60. The effect of the diameter of metal-on-metal bearings on systemic exposure to cobalt and chromium10.1302/0301-620X.88B4.17355 ↗
61. Mottard S, Roy AG, Dupont C, Lavigne M. Chromium and cobalt ion release following the Durom high carbon content, forged metal-on-metal surface replacement of the hip10.1302/0301-620X.89B4.18054 ↗
62. Metal ion release from bearing wear and corrosion with 28 mm and large-diameter metal-on-metal bearing articulations. A follow-up study10.1302/0301-620X.92B1.22226 ↗
63. Metal-on-metal hip resurfacing compared with 28-mm diameter metal-on-metal total hip replacement10.1302/0301-620X.95B11.31604 ↗
64. Blood and urine metal ion levels in young and active patients after Birmingham hip resurfacing arthroplasty10.1302/0301-620X.89B2.18519 ↗
65. Six-year results of a prospective study of metal ion levels in young patients with metal-on-metal hip resurfacings10.1302/0301-620X.91B2.21654 ↗
66. Transplacental transfer of cobalt and chromium in patients with metal-on-metal hip arthroplasty. A controlled study10.1302/0301-620X.89B3.18520 ↗
67. The validity of serum levels as a surrogate measure of systemic exposure to metal ions in hip replacement10.1302/0301-620X.89B6.18141 ↗
68. Distribution of chromium and cobalt ions in various blood fractions after resurfacing hip arthroplasty10.1016/j.arth.2007.07.003 ↗
69. Metal ion interpretation in resurfacing versus conventional hip arthroplasty and in whole blood versus serum. How should we interpret metal ion data?10.5301/HIP.2011.8643 ↗
70. Metal ion levels and functional results after either resurfacing hip arthroplasty or conventional metal-on-metal hip arthroplasty. Short-term outcome of a randomized controlled trial10.3109/17453674.2011.625533 ↗
71. Sensitivity and specificity of blood cobalt and chromium metal ions for predicting failure of metal-on-metal hip replacement10.1302/0301-620X.93B10.26249 ↗
72. Insufficient acetabular version increases blood metal ion levels after meta-on-metal hip resurfacing10.1007/s11999-011-1930-y ↗
73. Differential distribution of cobalt, chromium, and nickel between whole blood, plasma and urine in patients after metal-on-metal (MoM) hip arthroplasty10.1002/jor.22107 ↗
74. Cup inclination angle of greater than 50 degrees increases whole blood concentrations of cobalt and chromium ions after metal-on-metal hip resurfacing10.1177/112070000801800304 ↗
75. An analysis of metal ion levels in th joint fluid of symptomatic patients with metal-on-metal hip replacements10.1302/0301-620X.93B6.25804 ↗
76. The effect of component size and orientation on the concentrations of metal ions after resurfacing arthroplasty of the hip10.1302/0301-620X.90B9.20785 ↗
77. Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement. A consequence of excess wear10.1302/0301-620X.92B1.22770 ↗
78. Surgical variables influence metal ion levels after hip resurfacing10.1007/s11999-010-1636-6 ↗
79. Metal ion release with large-diameter metal-on-metal hip arthroplasty10.1016/j.arth.2009.12.013 ↗
80. In vivo serum titanium ion levels following modular neck total hip arthroplasty – 10 year results in 67 patients10.1016/j.actbio.2012.12.001 ↗
81. Significance of elevated blood metal ion levels in patients with metal-on-metal prostheses: an evaluation of oxidative stress markers10.2174/1874325001004010221 ↗
82. Development of a routine method for the determination of trace metals in whole blood by magnetic sector inductively coupled plasma mass spectrometry with particular relevance to patients with total hip and knee arthroplasty10.1093/clinchem/47.2.275 ↗
83. Changes in whole blood metal ion levels following resurfacing: serial measurements in a multi-centre study10.1177/112070000901900406 ↗
84. Femoral head size does not affect ion values in metal-on-metal total hips10.1007/s11999-010-1630-z ↗
85. Whole blood metal ion concentrations in correlation with activity level in three different metal-on-metal bearings10.1016/j.arth.2009.11.007 ↗
86. Comparison of whole-blood metal ion levels in four types of metal-on-metal large-diameter femoral head total hip arthroplasty: the potential influence of the adapter sleeve10.2106/JBJS.J.01885 ↗
87. Do ion levels in hip resurfacing differ from metal-on-metal THA at midterm?10.1007/s11999-010-1424-3 ↗
88. Do ion levels in metal-on-metal hip resurfacing differ from those in metal-on-metal THA at long-term followup?10.1007/s11999-013-2981-z ↗
89. What happens to serum metal ion levels after a metal-on-metal bearing is removed?10.1016/j.arth.2013.06.040 ↗
90. Serum and urine metal levels in patients with metal-on-metal surface arthroplasty10.1023/A:1021179029640 ↗
91. Elevated serum cobalt with metal-on-metal articulating surfaces10.1302/0301-620X.79B2.7326 ↗
92. Serum cobalt levels after metal-on-metal total hip arthroplasty10.2106/00004623-200311000-00017 ↗
93. Value and limits of determining serum cobalt levels in patients with metal on metal articulating prostheses
94. Levels of metal ions after small- and large-diameter metal-on-metal hip arthroplasty10.1302/0301-620X.85B6.14166 ↗
95. Ion release in stable hip arthroplasties using metal-on-metal articulating surfaces: a comparison between short- and medium-term results10.1002/jbm.a.10595 ↗
96. Levels of systemic metal ions in patients with intramedullary nails10.1080/17453670810016911 ↗
97. Metal serum levels in 48 patients bearing a chromium–cobalt total hip arthroplasty with a metal-on-polyethylene combination10.1016/S0035-1040(04)70100-2 ↗
98. Chronological changes of serum chromium levels after modern metal-on-metal total hip arthroplasty10.1080/00016470410001187-1 ↗
99. Seven years of chronological changes of serum chromium levels after metasul metal-on-metal total hip arthroplasty10.1016/j.arth.2009.08.016 ↗
100. Cobalt and chromium ion release after large-diameter metal-on-metal total hip arthroplasty10.1016/j.arth.2011.12.016 ↗