journal article·2026-05-02Open access

Mechanical Fatigue of Titanium Dental Implants After Implantoplasty: An In Vitro Study Combined with Finite Element Simulations

Esteban Padullés-Roig, Pablo Sevilla, Eugenio Velasco-Ortega, Miguel Cerrolaza, Darcio Fonseca, Jeanne Parache, Conrado Aparicio, Javier Gil

Abstract

The increasing prevalence of peri-implantitis has led to a growing clinical use of implantoplasty, a procedure involving intraoral machining of the dental implant surface to remove biofilm. The absence of standardized clinical protocols may contribute to premature fatigue failure of dental implants. The present study aimed to evaluate the influence of machining depth on the cyclic mechanical behavior of dental implants. A total of 250 commercially pure grade 4 titanium dental implants were distributed into four groups according to machining depth: untreated (original), 0.2 mm, 0.4 mm, and 0.6 mm wall reduction. The implant system featured an internal connection with a thread height of 0.4 mm. Finite element analysis was performed for each machining depth to evaluate von Mises stress distribution and simulate fatigue behavior. The numerical models were validated through experimental fatigue testing using a servo-hydraulic MTS Bionix testing machine under ISO 14801:2016 conditions, showing a high correlation between simulated and experimental results (correlation coefficients > 0.9). The results indicated that maximum von Mises stresses were concentrated at the junction between the implant thread and the implant body. The fatigue limit of the untreated implants was approximately 351 N. Implants subjected to 0.4 mm machining exhibited a fatigue limit of 301 N, whereas lower fatigue limits were observed for 0.2 mm (255 N) and 0.6 mm (185 N) reductions. These findings suggest a significant mechanical effect of thread removal: 0.4 mm implantoplasty may provide improved fatigue performance compared to 0.2 mm, potentially due to reduced stress concentration at the thread–body junction. At high applied loads, fracture occurred in the coronal region of the implant, whereas at lower loads failure shifted to the implant–abutment connection. Although a good agreement between numerical and experimental results was observed, these findings should be interpreted with caution due to the in vitro testing conditions and the assumptions inherent to the finite element simulations. Therefore, while the results suggest that implantoplasty depth should not exceed the original thread height, further validation under clinically relevant conditions is required to confirm its impact on long-term mechanical reliability.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Dental Implant Techniques and Outcomes · Dental materials and restorations · Titanium Alloys Microstructure and Properties

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Faculty of Dentistry, Universidad Internacional de Catalunya, Josep Trueta s/n, Sant Cugat del Vallés, 08195 Barcelona, Spain

Bioinspired Oral Biomaterials and Interfaces, Department Ciencia e Ingenieria de Materiales, Escola Enginyeria Barcelona Est, Universitat Politècnica de Catalunya, Av. Eduard Maristany 16, 08019 Barcelona, Spain

Mechanics Department, Escuela Universitaria Salesiana de Sarria, Pg. Sant Joan Bosco, 74, 74, 08017 Barcelona, Spain

Department of Stomatology, Faculty of Dentistry, University of Seville, c/Avicena s/n, 41009 Sevilla, Spain

School of Engineering, Science & Technology, Valencian International University, Calle Pintor Sorolla, 21, 46002 Valencia, Spain

ICREA, Institució Catalana de Recerca i Estudis Avançats, Generalitat de Catalunya, 08010 Barcelona, Spain

References (56)

Show source-provided references
1. Non-surgical treatment of peri-implant mucositis and peri-implantitis: A literature review10.1111/j.1600-051X.2008.01276.x ↗
2. Peri-implantitis10.1111/jcpe.12954 ↗
3. Efficacy of reconstructive surgical therapy at peri-implantitis-related bone defects. A systematic review and meta-analysis10.1111/jcpe.13070 ↗
4. Efficacy of alternative or adjunctive measures to conventional treatment of peri-implant mucositis and peri-implantitis: A systematic review and meta-analysis10.1186/s40729-015-0023-1 ↗
5. Principles of Combined Surgical Therapy for the Management of Peri-Implantitis10.1002/cap.10186 ↗
6. Diaz, P., Gonzalo, E., Villagra, L.J.G., Miegimolle, B., and Suarez, M.J. (2022). What is the prevalence of periimplantitis? A systematic review and meta-analysis. BMC Oral Health, 22.10.1186/s12903-022-02493-8 ↗
7. Thakkar, R. (2025, January 12). Trends in Dental Implants 2022. Available online: https://connect.aaid-implant.org/blog/trends-in-dental-implants-2022.
8. Impact of the Habit of Alcohol Consumption on the Success of the Implants: A Retrospective Study10.4103/jpbs.jpbs_430_23 ↗
9. The influence of the gingival phenotype on implant survival rate and clinical parameters: A systematic review10.1038/s41432-025-01114-x ↗
10. Revisiting peri-implant diseases in order to rethink the future of compromised dental implants: Considerations, perspectives, treatment, and prognosis10.17219/dmp/187215 ↗
11. Advancing peri-implantitis treatment: A scoping review of breakthroughs in implantoplasty and Er:YAG laser therapies10.1002/cre2.70104 ↗
12. Surgical and Nonsurgical Treatment Protocols for Peri-implantitis: An overview of Systematic Reviews10.11607/jomi.9659 ↗
13. Resolution of peri-implantitis by means of implantoplasty as adjunct to surgical therapy: A retrospective study10.1002/JPER.21-0103 ↗
14. Does implantoplasty affect the failure strength of narrow and regular diameter implants? A laboratory study10.1007/s00784-020-03534-8 ↗
15. Toledano-Serrabona, J., Gil, F.J., Camps-Font, O., Valmaseda-Castellón, E., Gay-Escoda, C., and Sánchez-Garcés, M.Á. (2021). Physicochemical and Biological Characterization of Ti6Al4V Particles Obtained by Implantoplasty: An In Vitro Study. Part I. Materials, 14.10.3390/ma14216507 ↗
16. Padulles-Gaspar, E., Padulles-Roig, E., Cabanes, G., Pérez, R.A., Gil, J., and Bosch, B.M. (2023). Effects of Hypochlorous Acid and Hydrogen Peroxide Treatment on Bacterial Disinfection Treatments in Implantoplasty Procedures. Materials, 16.10.3390/ma16082953 ↗
17. Evaluation of the inflammatory and osteogenic response induced by titanium particles released during implantoplasty of dental implants10.1038/s41598-022-20100-2 ↗
18. Particle release from implantoplasty of dental implants and impact on cells10.1186/s40729-020-00247-1 ↗
19. Dynamic assessment of titanium surface oxides following mechanical damage reveals only partial passivation under inflammatory conditions10.1038/s41529-024-00514-1 ↗
20. Electrolytic Cleaning and Regenerative Therapy of Peri-implantitis in the Esthetic Area: A Case Report10.1055/s-0042-1750773 ↗
21. Effect of Electrolytic Cleaning on Mechanical Properties for Titanium Dental Implants with Surface Contamination
22. The effect of antimicrobial therapy on periimplantitis lesions. An experimental study in the dog10.1034/j.1600-0501.1996.070404.x ↗
23. Giok, K.C., and Menon, R.K. (2023). The Microbiome of Peri-Implantitis: A Systematic Review of Next-Generation Sequencing Studies. Antibiotics, 12.10.3390/antibiotics12111610 ↗
24. Prevention and treatment of peri-implant diseases—The EFP S3 level clinical practice guideline10.1111/jcpe.13823 ↗
25. (2026, April 10). University of British Columbia. Available online: http://www.stat.ubc.ca/~rollin/stats/ssize/n2.html.
26. (2016). Dentistry—Implants—Dynamic Loading Test for Endosseous Dental Implants (Standard No. ISO 14801:2016).
27. Callister, W.D., and Rethwisch, D.G. (2018). Materials Science and Engineering: An Introduction, Wiley. [10th ed.].
28. Ramberg, W., and Osgood, W.R. (1943). Description of Stress-Strain Curves by Three Parameters, National Advisory Committee for Aeronautics. Technical Note No.902.
29. Mechanical properties of biomedical titanium alloys10.1016/S0921-5093(97)00806-X ↗
30. Budynas, R.G., and Nisbett, J.K. (2019). Shigley’s Mechanical Engineering Design, McGraw-Hill. [11th ed.].
31. Working stresses10.1115/1.4008629 ↗
32. Walczak, M., Okuniewski, W., Nowak, W.J., Chocyk, D., and Pasierbiewicz, K. (2025). Corrosion Behavior of Shot Peened Ti6Al4V Alloy Fabricated by Conventional and Additive Manufacturing. Materials, 18.10.3390/ma18102274 ↗
33. Can degradation products released from dental implants affect peri-implant tissues?
34. Acceleration of apatite nucleation on microrough bioactive titanium for bone-replacing implants10.1002/jbm.a.31164 ↗
35. Implant surface modifications and their impact on osseointegration and peri-implant diseases through epigenetic changes: A scoping review10.1111/jre.13273 ↗
36. The effect of surface roughness on early in vivo plaque colonization on titanium10.1902/jop.1997.68.6.556 ↗
37. Marin, E., and Lanzutti, A. (2024). Biomedical Applications of Titanium Alloys: A Comprehensive Review. Materials, 17.10.3390/ma17010114 ↗
38. Toledano-Serrabona, J., Sánchez-Garcés, M.A., Gay-Escoda, C., Valmaseda-Castellon, E., Camps-Font, O., Verdeguer, P., Molmeneu, M., and Gil, F.J. (2021). Mechanical properties and corrosión behavior of Ti6Al4V particles obtained by Implatoplasty. An in vivo study. Part. II. Materials, 14.10.3390/ma14216519 ↗
39. Clinical management of implant prostheses in patients with bruxism10.1155/2012/369063 ↗
40. Effect of surface treatments on the fatigue life of titanium for biomedical applications10.1016/j.jmbbm.2010.03.006 ↗
41. Effect of implant connection and restoration design (screwed vs. cemented) in reliability and failure modes of anterior crowns10.1111/j.1600-0722.2011.00837.x ↗
42. Osseintegration following treatment of peri-implantitis and replacement of implant components. An experimental study in the dog10.1034/j.1600-051x.2001.028003258.x ↗
43. Effects of tensile and compressive stress on bone resorption and formation parameters surrounding dental implants10.1016/j.jmbbm.2025.106928 ↗
44. Pereira, R., Maia, P., Rios-Santos, J.V., Herrero-Climent, M., Rios-Carrasco, B., Aparicio, C., and Gil, J. (2024). Influence of Titanium Surface Residual Stresses on Osteoblastic Response and Bacteria Colonization. Materials, 17.10.3390/ma17071626 ↗
45. Efficacy of implantoplasty in management of peri-implantitis: A systematic review10.4103/jips.jips_102_23 ↗
46. Two to six-year disease resolution and marginal bone stability rates of a modified resective-implantoplasty therapy in 32 peri-implantitis cases10.1111/cid.12773 ↗
47. Rondone, E.M., Leitão-Almeida, B., Pereira, M.S., Fernandes, G.V.O., and Borges, T. (2024). The Use of Tissue Grafts Associated with Immediate Implant Placement to Achieve Better Peri-Implant Stability and Efficacy: A Systematic Review and Meta-Analysis. J. Clin. Med., 13.10.3390/jcm13030821 ↗
48. Fonseca, D., de Tapia, B., Pons, R., Aparicio, C., Guerra, F., Messias, A., and Gil, J. (2024). The Effect of Implantoplasty on the Fatigue Behavior and Corrosion Resistance in Titanium Dental Implants. Materials, 17.10.3390/ma17122944 ↗
49. Klein, A., Rasperini, G., Gruber, R., Andrukhov, O., and Rausch-Fan, X. (2025). Electrolytic Cleaning of Dental Implants: A Scoping Review of Clinical Studies. Dent. J., 13.10.3390/dj13040172 ↗
50. Evidence of re-osseointegration after electrolytic cleaning and regenerative therapy of peri-implantitis in humans: A case report with four implants10.1007/s00784-021-04345-1 ↗
51. Schlee, M., Rathe, F., Brodbeck, U., Ratka, C., Weigl, P., and Zipprich, H. (2019). Treatment of Peri-implantitis-Electrolytic Cleaning Versus Mechanical and Electrolytic Cleaning-A Randomized Controlled Clinical Trial-Six-Month Results. J. Clin. Med., 8.10.3390/jcm8111909 ↗
52. Ambrosio, D., Caggiano, F., Acerra, M., Pisano, A., and Giordano, F. (2023). Is Ozone a Valid Adjuvant Therapy for Periodontitis and Peri-Implantitis? A Systematic Review. J. Pers. Med., 13.10.3390/jpm13040646 ↗
53. Fernandes, G.V.O., Martins, B.G.d.S., Fernandes, J.C.H., Gabet, Y., and Vizanski, A. (2025). The Novel iMPACT Tool and Quadrant Protocol for Peri-Implantitis: Surface Refinement and Re-Osseointegration Validated by SEM/EDS and Long-Term Clinical Case Reports. Medicina, 61.10.3390/medicina61061094 ↗
54. Eliaz, N. (2019). Corrosion of Metallic Biomaterials: A Review. Materials, 12.10.3390/ma12030407 ↗
55. Compressive fatigue in titanium dental implants submitted to fluoride ions action10.1590/S1678-77572007000400011 ↗
56. Corrosion and surface modification on biocompatible metals: A review10.1016/j.msec.2017.04.102 ↗