journal article·2025-01-22Open access

Comparative Biomechanical Evaluation of Novel Screwless Retained Dental Implant Prosthesis: A 3D Finite Element Analysis

Ki-Sun Lee, Jaeyeol Kim, JaeHyung Lim, Jae-Jun Ryu

Abstract

This study aimed to comparatively evaluate the biomechanical behaviors of three types of dental implant restorations: a screw-and-cement-retained prosthetic system (SCRP); a cementless screw-retained prosthetic system (SRP); and a novel screwless hook-retained prosthetic system (HRP). Three-dimensional finite element analysis (FEA) was used to evaluate biomechanical behavior. A comparative study of three dental implant prostheses was performed under two loading conditions: a vertical load of 100 N and an oblique load of 100 N at an angle of 30°. Under both loading conditions, the maximum von Mises stress values in the dental implant using the HRP system were lower (21.33 MPa) than those of the SCRP system (32.91 MPa), and the stress distribution of the implant prosthetic components tended to be more favorable than that of the SCRP system. Thus, the results show that the performance of the HRP system was comparable to that of a conventional SRP system under the same conditions. Regarding stress distribution, the novel screwless HRP system presents a viable alternative implant prosthodontic system to the conventional SCRP system.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Dental Implant Techniques and Outcomes · Dental materials and restorations · Bone Tissue Engineering Materials

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Prosthodontics, Korea University Ansan Hospital, Ansan-si 15355, Republic of Korea

Medical Science Research Center, Korea University Ansan Hospital, Ansan-si 15355, Republic of Korea

Department of Medicine, Graduate School, Korea University, Seoul 02841, Republic of Korea

Department of Oral & Maxillofacial Surgery, Korea University Ansan Hospital, Ansan-si 15355, Republic of Korea

Department of Prosthodontics, Korea University Anam Hospital, Seoul 02841, Republic of Korea

Funding

Korea Medical Device Development Fund grant funded by the Korea government (the Ministry of Science and ICT, the Ministry of Trade, Industry and Energy, the Ministry of Health & Welfare, the Ministry of Food and Drug Safety) · Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education · Korea University Ansan Hospital

References (27)

Show source-provided references
1. A meta-analysis of implants in partial edentulism10.1034/j.1600-0501.1998.090203.x ↗
2. Evaluation of survival and success rates of dental implants reported in longitudinal studies with a follow-up period of at least 10 years: A systematic review10.1016/j.ijom.2014.10.023 ↗
3. Kim, J.-J., Lee, J.-H., Kim, J.C., Lee, J.-B., and Yeo, I.-S.L. (2019). Biological responses to the transitional area of dental implants: Material-and structure-dependent responses of peri-implant tissue to abutments. Materials, 13.10.3390/ma13010072 ↗
4. Li, J., Jansen, J.A., Walboomers, X.F., and van den Beucken, J.J. (2020). Mechanical aspects of dental implants and osseointegration: A narrative review. J. Mech. Behav. Biomed. Mater., 103.10.1016/j.jmbbm.2019.103574 ↗
5. Alemayehu, D.B., and Jeng, Y.R. (2021). Three-Dimensional Finite Element Investigation into Effects of Implant Thread Design and Loading Rate on Stress Distribution in Dental Implants and Anisotropic Bone. Materials, 14.10.20944/preprints202109.0221.v1 ↗
6. Evaluation of design parameters of osseointegrated dental implants using finite element analysis10.1046/j.1365-2842.2002.00891.x ↗
7. Influence of Connections and Surfaces of Dental Implants on Marginal Bone Loss: A Retrospective Study Over 7 to 19 Years10.11607/jomi.8450 ↗
8. The influence of thread geometry on biomechanical load transfer to bone: A finite element analysis comparing two implant thread designs
9. Lee, J.-H., Jang, H.Y., and Lee, S.Y. (2021). Finite element analysis of dental implants with zirconia crown restorations: Conventional cement-retained vs. cementless screw-retained. Materials, 14.10.3390/ma14102666 ↗
10. Cement-retained versus screw-retained implant restorations: A critical review
11. Screw versus cemented implant supported restorations10.1038/sj.bdj.4814157 ↗
12. Comparative evaluation of implant designs: Influence of diameter, length, and taper on strains in the alveolar crest: A three-dimensional finite-element analysis10.1111/j.1600-0501.2005.01132.x ↗
13. Two dental implants designed for immediate loading: A finite element analysis
14. Effect of engaging abutment position in implant-borne, screw-retained three-unit fixed cantilevered prostheses
15. Passivity of fit and marginal opening in screw-or cement-retained implant fixed partial denture designs
16. The preliminary study for three-dimensional alveolar bone morphologic characteristics for alveolar bone restoration10.1186/s40902-019-0216-2 ↗
17. Feedback control during mastication of solid food textures—A clinical-experimental study10.1016/S0022-3913(98)70134-9 ↗
18. Influences of implant neck design and implant-abutment joint type on peri-implant bone stress and abutment micromovement: Three-dimensional finite element analysis10.1016/j.dental.2012.07.160 ↗
19. The influence of implant location and length on stress distribution for three-unit implant-supported posterior cantilever fixed partial dentures10.1016/j.prosdent.2003.12.017 ↗
20. Platform switching: Biomechanical evaluation using three-dimensional finite element analysis
21. Analysis of bone-implant interaction phenomena by using a numerical approach10.1111/j.1600-0501.2005.01162.x ↗
22. Stress distribution pattern of screw-retained restorations with segmented vs. non-segmented abutments: A finite element analysis10.15171/joddd.2017.027 ↗
23. Vertical load distribution on a three-unit prosthesis supported by a natural tooth and a single Brånemark implant. An in vivo study10.1034/j.1600-0501.1995.060105.x ↗
24. Effect of passive fit absence in the prosthesis/implant/retaining screw system: A two-dimensional finite element analysis10.1097/SCS.0b013e3181bd2df8 ↗
25. A biomechanical effect of wide implant placement and offset placement of three implants in the posterior partially edentulous region10.1046/j.1365-2842.2000.00475.x ↗
26. Effect of offset implant placement on the stress distribution around a dental implant: A three-dimensional finite element analysis10.1563/AAID-JOI-D-13-00163 ↗
27. Influence of implant angulation with different crowns on stress distribution10.1097/SCS.0b013e318207477c ↗