Fracture Resistance of a Two-Piece Zirconia Implant System after Artificial Loading and/or Hydrothermal Aging—An In Vitro Investigation
Source abstract
Abstract
The purpose of the present study was to assess the fracture resistance of a two-piece alumina-toughened zirconia implant system with a carbon-reinforced PEEK abutment screw. Methods: Thirty-two implants with screw-retained zirconia abutments were divided into four groups of eight samples each. Group 0 (control group) was neither loaded nor aged in a chewing simulator; group H was hydrothermally aged; group L was loaded with 98 N; and group HL was subjected to both hydrothermal aging and loading in a chewing simulator. One sample of each group was evaluated for t-m phase transformation, and the others were loaded until fracture. A one-way ANOVA was applied to evaluate differences between the groups. Results: No implant fracture occurred during the artificial chewing simulation. Furthermore, there were no statistically significant differences (p > 0.05) between the groups in terms of fracture resistance (group 0: 783 ± 43 N; group H: 742 ± 43 N; group L: 757 ± 86 N; group HL: 740 ± 43 N) and bending moment (group 0: 433 ± 26 Ncm; group H: 413 ± 23 Ncm; group L: 422 ± 49 Ncm; group HL: 408 ± 27 Ncm). Conclusions: Within the limitations of the present investigation, it can be concluded that artificial loading and hydrothermal aging do not reduce the fracture resistance of the investigated implant system.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Dental Implant Techniques and Outcomes · Dental materials and restorations · Bone Tissue Engineering Materials
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Medical Center—University of Freiburg, Center for Dental Medicine, Department of Prosthetic Dentistry, Faculty of Medicine, University of Freiburg, 79106 Freiburg, Germany
RISE Research Institutes of Sweden, 431 53 Mölndal, Sweden
Medical Center—University of Freiburg, Institute for Medical Biometry and Statistics, Faculty of Medicine, University of Freiburg, 79104 Freiburg, Germany
Private Dental Clinic, 78658 Zimmern ob Rottweil, Germany
References (56)
Show source-provided references
1. A 15-year study of osseointegrated implants in the treatment of the edentulous jaw10.1016/S0300-9785(81)80077-4 ↗
2. The long-term efficacy of currently used dental implants: A review and proposed criteria of success
3. Herrero-Climent, M., Lopez-Jarana, P., Lemos, B.F., Gil, F.J., Falcao, C., Rios-Santos, J.V., and Rios-Carrasco, B. (2020). Relevant design aspects to improve the stability of titanium dental implants. Materials, 13.10.3390/ma13081910 ↗
4. Low-temperature degradation of y-tzp ceramics: A systematic review and meta-analysis10.1016/j.jmbbm.2015.10.017 ↗
5. Osseointegrated implants in the treatment of the edentulous jaw. Experience from a 10-year period
6. Osseointegrated titanium fixtures in the treatment of edentulousness10.1016/0142-9612(83)90065-0 ↗
7. Survival of the Brånemark implant in partially edentulous jaws: A 10-year prospective multicenter study
8. A mechanical evaluation of implants placed with different surgical techniques into the trabecular bone of goats10.1563/0-827.1 ↗
9. Influence of the surgical technique and surface roughness on the primary stability of an implant in artificial bone with a density equivalent to maxillary bone: A laboratory study10.1111/j.1600-0501.2008.01692.x ↗
10. Longevity of dental implants in type iv bone: A systematic review10.1016/j.ijom.2014.02.016 ↗
11. Callejas, J.A., Gil, J., Brizuela, A., Perez, R.A., and Bosch, B.M. (2022). Effect of the size of titanium particles released from dental implants on immunological response. Int. J. Mol. Sci., 23.10.3390/ijms23137333 ↗
12. Hypersensitivity to titanium: Clinical and laboratory evidence
13. Titanium allergy in dental implant patients: A clinical study on 1500 consecutive patients10.1111/j.1600-0501.2008.01544.x ↗
14. Changes in the esthetic, physical, and biological properties of a titanium alloy abutment treated by anodic oxidation10.1016/j.prosdent.2018.03.024 ↗
15. Transformation toughening. Part 3 experimental observations in the ZrO2-Y2O3 system10.1007/BF00809059 ↗
16. In vitro biofilm formation on titanium and zirconia implant surfaces10.1902/jop.2016.160245 ↗
17. In vitro evaluation of the mutagenic and carcinogenic power of high purity zirconia ceramic10.1016/S0142-9612(98)00182-3 ↗
18. Clinical outcomes of zirconia dental implants: A systematic review10.1177/0022034516664043 ↗
19. Performance and outcome of zirconia dental implants in clinical studies: A meta-analysis10.1111/clr.13352 ↗
20. Reliability of one-piece ceramic implant10.1002/jbm.b.31113 ↗
21. In vitro performance of zirconia and titanium implant/abutment systems for anterior application10.1016/j.jdent.2014.03.010 ↗
22. Clinical performance of two-piece zirconia implants in the posterior mandible and maxilla: A prospective cohort study over 2 years10.1111/clr.12610 ↗
23. Two-piece zirconia implants supporting all-ceramic crowns: A prospective clinical study10.1111/clr.12370 ↗
24. All-ceramic restoration of zirconia two-piece implants—A randomized controlled clinical trial10.1111/clr.12342 ↗
25. Two-piece zirconia oral implants withstand masticatory loads: An investigation in the artificial mouth10.1016/j.jmbbm.2015.07.005 ↗
26. Stability and aging resistance of a zirconia oral implant using a carbon fiber-reinforced screw for implant-abutment connection10.1016/j.dental.2018.08.290 ↗
27. Fracture resistance and crystal phase transformation of a one- and a two-piece zirconia implant with and without simultaneous loading and aging-an in vitro study10.1111/clr.13825 ↗
28. Stability of prototype two-piece zirconia and titanium implants after artificial aging: An in vitro pilot study10.1111/j.1708-8208.2008.00116.x ↗
29. Influence of loading and aging on the fracture strength of an injection-molded two-piece zirconia implant restored with a zirconia abutment10.1111/clr.14022 ↗
30. Osseointegration of zirconia in the presence of multinucleated giant cells10.1111/cid.12375 ↗
31. (2017). Dentistry—Implants—Dynamic Fatigue Test for Endosseous Dental Implants (Standard No. ISO 14801:2016).
32. Average grain size in polycrystalline ceramics10.1111/j.1151-2916.1969.tb11975.x ↗
33. Towards long lasting zirconia-based composites for dental implants: Transformation induced plasticity and its consequence on ceramic reliability10.1016/j.actbio.2016.11.040 ↗
34. Burkhardt, F., Harlass, M., Adolfsson, E., Vach, K., Spies, B.C., and Kohal, R.J. (2021). A novel zirconia-based composite presents an aging resistant material for narrow-diameter ceramic implants. Materials, 14.10.3390/ma14092151 ↗
35. Kohal, R.J., Trinkner, A., Burkhardt, F., Patzelt, S.B.M., Vach, K., Kuster, M., Abram, A., Kocjan, A., and Nold, J. (2023). Long-term stability of hydrothermally aged and/or dynamically loaded one-piece diameter reduced zirconia oral implants. J. Funct. Biomater., 14.10.3390/jfb14030123 ↗
36. Bethke, A., Pieralli, S., Kohal, R.J., Burkhardt, F., von Stein-Lausnitz, M., Vach, K., and Spies, B.C. (2020). Fracture resistance of zirconia oral implants in vitro: A systematic review and meta-analysis. Materials, 13.10.3390/ma13030562 ↗
37. In vivo forces on implants influenced by occlusal scheme and food consistency
38. Zhang, F., Meyer Zur Heide, C., Chevalier, J., Vleugels, J., Van Meerbeek, B., Wesemann, C., Camargo Dos Santos, B., Sergo, V., Kohal, R.J., and Adolfsson, E. (2020). Reliability of an injection-moulded two-piece zirconia implant with pekk abutment after long-term thermo-mechanical loading. J. Mech. Behav. Biomed. Mater., 110.10.1016/j.jmbbm.2020.103967 ↗
39. Time-frequency analysis of chewing activity in the natural environment10.1177/0022034511416669 ↗
40. Influence of stress simulation parameters on the fracture strength of all-ceramic fixed-partial dentures10.1016/j.dental.2005.04.024 ↗
41. The wear of dental amalgam in an artificial mouth: A clinical correlation10.1016/S0109-5641(85)80050-6 ↗
42. The wear of a posterior composite in an artificial mouth: A clinical correlation10.1016/S0109-5641(86)80034-3 ↗
43. In-vitro-testverfahren zur evaluation dentaler restaurationssysteme. 3. Korrelation mit in-vivo-resultaten
44. An artificial oral environment for testing dental materials10.1109/10.133228 ↗
45. Quality assessment of the cortical bone of the human mandible10.1016/S8756-3282(02)00921-3 ↗
46. Kohal, R.J., Spies, B.C., Vach, K., Balmer, M., and Pieralli, S. (2020). A prospective clinical cohort investigation on zirconia implants: 5-year results. J. Clin. Med., 9.10.3390/jcm9082585 ↗
47. Prospective clinical multicenter study evaluating the 5-year performance of zirconia implants in single-tooth gaps10.11607/jomi.9289 ↗
48. Influence of hydrothermal and mechanical conditions on the strength of zirconia10.1016/j.actbio.2010.07.025 ↗
49. Low temperature degradation -aging- of zirconia: A critical review of the relevant aspects in dentistry10.1016/j.dental.2010.04.006 ↗
50. Transformation toughening in zirconia-containing ceramics10.1111/j.1151-2916.2000.tb01221.x ↗
51. On the kinetics and impact of tetragonal to monoclinic transformation in an alumina/zirconia composite for arthroplasty applications10.1016/j.biomaterials.2009.06.022 ↗
52. Ban, S. (2021). Classification and properties of dental zirconia as implant fixtures and superstructures. Materials, 14.10.3390/ma14174879 ↗
53. Low-temperature degradation of different zirconia ceramics for dental applications10.1016/j.actbio.2011.11.016 ↗
54. In vitro performance of two-piece zirconia implant systems for anterior application10.1016/j.dental.2016.03.028 ↗
55. Comparison between peek and Ti6AL4V concerning micro-scale abrasion wear on dental applications10.1016/j.jmbbm.2015.12.038 ↗
56. Fracture resistance of abutment screws made of titanium, polyetheretherketone, and carbon fiber-reinforced polyetheretherketone10.1590/1807-3107BOR-2014.vol28.0028 ↗