Finite element analysis of the foreshortening property of customized coronary stents
Source abstract
Abstract
Foreshortening is an important property of customized coronary stents,deciding the choice of stent and precise positioning of stent length.The primary aim of this investigation was to investigate the effects of varying customized stent design and expansion diameter on the foreshortening of stent by using the finite element method(FEM).The results show that L-stent forshortening was the best and the M-stent the worst.The amplitude and expansion diameter are the dominant parameter in this study.Increasing amplitude decreases forshortening rate.However,increasing expansion of stent increases forshortening rate of stent.Besides,strut thickness,strut width,amplitude or radius of curvature and the length of link influence the forshortening hardly.In conclusion,FEM can quantify some mechanical behaviors of stent and help surgeons to select stent and designers to optimize stent.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Coronary Interventions and Diagnostics
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References