journal article·2024-02Open access

Synchrotron microtomography reveals insights into the degradation kinetics of bio-degradable coronary magnesium scaffolds

Roman Menze, Bernhard Hesse, Maciej Kusmierczuk, Duote Chen, Timm Weitkamp, Stephanie Bettink, Bruno Scheller

Abstract

Bioresorbable magnesium scaffolds are a promising future treatment option for coronary artery stenosis, especially for young adults. Due to the degradation of these scaffolds (<1 year), long-term device-related clinical events could be reduced compared to treatments with conventional drug eluting stents. First clinical trials indicate a return of vasomotion after one year, which may be associated with improved long-term clinical outcomes. However, even after decades of development, the degradation process, ideal degradation time and biological response in vivo are still not fully understood. The present study investigates the in vivo degradation of magnesium scaffolds in the coronary arteries of pigs influenced by different strut thicknesses and the presence of antiproliferative drugs. Due to high 3D image contrast of synchrotron-based micro-CT with phase contrast (SR-μCT), a qualitative and quantitative evaluation of the degradation morphology of magnesium scaffolds was obtained. For the segmentation of the μCT images a convolutional network architecture (U-net) was exploited, demonstrating the huge potential of merging high resolution SR-μCT with deep learning (DL) supported data analysis. In total, 30 scaffolds, made of the rare earth alloy Resoloy®, with different strut designs were implanted into the coronary arteries of 10 domestic pigs for 28 days using drug-coated or uncoated angioplasty balloons for post-dilatation. The degradation morphology was analyzed using scanning electron microscopy, energy dispersive x-ray spectroscopy and SR-μCT. The data from these methods were then related to data from angiography, optical coherence tomography and histology. A thinner strut size (95 vs. 130 μm) and the presence of paclitaxel indicated a slower degradation rate at 28 d in vivo, which positively influences the late lumen loss (0.5 and 0.6 mm vs. 1.0 and 1.1 mm) and recoil values (0 and 1.7% vs. 6.1 and 22%).
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Coronary Interventions and Diagnostics · Electron and X-Ray Spectroscopy Techniques · Advanced Materials Characterization Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (57)

Show source-provided references
1. Comparison of acute thrombogenicity for magnesium versus stainless steel stents in a porcine arteriovenous shunt model10.4244/EIJ-D-17-00958 ↗
2. Preclinical evaluation of degradation kinetics and elemental mapping of first- and second-generation bioresorbable magnesium scaffolds10.4244/EIJ-D-17-00708 ↗
3. Comparison of an everolimus-eluting bioresorbable scaffold with an everolimus-eluting metallic stent for the treatment of coronary artery stenosis (ABSORB II): a 3 year, randomised, controlled, single-blind, multicentre clinical trial10.1016/S0140-6736(16)32050-5 ↗
4. Verheye · 2020
5. Sustained safety and performance of a second-generation sirolimus-eluting absorbable metal scaffold: long-term data of the BIOSOLVE-II first-in-man trial at 5 years10.1016/j.carrev.2021.07.017 ↗
6. Seguchi · 2023
7. Hijazi · 2015 · vol. 7 · 245
8. Alexy · 2013
9. Zartner · 2020
10. Quantitative characterization of degradation processes in situ by means of a bioreactor coupled flow chamber under physiological conditions using time-lapse SRμCT10.1002/maco.201709514 ↗
11. Magnesium degradation observed in situ under flow by synchrotron radiation based microtomography10.1117/12.2241085 ↗
12. Magnesium degradation under physiological conditions – best practice
13. Pfeiffer · X-ray sources · 2006 · 2 · 258
14. Biodegradable magnesium fixation screw for barrier membranes used in guided bone regeneration
15. Biodegradation of a magnesium alloy fixation screw used in a guided bone regeneration model in beagle dogs
16. Priors for X-ray in-line phase tomography of heterogeneous objects
17. Lienkamp, thomas brox OR. 3D U_net
18. Medical image computing and computer-assisted intervention - MICCAI 2015: 18th international conference Munich, Germany, october 5-9, 2015 proceedings, part III. Lect. Notes comput
19. Jung · 2021
20. Scaling the U - net : segmentation of biodegradable bone implants in high - resolution synchrotron radiation microtomograms
21. Hara · 2006 · vol. 58 · 377
22. Drug-eluting or bare-metal stents for coronary artery disease10.1056/NEJMoa1607991 ↗
23. Comparison of acute thrombogenicity for metallic and polymeric bioabsorbable scaffolds: Magmaris versus absorb in a porcine arteriovenous shunt model10.1161/CIRCINTERVENTIONS.116.004762 ↗
24. Three-year outcomes with the absorb bioresorbable scaffold10.1161/CIRCULATIONAHA.117.031843 ↗
25. Fully bioresorbable vascular scaffolds: lessons learned and future directions10.1038/s41569-018-0124-7 ↗
26. In-stent restenosis : contributions of inflammatory responses and arterial injury to neointimal hyperplasia10.1016/S0735-1097(97)00450-6 ↗
27. The tomography beamline ANATOMIX at Synchrotron SOLEIL10.1088/1742-6596/849/1/012037 ↗
28. Microtomography on the ANATOMIX beamline at synchrotron SOLEIL
29. The PyHST2 hybrid distributed code for high speed tomographic reconstruction with iterative reconstruction and a priori knowledge capabilities10.1016/j.nimb.2013.09.030 ↗
30. Scientific · 2002 · vol. 206 · 33
31. In vitro and in vivo evaluation of a novel bioresorbable magnesium scaffold with different surface modifications
32. The degradation and transport mechanism of a Mg-Nd-Zn-Zr stent in rabbit common carotid artery: a 20-month study10.1016/j.actbio.2018.01.018 ↗
33. Enhanced efficacy of sirolimus-eluting bioabsorbable magnesium alloy stents in the prevention of restenosis10.1583/10-3353.1 ↗
34. Macrophage phagocytosis of biomedical Mg alloy degradation products prepared by electrochemical method10.1016/j.msec.2017.02.126 ↗
35. Assessing the long-term in vivo degradation behavior of magnesium alloys - a high resolution synchrotron radiation micro computed tomography study10.3389/fbiom.2022.925471 ↗
36. Corrosion bending fatigue of RESOLOY® and WE43 magnesium alloy wires
37. Influence of solution heat treatment on the microstructure, hardness and stress corrosion behavior of extruded Resoloy10.1007/s11837-020-04077-9 ↗
38. Microstructure and fracture toughness of an extruded Mg-Dy-Nd-Zn-Zr alloy influenced by heat treatment
39. In vivo and in vitro evaluation of a biodegradable magnesium vascular stent designed by shape optimization strategy10.1016/j.biomaterials.2019.119414 ↗
40. Experimental data confirm numerical modeling of the degradation process of magnesium alloys stents10.1016/j.actbio.2012.10.035 ↗
41. Bioresorbable drug-eluting magnesium-alloy scaffold: design and feasibility in a porcine coronary model10.4244/EIJV8I12A218 ↗
42. In vivo degradation of magnesium alloy LA63 scaffolds for temporary stabilization of biological myocardial grafts in a swine model10.1515/bmt-2012-0047 ↗
43. Inhibition of vascular endothelial growth factor-mediated neointima progression with angiostatin or paclitaxel10.1016/S1051-0443(07)61847-8 ↗
44. Sirolimus and paclitaxel on polymer-based drug-eluting stents: similar but different10.1016/j.jacc.2005.09.047 ↗
45. Surface biodegradation behavior of rare earth- containing magnesium alloys with different microstructure: the impact on apatite coating formation on the surface10.1080/10667857.2018.1462025 ↗
46. Degradation behaviour of magnesium-rare earth biomedical alloys10.1080/10667857.2016.1213550 ↗
47. Recent advances on development of hydroxyapatite coating on biodegradable magnesium alloys: a review
48. Grain structure dependent self-assembled bioactive coating on Mg-2Zn-2Gd alloy: mechanism of degradation at biointerfaces10.1016/j.surfcoat.2017.02.052 ↗
49. In vivo corrosion and corrosion protection of magnesium alloy LAE44210.1016/j.actbio.2009.10.012 ↗
50. Impact of stent strut design in metallic stents and biodegradable scaffolds10.1016/j.ijcard.2014.09.143 ↗
51. Hemodynamically driven stent strut design10.1007/s10439-009-9719-9 ↗
52. Degradation mechanism of magnesium alloy stent under simulated human micro-stress environment10.1016/j.msec.2017.12.001 ↗
53. Absorbable magnesium-based stent: physiological factors to consider for in vitro degradation assessments10.1093/rb/rbu015 ↗
54. Flow-induced corrosion behavior of absorbable magnesium-based stents10.1016/j.actbio.2014.08.034 ↗
55. Early collapse of a magnesium bioresorbable scaffold10.1016/j.jcin.2017.07.037 ↗
56. Early collapse causing stenosis in a resorbable magnesium scaffold
57. Restenosis in Magmaris stents due to significant collapse10.1016/j.jcin.2018.02.040 ↗