The preparation of platelet glycoprotein monoclonal antibody coating stent based on ultrasonic atomization spray and its study on biocompatibility
Source abstract
Abstract
Platelet glycoprotein monoclonal antibody coating stent was prepared using ultrasonic atomization spray.Chitosan-monoclonal antibody coating parameters was optimized with coating uniformity and the size of contact angle.In order to control release rate of the monoclonal antibody SZ-21,levorotatory polylactic acid was coated as shielding layer.The blood compatibility of coating stent was tested with hemolysis rate.The cell compatibility of coating stent was investigated by MTT test and cell adhesion.Stents surface coating was analyzed by SEM and EDS,the results showed that polymer and monoclonal antibody was coated successful to the stent surface and the coating was uniform.The hemolysis rate of coating stent was well below the accepted threshold value of 5%.Cell toxicity test showed that the coating stent had no obvious side effects on vascular endothelial cell.The coating stent has good cell compatibility,which conduced to cell growth.Comparing with the monoclonal antibody coating stent that lacked shielding layer,the release of SZ-21in monoclonal antibody coating stent with shielding layer was significantly slow release,and there was no burst releases of massive amounts of drug.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Nanofabrication and Lithography Techniques · Coronary Interventions and Diagnostics · Polymer Surface Interaction Studies