journal article·2023-06-05Open access

Simulation and Experimental Investigation of Balloon Folding and Inserting Performance for Angioplasty: A Comparison of Two Materials, Polyamide-12 and Pebax

Tao Li, Zhuo Zhang, Wenyuan Wang, Aijia Mao, Yu Chen, Yan Xiong, Fei Gao

Abstract

Background: A balloon dilatation catheter is a vital tool in percutaneous transluminal angioplasty. Various factors, including the material used, influence the ability of different types of balloons to navigate through lesions during delivery. Objective: Thus far, numerical simulation studies comparing the impacts of different materials on the trackability of balloon catheters has been limited. This project seeks to unveil the underlying patterns more effectively by utilizing a highly realistic balloon-folding simulation method to compare the trackability of balloons made from different materials. Methods: Two materials, nylon-12 and Pebax, were examined for their insertion forces via a bench test and a numerical simulation. The simulation built a model identical to the bench test’s groove and simulated the balloon’s folding process prior to insertion to better replicate the experimental conditions. Results: In the bench test, nylon-12 demonstrated the highest insertion force, peaking at 0.866 N, significantly outstripping the 0.156 N force exhibited by the Pebax balloon. In the simulation, nylon-12 experienced a higher level of stress after folding, while Pebax had demonstrated a higher effective strain and surface energy density. In terms of insertion force, nylon-12 was higher than Pebax in specific areas. Conclusion: nylon-12 exerts greater pressure on the vessel wall in curved pathways when compared to Pebax. The simulated insertion forces of nylon-12 align with the experimental results. However, when using the same friction coefficient, the difference in insertion forces between the two materials is minimal. The numerical simulation method used in this study can be used for relevant research. This method can assess the performance of balloons made from diverse materials navigating curved paths and can yield more precise and detailed data feedback compared to benchtop experiments.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Coronary Interventions and Diagnostics · Soft Robotics and Applications · Elasticity and Material Modeling

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

College of Mechanical Engineering, Sichuan University, Chengdu 610065, China

Chengdu Neurotrans Medical Technology Co., Ltd., Chengdu 610065, China

Department of Applied Mechanics, Sichuan University, Chengdu 610065, China

Funding

Key Research Programs of Science & Technology Department of Sichuan Province · National Natural Science Research Foundation of China

References (41)

Show source-provided references
1. Evidence-based medical management of peripheral artery disease10.1161/ATVBAHA.119.312142 ↗
2. Drug-coated balloon versus standard percutaneous transluminal angioplasty for the treatment of superficial femoral and popliteal peripheral artery disease: 12-month results from the IN.PACT SFA rando10.1161/CIRCULATIONAHA.114.011004 ↗
3. Update on peripheral artery disease: Epidemiology and evidence-based facts10.1016/j.atherosclerosis.2018.05.033 ↗
4. Systematic review and updated meta-analysis of the use of drug-coated balloon angioplasty versus plain old balloon angioplasty for femoropopliteal arterial disease10.1016/j.jvs.2019.01.080 ↗
5. Lower limb multilevel treatment with drug-eluting balloons: 6-month results from the DEBELLUM randomized trial10.1583/JEVT-12-3926MR.1 ↗
6. The LEVANT I (Lutonix paclitaxel-coated balloon for the prevention of femoropopliteal restenosis) trial for femoropopliteal revascularization: First-in-human randomized trial of low-dose drug-coated balloon versus uncoated balloon angioplasty10.1016/j.jcin.2013.05.022 ↗
7. Paclitaxel-coated balloons reduce restenosis after femoro-popliteal angioplasty: Evidence from the randomized PACIFIER trial10.1161/CIRCINTERVENTIONS.112.971630 ↗
8. Drug-eluting balloon: Design, technology and clinical aspects10.1088/1748-605X/aaa0aa ↗
9. Comparative Investigation on Polymer Drug-coating Balloons Used in Infrapopliteal Angioplasty Based on Angiosomes Concept10.37358/MP.22.2.5597 ↗
10. Amstutz, C., Weisse, B., Haeberlin, A., Burger, J., and Zurbuchen, A. (2022). Inverse Finite Element Approach to Identify the Post-Necking Hardening Behavior of Polyamide 12 under Uniaxial Tension. Polymers, 14.10.3390/polym14173476 ↗
11. Numerical Methodology to Evaluate Trackability and Pushability of PTCA Balloon Catheter10.1007/s13239-022-00653-z ↗
12. Quality Determining Parameters of Balloon Angioplasty Catheters
13. (2013). Standard Guide for Measuring Securement of Balloon Expandable Vascular Stent Mounted on Delivery System. Standard No. ASTMF2394-07.
14. A comparison of the mechanical performance characteristics of seven drug-eluting stent systems10.1002/ccd.21832 ↗
15. Geith, M.A., Eckmann, J.D., Haspinger, D.C., Agrafiotis, E., Maier, D., Szabo, P., Sommer, G., Schratzenstaller, T.G., and Holzapfel, G.A. (2020). Experimental and mathematical characterization of coronary polyamide-12 balloon catheter membranes. PLoS ONE, 15.10.1371/journal.pone.0234340 ↗
16. Influence of balloon design, plaque material composition, and balloon sizing on acute post angioplasty outcomes: An implicit finite element analysis10.1002/cnm.3499 ↗
17. Mechanical performances of balloon post-dilation for improving stent expansion in calcified coronary artery: Computational and experimental investigations10.1016/j.jmbbm.2021.104609 ↗
18. Numerical Analysis for Non-Uniformity of Balloon-Expandable Stent Deployment Driven by Dogboning and Foreshortening10.1007/s13239-021-00573-4 ↗
19. Realistic finite element-based stent design: The impact of balloon folding10.1016/j.jbiomech.2007.08.014 ↗
20. Wiesent, L., Schultheiß, U., Schmid, C., Schratzenstaller, T., and Nonn, A. (2019). Experimentally validated simulation of coronary stents considering different dogboning ratios and asymmetric stent positioning. PLoS ONE, 14.10.1371/journal.pone.0224026 ↗
21. Finite element analysis of the percutaneous coronary intervention in a coronary bifurcation
22. Intravascular optical coherence tomography to validate finite-element simulation of angioplasty balloon inflation10.1088/1361-6560/ab0d58 ↗
23. The influence of the nylon balloon stiffness on the efficiency of the intra-aortic balloon occlusion10.1002/cnm.3173 ↗
24. On the importance of modeling balloon folding, pleating, and stent crimping: An FE study comparing experimental inflation tests. I10.1002/cnm.3249 ↗
25. Characterization of Pebax angioplasty balloon surfaces with AFM, SEM, TEM, and SAXS10.1002/jbm.b.33414 ↗
26. Characterization of polymeric biomedical balloon: Physical and mechanical properties10.1515/polyeng-2021-0203 ↗
27. Simulation of the Single Point Incremental Forming of Polyamide and Polyethylene Sheets10.1051/matecconf/201929003014 ↗
28. The influence of Young’s modulus of loaded implants on bone remodeling: An experimental and numerical study in the goat knee10.1002/jbm.a.32145 ↗
29. Hallquist, J.O. (2013). LS-DYNA® Keyword User’s Manual Volume II Material Models, Livermore Software Technology Corporation.
30. Zhang, K., Li, W., Zheng, Y., Yao, W., and Zhao, C. (2021). Compressive Properties and Constitutive Model of Semicrystalline Polyethylene. Polymers, 13.10.3390/polym13172895 ↗
31. Effect of time scaling and mass scaling in numerical simulation of incremental forming10.4028/www.scientific.net/AMM.612.105 ↗
32. A Rigid Multibody Model to Study the Translational Motion of Guidewires Based on Their Mechanical Properties10.1115/1.4043618 ↗
33. Salaha, Z.F.M., Ammarullah, M.I., Abdullah, N.N.A.A., Aziz, A.U.A., Gan, H.-S., Abdullah, A.H., Abdul Kadir, M.R., and Ramlee, M.H. (2023). Biomechanical Effects of the Porous Structure of Gyroid and Voronoi Hip Implants: A Finite Element Analysis Using an Experimentally Validated Model. Materials, 16.10.3390/ma16093298 ↗
34. Dong, S., Sheldon, A., Pydimarry, K., and Dapino, M. (2016, January 12–14). Friction in LS-DYNA®: Experimental Characterization and Modeling Application. Proceedings of the 14th International LSDYNA Users Conference, Detroit, MI, USA.
35. Determination of urethral catheter surface lubricity10.1007/s10856-007-3339-4 ↗
36. Comparison of angiographic dissection patterns caused by long vs short balloons during balloon angioplasty of chronic femoropopliteal occlusions10.1177/1526602818756610 ↗
37. Impact of total stent length after cobalt chromium everolimus-eluting stent implantation on 3-year clinical outcomes10.1002/ccd.26455 ↗
38. Paclitaxel-eluting balloons, paclitaxel-eluting stents, and balloon angioplasty in patients with restenosis after implantation of a drug-eluting stent (ISAR-DESIRE 3): A randomised, open-label trial10.1016/S0140-6736(12)61964-3 ↗
39. Computational analysis of mechanical stress–strain interaction of a bioresorbable scaffold with blood vessel10.1016/j.jbiomech.2016.05.035 ↗
40. Xiaodong, Z., Mitsuo, U., and Kiyotaka, I. (2021). Finite element analysis of cutting balloon expansion in a calcified artery model of circular angle 180°: Effects of balloon-to-diameter ratio and number of blades facing calcification on potential calcification fracturing and perforation reduction. PLoS ONE, 16.10.1371/journal.pone.0251404 ↗
41. A Novel Simulation Strategy for Stent Insertion and Deployment in Curved Coronary Bifurcations: Comparison of Three Drug-Eluting Stents10.1007/s10439-009-9836-5 ↗