Effect of coronary stent design parameters on the mechanical property of the elastic recoil of stent
Source abstract
Abstract
Coronary stent is an important device used in percutaneous transluminal coronary angioplasty(PTCA),which is usually implanted to support the stenosed artery.Low elastic recoil has become an essential feature of new coronary stents.The primary aim of this investigation was to investigate the effects of varying stent design on the elastic recoil of stent by using the finite element method(FEM).The finite element model to evaluate elastic recoil of stent was set up based on the stent prototype testing.The results showed that the amplitude and the number of crowns were the dominant parameter in this study.Increasing either strut width or thickness generally decreased elastic recoil of stent.However,increasing either amplitude or radius of curvature or the number of crowns generally increased it.In conclusion,FEM could quantify some mechanical behaviors of stent and help surgeons to select stent and designers to optimize stent.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Coronary Interventions and Diagnostics · Coronary Artery Anomalies
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References
Reference metadata was not supplied for this record.