A Method for Determining Detector Efficiencies In-Situ by Variation of the Radiation Incidence Angle
Source abstract
Abstract
The method of semiconductor detector calibration by variation of the radiation incidence angle is discussed in detail and further developed with regard to the use of particle or radiation beam excited samples as radiation sources. Special aspects introduced by this generalization of the theory, such as geometrical considerations or the choice of appropriate instrumental settings, are discussed in detail to provide the user with a comprehensive and ready-to-use theory of this method. The model described combines the advantages of the method of radiation incidence angle variation, viz., that neither a reference detector nor calibrated radiation standards are needed, with those offered by the possibility to use radiation usually generated in the instrument, e.g. electron impact caused radiation in an electron microprobe, for the determination of the detector efficiency.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Electron and X-Ray Spectroscopy Techniques · X-ray Spectroscopy and Fluorescence Analysis · Nuclear Physics and Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References
Reference metadata was not supplied for this record.