Ultrabroad Dynamic All-Solid-State Radiation Dose Detector Based on a 0D Cs3Cu2I5 Perovskite-Like Single Crystal

Xiaojing Song, Linyue Liu, Pengying Wan, Xiao Ouyang, Tao Bai, Qian Wang, Runlong Gao, Rui Jia, Qiang Gao, Yuntao Wu, Wei Zheng, Xiaoping Ouyang

ACS Applied Electronic Materials· Volume 5· Issue 12
Publisher page

Abstract

Radiation dose detectors based on a gas-filled detecting mode play an important role in cancer radiation therapy, nuclear accident early warning, and radiation protection but still suffer from frequent air density corrections, a high working voltage, and a long stability time. Therefore, to tackle these issues, the development of high-performance all-solid-state radiation dose detectors is urgently needed. Here, we demonstrated a prototype of a highly sensitive all-solid-state radiation dose detector based on a highly efficient Cs 3 Cu 2 I 5 perovskite-like single crystal and a Si photodetector with a heterojunction structure. The Cs 3 Cu 2 I 5 -based detector exhibits an ultrabroad dynamic range of 11.45 mGy·h –1 to 107.3 Gy·h –1, covering nearly 5 orders of magnitude for γ-ray with the upper limit being 100 times higher than that of Cs 3 Cu 2 I nanocrystal of 0.846 Gy·h –1 and the detection limit approaching the lower limit of the radiotherapy level (10 mGy·h –1 ). Moreover, it has an outstanding linear response with a linear correlation coefficient ( R 2 ) of 0.9999 for the X/γ-ray incident dose rate and a good repeatable response deviation of less than 0.3% for γ-rays, which is comparable to classical ionization chamber detectors. This work opens a horizon of developing ultrabroad dynamic, highly sensitive, and low-power consumption all-solid-state radiation dose detectors based on perovskite-like single-crystal scintillators.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Radiation Detection and Scintillator Technologies · Luminescence Properties of Advanced Materials · Perovskite Materials and Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

School of Nuclear Science and Technology, Xi’an Jiaotong University, Xi’an 710049, China

Northwest Institute of Nuclear Technology, Xi’an 710024, China

School of Microelectronics, Xi’an Jiaotong University, Xi’an 710049, China

Key Laboratory of Beam Technology of Ministry of Education, College of Nuclear Science and Technology, Beijing Normal University, Beijing 100875, China

Shanghai Institute of Ceramics, Chinese Academy of Sciences, Shanghai 201899, China

Sino-French Institute of Nuclear Engineering and Technology, Sun Yat-Sen University, Zhuhai 519082, China

Institute of Microelectronics, Chinese Academy of Sciences, Beijing 100029, China

School of Materials, Sun Yat-Sen University, Guangzhou 510275, China

Funding

Natural Science Foundation of Shanghai Municipality · National Key Research and Development Program of China · China Institute of Atomic Energy · National Natural Science Foundation of China · Chinese Academy of Sciences

References (44)

Show source-provided references
1. 10.2298/NTRP0902145L10.2298/NTRP0902145L ↗
2. 10.1118/1.475806510.1118/1.4758065 ↗
3. 10.1080/00223131.2012.66001410.1080/00223131.2012.660014 ↗
4. 10.1016/j.radphyschem.2018.12.02910.1016/j.radphyschem.2018.12.029 ↗
5. 10.1016/j.radphyschem.2018.08.02810.1016/j.radphyschem.2018.08.028 ↗
6. 10.1364/CLEOPR.2018.W3A.7410.1364/CLEOPR.2018.W3A.74 ↗
7. 10.1080/00223131.2008.1087587010.1080/00223131.2008.10875870 ↗
8. 10.1016/j.radphyschem.2017.05.00610.1016/j.radphyschem.2017.05.006 ↗
9. 10.1007/s11433-022-2021-610.1007/s11433-022-2021-6 ↗
10. International Electrotechnical Commission, IEC 60731:2011+AMD1:2016 CSV. Medical Electrical Equipment-Dosimeters with Ionization Chambers as Used in Radiotherapy; IEC: Switzerland, 2016, 15–19.
11. 10.1007/s10512-020-00669-510.1007/s10512-020-00669-5 ↗
12. 10.1007/s12648-021-02210-110.1007/s12648-021-02210-1 ↗
13. 10.1016/j.nima.2019.16312010.1016/j.nima.2019.163120 ↗
14. 10.1016/j.radmeas.2009.09.01510.1016/j.radmeas.2009.09.015 ↗
15. 10.1016/j.nima.2019.16314110.1016/j.nima.2019.163141 ↗
16. 10.1002/adma.20190471110.1002/adma.201904711 ↗
17. 10.1117/12.260117510.1117/12.2601175 ↗
18. 10.1002/adom.20180067410.1002/adom.201800674 ↗
19. 10.1038/s41566-020-00727-110.1038/s41566-020-00727-1 ↗
20. 10.1021/acsami.0c1032310.1021/acsami.0c10323 ↗
21. 10.1002/adom.20220215710.1002/adom.202202157 ↗
22. 10.1002/pssr.20200037410.1002/pssr.202000374 ↗
23. 10.1002/adom.20210046010.1002/adom.202100460 ↗
24. 10.1002/pssr.20210042210.1002/pssr.202100422 ↗
25. 10.1021/acsanm.2c0199610.1021/acsanm.2c01996 ↗
26. 10.1002/adom.20220030410.1002/adom.202200304 ↗
27. 10.1021/acs.jpcc.2c0345310.1021/acs.jpcc.2c03453 ↗
28. 10.1002/advs.20200019510.1002/advs.202000195 ↗
29. 10.1021/acs.jpclett.2c0045610.1021/acs.jpclett.2c00456 ↗
30. 10.1364/PRJ.41295910.1364/PRJ.412959 ↗
31. 10.1021/acsami.0c0904710.1021/acsami.0c09047 ↗
32. 10.1016/j.nima.2020.16496310.1016/j.nima.2020.164963 ↗
33. 10.1002/adfm.20191064810.1002/adfm.201910648 ↗
34. 10.1002/adfm.20210914910.1002/adfm.202109149 ↗
35. 10.1021/acsami.1c2105510.1021/acsami.1c21055 ↗
36. 10.1038/s41566-017-0012-410.1038/s41566-017-0012-4 ↗
37. 10.1038/s41427-021-00308-w10.1038/s41427-021-00308-w ↗
38. 10.1002/adma.20190440510.1002/adma.201904405 ↗
39. Wang, K.; Yang, X. Y.; Hu, J. C. JJG 2044–2010 Metrological Instruments for Air Kerma of γ Rays; AQSIQ: CH 2010, 2–5.
40. Hu, J. C.; Zhang, H.; Lv, Y. Z.; Hu, S. Z.; Guo, H. T.; Yang, Y. JJG 912–2010 Dosimeters with Ionization Chambers as Used in Radiotherapy; AQSIQ: CH 2010, 2–8.
41. 10.1016/j.nima.2021.16589410.1016/j.nima.2021.165894 ↗
42. 10.1021/acs.jpclett.0c0116610.1021/acs.jpclett.0c01166 ↗
43. Hubbell, J. H.; Seltzer, S. M. NISTIR 5632 Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients 1 keV to 20 MeV for Elements Z = 1 to 92 and 48 Additional Substances of Dosimetric Interest; NIST: US 1995, 24–54.10.6028/NIST.IR.5632 ↗
44. 10.1109/TNS.2021.306956810.1109/TNS.2021.3069568 ↗