Quantifying Moisture Penetration in Encapsulated Devices by Heavy Water Mass Spectrometry: A Standard Moisture Leak Using Poly(ether-ether-ketone)
Source abstract
Abstract
Moisture penetration into active biomedical implants such as the bionic ear and eye is a major problem in healthcare since surgery is required to replace devices affected by corrosion. Existing methods for measuring moisture leak rates such as the commercially available dynamic relative humidity method are not sufficiently sensitive to guarantee security against moisture penetration. Helium leak detection is highly sensitive but is challenged by the unknown relation to the moisture leak rate because of mixed flow modes involving liquid water. A standard moisture leak traceable to fundamental units is not currently available, preventing direct comparison of moisture and helium leak rates in the same device. Here, we demonstrate a practical calibrated moisture leak based on the stable polymer poly(ether-ether-ketone), for calibrating heavy water mass spectrometry. Using biomedical test structures from manufactured encapsulations, we show that in the majority of cases, calibrated measurements of molar moisture leak rates exceed the helium leak rate, especially for very small and large leaks. Comparison with theory shows that LaPlace pressure is the driving force for the enhanced moisture flows. We recommend that the compliance limit for helium testing in biomedical devices be reduced by one order of magnitude.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Nanopore and Nanochannel Transport Studies · Fuel Cells and Related Materials · Microfluidic and Capillary Electrophoresis Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
National Measurement Institute, 36 Bradfield Road, West Lindfield, New South Wales 2071, Australia
School of Physics, University of Sydney, New South Wales 2006, Australia
ANSTO, New Illawarra Road, Lucas Heights, New South Wales 2232, Australia
Funding
Australian Research Council
References (51)
Show source-provided references
1. 10.1109/EMBC.2012.634654510.1109/EMBC.2012.6346545 ↗
2. 10.1016/j.amjoto.2017.12.01410.1016/j.amjoto.2017.12.014 ↗
3. 10.3390/mi1008050810.3390/mi10080508 ↗
4. Jiang G. · 2010
5. 10.1109/tcpmt.2020.296812610.1109/tcpmt.2020.2968126 ↗
6. 10.1002/lary.2414910.1002/lary.24149 ↗
7. AQUATRAN Model 3, Water Vapor Permeation Analyzer, MOCON, MN, USA. https://www.ametekmocon.com/products/permeationanalyzers/wvtr-permeation-analyzers/aquatran-3 (accessed July 2020).
8. Caironi M. · 201510.1002/9783527679973 ↗
9. 10.1063/1.180460210.1063/1.1804602 ↗
10. 10.1016/j.jcis.2005.06.03710.1016/j.jcis.2005.06.037 ↗
11. 10.1073/pnas.113684410010.1073/pnas.1136844100 ↗
12. 10.1209/0295-5075/125/1400310.1209/0295-5075/125/14003 ↗
13. 10.1103/physrevfluids.4.10390110.1103/physrevfluids.4.103901 ↗
14. Greenhouse H. · 1999
15. 10.1021/nl072385q10.1021/nl072385q ↗
16. 10.1038/438044a10.1038/438044a ↗
17. 10.1021/nn200222g10.1021/nn200222g ↗
18. 10.1038/s41565-017-0031-910.1038/s41565-017-0031-9 ↗
19. 10.1007/s10404-010-0612-510.1007/s10404-010-0612-5 ↗
20. 10.1103/physrevlett.102.18450210.1103/physrevlett.102.184502 ↗
21. 10.1016/j.ijthermalsci.2009.07.00810.1016/j.ijthermalsci.2009.07.008 ↗
22. 10.1126/science.112629810.1126/science.1126298 ↗
23. 10.1021/la200995r10.1021/la200995r ↗
24. 10.1021/acsnano.9b0432810.1021/acsnano.9b04328 ↗
25. 10.1111/j.1525-1594.2011.01222.x10.1111/j.1525-1594.2011.01222.x ↗
26. García-Blanco S. M. · 2012 · 82500H
27. 10.1021/acs.langmuir.5b0218310.1021/acs.langmuir.5b02183 ↗
28. 10.1088/0034-4885/79/2/02590110.1088/0034-4885/79/2/025901 ↗
29. 10.1016/0021-9797(86)90407-810.1016/0021-9797(86)90407-8 ↗
30. 10.1016/j.desal.2010.04.04210.1016/j.desal.2010.04.042 ↗
31. 10.1016/j.biomaterials.2007.07.01310.1016/j.biomaterials.2007.07.013 ↗
32. 10.1016/j.actbio.2006.11.00510.1016/j.actbio.2006.11.005 ↗
33. Kurtz S. M. · 2019
34. 10.1016/j.surfin.2017.09.00910.1016/j.surfin.2017.09.009 ↗
35. 10.1016/j.surfcoat.2017.12.05610.1016/j.surfcoat.2017.12.056 ↗
36. 10.1021/j100624a02510.1021/j100624a025 ↗
37. 10.1016/j.orgel.2014.10.01410.1016/j.orgel.2014.10.014 ↗
38. 10.1016/j.tsf.2015.08.03810.1016/j.tsf.2015.08.038 ↗
39. Standard Test Method for Water Vapor Transmission Rate through Plastic Film and Sheeting Using a Modulated Infrared Sensor; ASTM International: West Conshohocken, PA, United States., 2001.
40. McKeen L. W. · 2012
41. Kristen S. · 2014
42. Dahan, N. The application of PEEK to the packaging of implantable electronic devices. Ph.D. Thesis, University College London, 2013.
43. 10.1063/1.167468710.1063/1.1674687 ↗
44. 10.1116/1.57962210.1116/1.579622 ↗
45. 10.1021/acs.jpcc.5b0624110.1021/acs.jpcc.5b06241 ↗
46. 10.1021/nl200843g10.1021/nl200843g ↗
47. 10.1016/s0001-8686(98)00053-010.1016/s0001-8686(98)00053-0 ↗
48. 10.1021/acs.jpclett.6b0102110.1021/acs.jpclett.6b01021 ↗
49. 10.1006/jcis.1999.659710.1006/jcis.1999.6597 ↗
50. 10.1103/physreve.79.06730110.1103/physreve.79.067301 ↗
51. 10.1039/c5cp01862e10.1039/c5cp01862e ↗