Evaluating the Impact of Pontic Geometry on Load to Failure and Displacement in Implant-Supported Monolithic Zirconia Prostheses: An In Vitro Analysis
Source abstract
Abstract
The pontic design may influence the load-to-failure performance of fixed implant-supported screw-retained monolithic zirconia prostheses. This study aimed to evaluate the effect of pontic geometry on the fracture resistance of such restorations. Forty restorations were designed using dental CAD software and divided into four groups (n = 10 each): (A) Flat + Wide—pontics with a flat contour, 10 mm in width and 8 mm in height; (B) Concave + Wide—pontics with a concave contour, 10 mm in width and 5.5 mm in height; (C) Flat + Narrow—pontics with a flat contour, 6 mm in width and 8 mm in height; and (D) Concave + Narrow—pontics with a concave contour, 6 mm in width and 5.5 mm in height. All specimens underwent thermal and mechanical cycling, followed by a fracture load test using a three-point bending setup. Maximum fracture loads and displacements were analyzed using one-way ANOVA. Statistically significant differences were observed among the groups for both load to failure (p = 0.001) and displacement (p = 0.002). These findings indicate that pontic geometry significantly influences the fracture resistance and deformation behavior of monolithic zirconia prostheses.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Dental Implant Techniques and Outcomes · Dental materials and restorations · Orthopaedic implants and arthroplasty
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Faculty of Health Sciences, Alfonso X el Sabio University, 28691 Madrid, Spain
Department of Surgery, Faculty of Medicine, University of Salamanca, 37008 Salamanca, Spain
Department of Stomatology, Faculty of Medicine and Dentistry, University of Valencia, 46010 Valencia, Spain
References (65)
Show source-provided references
1. Systematic review of the survival rate and the incidence of biological, technical, and aesthetic complications of single crowns on implants reported in longitudinal studies with a mean follow-up of 5 years10.1111/j.1600-0501.2012.02547.x ↗
2. Gingival contour and clinical crown lengththeir effect on the esthetic appearance of maxillary anterior teeth10.1016/0002-9416(84)90300-2 ↗
3. Thanissorn, C., Guo, J., Jing Ying Chan, D., Koyi, B., Kujan, O., Khzam, N., and Miranda, L.A. (2022). Success Rates and Complications Associated with Single Immediate Implants: A Systematic Review. Dent. J., 10.10.3390/dj10020031 ↗
4. Soft tissue-related complications around anterior implants: Commentary and clinical checklist
5. Zirconia as a dental implant abutment material: A systematic review
6. A systematic review of the performance of ceramic and metal implant abutments supporting fixed implant reconstructions10.1111/j.1600-0501.2009.01787.x ↗
7. Implant- supported mandibular overdentures: A cross-sectional study10.1111/j.1600-0501.2011.02225.x ↗
8. Ridge augmentation by applying bioresorbable membranes and deproteinized bovine bone mineral: A report of twelve consecutive cases10.1111/j.1600-0501.2007.01407.x ↗
9. Esthetic high-strength implant abutments. Part I10.1111/j.1708-8240.1993.tb00741.x ↗
10. Current classification of zirconia in dentistry: An updated review10.7717/peerj.15669 ↗
11. CAD/CAM and telescopic technology: Design options for implant-supported overdentures
12. De Angelis, P., Gasparini, G., Rella, E., De Angelis, S., Grippaudo, C., D’Addona, A., and Manicone, P.F. (2021). Patient Satisfaction with Implant-Supported Monolithic and Partially Veneered Zirconia Restorations. Biomed. Res. Int., 2021.10.1155/2021/6692939 ↗
13. Kongkiatkamon, S., Booranasophone, K., Tongtaksin, A., Kiatthanakorn, V., and Rokaya, D. (2021). Comparison of fracture load of the four translucent zirconia crowns. Molecules, 26.10.3390/molecules26175308 ↗
14. Zirconia in dentistry: Evidence-based clinical breakthrough
15. Evaluating dental zirconia10.1016/j.dental.2018.08.291 ↗
16. Are zirconia crowns the superior choice when restoring primary posterior molars?10.1038/s41432-022-0275-6 ↗
17. Ban, S. (2021). Classification and Properties of Dental Zirconia as Implant Fixtures and Superstructures. Materials, 14.10.3390/ma14174879 ↗
18. Can the Application of Multi-Mode Adhesive be a Substitute to Silicatized/Silanized Y-TZP Ceramics?10.1590/0103-6440201801862 ↗
19. Heboyan, A., Vardanyan, A., Karobari, M.I., Marya, A., Avagyan, T., Tebyaniyan, H., Mustafa, M., Rokaya, D., and Avetisyan, A. (2023). Dental Luting Cements: An Updated Comprehensive Review. Molecules, 28.10.3390/molecules28041619 ↗
20. Kontonasaki, E., Rigos, A.E., Ilia, C., and Istantsos, T. (2019). Monolithic Zirconia: An Update to Current Knowledge. Optical Properties, Wear, and Clinical Performance. Dent. J., 7.10.3390/dj7030090 ↗
21. A revised relationship between fracture toughness and Y2O3 content in modern dental zirconias10.1016/j.jeurceramsoc.2021.08.003 ↗
22. Survey of the mechanical and physical behaviors of yttria-stabilized zirconia from multiple dental laboratories
23. Clinical evaluation of monolithic zirconia crowns for posterior teeth restorations10.1097/MD.0000000000017385 ↗
24. Clinical outcomes of three different crown systems with CAD/CAM technology10.1016/j.prosdent.2014.05.002 ↗
25. Effect of Surface Treatment and Cement on Fracture Load of Traditional Zirconia (3Y), Translucent Zirconia (5Y), and Lithium DisilicateCrowns10.1111/jopr.13088 ↗
26. Effectiveness of zirconia crowns compared with stainless steel crowns in primary posterior teeth rehabilitation: A systematic review and meta-analysis10.1016/j.adaj.2021.08.005 ↗
27. Comparison of survival and complication rates of tooth-supported fixed dental prostheses (FDPs) and implant-supported FDPs and single crowns (SCs)10.1111/j.1600-0501.2007.01439.x ↗
28. Comparing the influence of cuspal angulation, occlusal loading, and connector widths between tooth- and implant-supported zirconia fixed dental prosthesis10.1016/j.mjafi.2023.10.004 ↗
29. Influence of connector cross-sectional geometry on the load-bearing capacity under fatigue of implant-supported zirconia fixed partial prosthesis10.1016/j.prosdent.2022.10.008 ↗
30. Selva-Otaolaurruchi, E.J., Fernández-Estevan, L., Solá-Ruiz, M.F., García-Sala-Bonmati, F., Selva-Ribera, I., and Agustín-Panadero, R. (2023). Graphene-Doped Polymethyl Methacrylate (PMMA) as a New Restorative Material in Implant-Prosthetics: In Vitro Analysis of Resistance to Mechanical Fatigue. J. Clin. Med., 12.10.3390/jcm12041269 ↗
31. Approach for valuating the influence of laboratory simulation10.1016/j.dental.2008.08.009 ↗
32. Spitznagel, F.A., Hoppe, J.S., Bonfante, E.A., Campos, T.M.B., Langner, R., and Gierthmuehlen, P.C. (2022). Failure Load and Fatigue Behavior of Monolithic and Bi-Layer Zirconia Fixed Dental Prostheses Bonded to One-Piece Zirconia Implants. Materials, 15.10.3390/ma15238465 ↗
33. García-Sala Bonmatí, F., Bustamante-Hernández, N., Alonso Pérez-Barquero, J., Maneiro-Lojo, J., Fons-Badal, C., Labaig-Caturla, C., Fernández-Estevan, L., and Agustín-Panadero, R. (2023). Implant-Supported Fixed Partial Dentures with Posterior Cantilevers: In Vitro Study of Mechanical Behavior. Materials, 16.10.3390/ma16206805 ↗
34. Evaluation of an endosseous oral implant system according to UNI EN ISO 14801 fatigue test protocol
35. Finite Element Analysis: Connector Designs and Pontic Stress Distribution of Fixed Partial Denture Implant-Supported Metal Framework10.1615/JLongTermEffMedImplants.2023048378 ↗
36. Lifetime-limiting strength degradation from contact fatigue in dental ceramics10.1177/00220345000790020501 ↗
37. Fracture strength and failure mode of five different single-tooth implant-abutment combinations
38. Influence of cyclic loading and luting agents on the fracture load of two all-ceramic crown systems10.1016/j.prosdent.2004.09.002 ↗
39. Fracture resistance of yttrium oxide partially-stabilized zirconia all-ceramic bridges after veneering and mechanical fatigue testing10.1016/j.dental.2004.07.013 ↗
40. Damage accumulation and fatigue life of particle-abraded ceramics
41. Fatigue of zirconia under cyclic loading in water and its implications for the design of dental bridges10.1016/j.dental.2005.12.008 ↗
42. Influence of cyclic fatigue in water on the load-bearing capacity of dental bridges made of zirconia10.1016/j.actbio.2008.04.012 ↗
43. Influence of artificial aging on the load-bearing capability of straight or angulated zirconia abutments in implant/tooth-supported fixed partial dentures
44. Fracture behavior of straight or angulated zirconia implant abutments supporting anterior single crowns10.1007/s00784-009-0377-y ↗
45. Low-temperature degradation of different zirconia ceramics for dental applications10.1016/j.actbio.2011.11.016 ↗
46. Influence of surface treatment of yttria-stabilized tetragonal zirconia polycrystal with hot isostatic pressing on cyclic fatigue strength10.4012/dmj.2012-247 ↗
47. Alumina reinforced zirconia implants: Effects of cyclic loading and abutment modification on fracture resistance10.1016/j.dental.2014.12.013 ↗
48. In vitro fatigue and fracture resistance of one- and two-piece CAD/CAM zirconia implant abutments10.11607/jomi.3942 ↗
49. Approach for valuating the significance of laboratory simulation10.1016/j.jdent.2008.09.001 ↗
50. Bending moments of zirconia and titanium implant abutments supporting all-ceramic crowns after aging10.1111/clr.12192 ↗
51. Survival rate, fracture strength and failure mode of ceramic implant abutments after chewing simulation10.1111/j.1365-2842.2005.01515.x ↗
52. Static and fatigue mechanical behavior of three dental CAD/CAM ceramics10.1016/j.jmbbm.2016.01.023 ↗
53. The influence of support properties and complexity on fracture strength and fracture mode of all-ceramic fixed dental prostheses10.3109/00016357.2010.549508 ↗
54. The fracture resistance of all-ceramic crowns on supporting structures with different elastic moduli
55. Fracture resistance of 3-unitmonolithic ZrO2 ceramics FPDs with different preparation designs ofthe distal abutment—An in-vitro study10.3897/folmed.65.e77442 ↗
56. Effect of materials on axial displacement and internal discrepancy of cement-retained implant-supported prostheses10.1016/j.prosdent.2020.08.044 ↗
57. Dotooth-supported zirconia restorations present more technical failures related to fracture or loss of retention? Systematic review and meta-analysis10.1007/s00784-022-04573-z ↗
58. Zirconium Oxide Three-Unit Fixed Partial Denture Frameworks Supported by Dental Implants in Acceptable and Reduced Interocclusal Space Possibilities: Pilot In Vitro Fracture Strength and Fractographic Analyses10.11607/jomi.7009 ↗
59. Applications of three-dimensional printers in prosthetic dentistry10.2334/josnusd.21-0072 ↗
60. Clinical performance of monolithic CAD/CAM tooth-supported zirconia restorations: Systematic review and meta-analysis10.2186/jpr.JPR_D_21_00081 ↗
61. Fracture loads of screw-retained implant-supported zirconia prostheses after thermal and mechanical stress10.1016/j.jpor.2019.09.003 ↗
62. Ten-year clinical performance of zirconia posterior fixed partial dentures10.1111/joor.13276 ↗
63. The effect of cyclic loading on the fracture resistance of 3D-printed and CAD/CAM milled zirconia crowns-an in vitro study10.1007/s00784-023-05229-2 ↗
64. Clinical outcomes of implant-supported monolithiczirconia crowns and fixed partial dentures: A systematic review10.1111/jopr.13575 ↗
65. Fracture rate of 188695lithium disilicate and zirconia ceramic restorations after up to 7.5 years ofclinical service: A dental laboratory survey10.1016/j.prosdent.2019.06.011 ↗