Recommendation for modifying current cytotoxicity testing standards for biodegradable magnesium-based materials
Source abstract
Abstract
As one of the most promising medical metal implants, magnesium (Mg) or its alloys have shown significant advantages over other candidates attributed to not only their excellent biodegradability and suitable mechanical properties but also their osteopromotive effects for bone applications. Prior to approval mandated by the governmental regulatory body, the access to the medical market for Mg-based implants requires a series of testing for assurance of their safety and efficacy via preclinical evaluations and clinical tests including phase 1 and 2 evaluations, and phase 3 of multi-center randomized double blind and placebo-controlled clinical trials. However, as the most widely used protocols for biosafety evaluation of medical devices, current ISO 10993 standards should be carefully reevaluated when directly applying them to predict potential health risks of degradable Mg based biomaterials via cytotoxicity tests due to the huge gap between in vitro and in vivo conditions. Therefore, instead of a direct adoption, modification of current ISO standards for in vitro cytotoxicity test is desirable and justified. The differences in sensitivities of cells to in vitro and in vivo Mg ions and the capability of in vivo circulation system to dilute local degradation products were fully considered to propose modification of current ISO standards. This paper recommended a minimal 6 times to a maximal 10 times dilution of extracts for in vitro cytotoxicity test specified in ISO 10993 part 5 for pure Mg developed as potential orthopedic implants based on literature review and our specifically designed in vitro and in vivo tests presented in the study. Our work may contribute to the progress of biodegradable metals involved translational work.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Magnesium Alloys: Properties and Applications · Magnesium in Health and Disease · Aluminum Alloys Composites Properties
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (50)
Show source-provided references
1. The history of biodegradable magnesium implants: a review10.1016/j.actbio.2010.02.028 ↗
2. Surface coating reduces degradation rate of magnesium alloy developed for orthopaedic applications
3. Degradation and in vitro cell–material interaction studies on hydroxyapatite-coated biodegradable porous iron for hard tissue scaffolds
4. Similarities and differences in coatings for magnesium-based stents and orthopaedic implants
5. Progress and challenge for magnesium alloys as biomaterials10.1002/adem.200800035 ↗
6. Biodegradable metals10.1016/j.mser.2014.01.001 ↗
7. Magnesium and microvascular endothelial cells: a role in inflammation and angiogenesis10.2741/1610 ↗
8. In vitro and in vivo studies on biodegradable CaMgZnSrYb high-entropy bulk metallic glass10.1016/j.actbio.2013.01.029 ↗
9. In vitro and in vivo studies on a Mg–Sr binary alloy system developed as a new kind of biodegradable metal10.1016/j.actbio.2012.02.018 ↗
10. Biodegradable CaMgZn bulk metallic glass for potential skeletal application10.1016/j.actbio.2011.04.027 ↗
11. Regulation of medical devices in the United States and European Union10.1056/NEJMhle1113918 ↗
12. Biodegradable magnesium-based screw clinically equivalent to titanium screw in hallux valgus surgery: short term results of the first prospective, randomized, controlled clinical pilot study
13. In vivo evaluation of biodegradable magnesium alloy bone implant in the first 6 months implantation10.1002/jbm.a.32132 ↗
14. In vitro corrosion and biocompatibility of binary magnesium alloys10.1016/j.biomaterials.2008.10.021 ↗
15. ISO 10993-5: biological evaluation of medical devices: tests for in vitro cytotoxicity. International Organization of Standards; 2009.
16. ISO 10993-12: biological evaluation of medical devices: sample preparation and reference materials. International Organization for Standardization; 2012.
17. Mechanisms of polymer degradation and erosion10.1016/0142-9612(96)85755-3 ↗
18. Assessing the corrosion of biodegradable magnesium implants: a critical review of current methodologies and their limitations10.1016/j.actbio.2011.11.014 ↗
19. Effect of inorganic salts, amino acids and proteins on the degradation of pure magnesium in vitro10.1016/j.msec.2008.12.015 ↗
20. Improved cytotoxicity testing of magnesium materials10.1016/j.mseb.2011.04.008 ↗
21. Comparison of different in vitro tests for biocompatibility screening of Mg alloys10.1016/j.actbio.2013.02.020 ↗
22. Luo TJ, Yang YS, Shi BL, Feng XH, Li YJ. The method and device for preparation of ultra high purity magnesium by industrial pure magnesium. China Patent No. CN102808090, 2012.
23. Influence of nicotine on the biological activity of rabbit osteoblasts10.1111/j.1600-0501.2010.02088.x ↗
24. Leung · 2008
25. Exogenous phytoestrogenic molecule icaritin incorporated into a porous scaffold for enhancing bone defect repair10.1002/jor.22188 ↗
26. Importance of melastatin-like transient receptor potential 7 and cations (magnesium, calcium) in human osteoblast-like cell proliferation10.1111/j.1365-2184.2007.00476.x ↗
27. Analysis of apoptosis by propidium iodide staining and flow cytometry10.1038/nprot.2006.238 ↗
28. Effect of cell line on in vitro metal ion cytotoxicity10.1016/0109-5641(94)90025-6 ↗
29. Alkaline stress-induced apoptosis in human pulmonary artery endothelial cells10.1007/s10495-005-1402-5 ↗
30. High osmolality activates the G1 and G2 cell cycle checkpoints and affects the DNA integrity of nucleus pulposus intervertebral disc cells triggering an enhanced DNA repair response10.1016/j.dnarep.2009.05.005 ↗
31. The structure and regulation of magnesium selective ion channels10.1016/j.bbamem.2013.08.002 ↗
32. Diffusion permeability and ultrafiltration-reflection-coefficients of Narand Cl- in the near-term placenta of the sheep
33. Hyperosmotic stress signaling to the nucleus disrupts the ran gradient and the production of ranGTP10.1091/mbc.E07-01-0089 ↗
34. Magnesium. An update on physiological, clinical and analytical aspects10.1016/S0009-8981(99)00258-2 ↗
35. High extracellular magnesium inhibits mineralized matrix deposition and modulates intracellular calcium signaling in human bone marrow-derived mesenchymal stem cells
36. Mg2+ block unmasks Ca2+/Ba2+ selectivity of alpha 1G T-type calcium channels10.1016/S0006-3495(00)76540-4 ↗
37. A new class of ligand-gated ion channel defined by P2x receptor for extracellular ATP10.1038/371516a0 ↗
38. Ligand-gated calcium channels inside and out10.1016/S0955-0674(96)80025-1 ↗
39. Competitive Mg2+ block of a large-conductance, Ca2+-activated K+ channel in rat skeletal-muscle – Ca2+, Sr2+, and Ni2+ also block10.1085/jgp.98.1.163 ↗
40. Measurement of cytosolic free calcium concentration in isolated rat ventricular myocytes with quin 210.1161/01.RES.55.6.830 ↗
41. Calcium, magnesium, and growth control in the WI-38 human fibroblast cell10.1083/jcb.83.2.487 ↗
42. LTRPC7 is a Mg·ATP-regulated divalent cation channel required for cell viability10.1038/35079092 ↗
43. Mammalian MagT1 and TUSC3 are required for cellular magnesium uptake and vertebrate embryonic development10.1073/pnas.0908332106 ↗
44. An in vitro mechanism study on the proliferation and pluripotency of human embryonic stems cells in response to magnesium degradation10.1371/journal.pone.0076547 ↗
45. Magnesium sulfate in eclampsia and pre-eclampsia: pharmacokinetic principles10.2165/00003088-200038040-00002 ↗
46. In vitro and in vivo corrosion measurements of magnesium alloys10.1016/j.biomaterials.2005.07.037 ↗
47. In vivo evaluation of biodegradable magnesium alloy bone implant in the first 6 months implantation10.1002/jbm.a.32132 ↗
48. Magnesium alloys for temporary implants in osteosynthesis: in vivo studies of their degradation and interaction with bone10.1016/j.actbio.2011.11.008 ↗
49. FDA. Use of International Standard ISO- 10993, Biological evaluation of medical devices part 1: evaluation and testing. Draft guidance for industry and food and drug administration staff. U.S. Department of Health and Human Services: Food and Drug Administration; 2013.
50. Evaluation of short-term effects of rare earth and other elements used in magnesium alloys on primary cells and cell lines10.1016/j.actbio.2009.09.024 ↗