Heat Generation and Transfer Behaviors of Ti-Coated Carbon Steel Rod Adaptable for Ablation Therapy of Oral Cancer
Source abstract
Abstract
For the purpose of developing a novel ablation therapy for oral cancer, the heat generation and transfer properties of a Ti-coated carbon steel rod with 20-mm length and 1.8-mm outer diameter were investigated by means of a high-frequency induction technique at 300 kHz. The heat generation measurement performed using water (15 mL) revealed that the difference of the inclination angles (θ = 0°, 45° and 90°) relative to the magnetic flux direction only slightly affects the heating behavior, exhibiting the overlapped temperature curves during an induction time of 1200 s. These results suggest that the effect of the shape magnetic anisotropy is almost eliminated, being convenient for the precise control of the ablation temperature in clinical use. In the experiments utilizing a tissue-mimicking phantom, the heat transfer concentrically occurred in the lateral direction for both the planar surface and a 10-mm deep cross-section. However, the former exhibited a considerably lower increase in temperature (ΔT), probably due to the effect of heat dissipation to the ambient air. No significant heat transfer was found to occur to the lower side of the inserted Ti-coated carbon steel rod, which is situated in the longitudinal direction.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Oral and Maxillofacial Pathology · Oral health in cancer treatment · Voice and Speech Disorders
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Graduate School of Science and Engineering, Ehime University, Matsuyama 790-8577, Japan
Department of Environmental Materials Engineering, Niihama National College of Technology, Niihama 792-8580, Japan
Department of Electronic Control Engineering, Niihama National College of Technology, Niihama 792-8580, Japan
Department of Surgery, Graduate School of Medicine, Ehime University, Toon 791-0295, Japan
References (22)
Show source-provided references
1. Oral cancer and precancerous lesions10.3322/canjclin.52.4.195 ↗
2. Oral and neck examination for early detection of oral cancer—A practical guide
3. Werning, J.W. (2007). Oral Cancer: Diagnosis, Management, and Rehabilitation, Thieme Medical Publishers, Inc.10.1055/b-002-85470 ↗
4. Oral cancer: The importance of early diagnosis and treatment10.2165/00128071-200102040-00005 ↗
5. Recent advances in oral oncology 2007: Imaging, treatment and treatment incomes10.1016/j.oraloncology.2008.01.006 ↗
6. Current surgical treatment of squamous cell carcinoma of the head and neck10.1016/j.oraloncology.2006.04.013 ↗
7. Dysphagia following head and neck cancer surgery10.1007/BF00301917 ↗
8. Davies, A., and Epstein, J. (2010). Oral Complications of Oral Cancer and Its Management, Oxford University Press.10.1093/med/9780199543588.001.0001 ↗
9. Oral mucositis in cancer therapy
10. Oral sequelae of head and neck radiotherapy10.1177/154411130301400305 ↗
11. Oral complication of cancer radiotherapy10.1080/00325481.1977.11712115 ↗
12. Concurrent chemotherapy and radiotherapy for head and neck cancer10.1586/14737140.9.3.293 ↗
13. Current chemotherapies for recurrent/metastatic head and neck cancer10.1097/CAD.0b013e3283421f7c ↗
14. Dental treatment consideration in the chemotherapy patient
15. Oral mucositis induced by anticancer treatments: Physiopathology and treatments10.2147/tcrm.2006.2.2.159 ↗
16. Hair loss as a consequence of cancer chemotherapy—Physical methods of prevention. A review of the literature
17. Development of Ti-coated ferromagnetic needle adaptable for ablation cancer therapy by high-frequency induction heating10.3390/jfb3010163 ↗
18. A new muscle-equivalent phantom for SAR estimation
19. Biological tissue-equivalent agar-based solid phantom and SAR estimation using the thermographic method in the range of 3–6 GHz10.1093/ietcom/e88-b.9.3733 ↗
20. Heat generation ability in AC magnetic field of needle-type Ti-coated mild steel for ablation cancer therapy10.1108/03321641111152739 ↗
21. Cullity, B.D., and Graham, C.D. (2008). Introduction to Magnetic Materials, Willey IEEE Press.10.1002/9780470386323 ↗
22. Minimally required heat doses for various tumor sizes in induction heating cancer therapy determined by computer simulation using experimental data10.3109/02656731003681028 ↗