Comparison of biocompatibility of electrochemical polished surface and anodic oxidized surface of TA4 pure titanium
Source abstract
Abstract
The surface morphology,corrosion resistance,fretting friction and wear properties and bioactivity of electrochemical polished surface and anodic oxidized surface on TA4 pure titanium were investigated.The results show that compared to the anodic oxidized surface,the electrochemical polished surface has more nano-scale cavities(fewer total cavitities),lower surface rougness(about an order lower),lower corrosion resistance,lower friction coefficient(0.134 vs 0.286),worse wear resistance,and higher bioactivity(growth rate of Ca-P layer in simulated body fluid).
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Titanium Alloys Microstructure and Properties · Bone Tissue Engineering Materials · Orthopaedic implants and arthroplasty
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References
Reference metadata was not supplied for this record.